NivoRadar
®
Continuous level measurement
Series NR 4000
Technical information / Instruction manual
Mounting
page 23
interfering reflections from the agitators are saved with the blades in
different positions.
Fig. 20: Agitators
Through the action of filling, stirring and other processes in the ves-
sel, compact foams which considerably damp the emitted signals
may form on the medium surface.
Note:
If foams lead to measurement errors, you should use the biggest
possible radar antennas or as an alternative, sensors with guided
radar.
Foam generation