manualshive.com logo in svg
background image

DM SERIES  

 

 

 

ELECTRONIC WEIGHING SCALE 

 

 

OPERATION MANUAL 

 

 

 

PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE 

ATTEMPT TO OPERATE THE SCALE 

 

 

 

 

 

 

 

 

FEBRUARY 2002  

 

Specifications subject to change without prior notice

 

CONTENTS 

 

CAUTIONS 

 

1. INSTALLATION 

 

2. SPECIFICATIONS 

 

3. KEYBOARD LAYOUT AND DESCRIPTION 

 

4. INITIAL SETUP 

 

5. INSTRUCTION FOR USE 

 

6. TROUBLE SHOOTING 

 

7. DAILY CARE AND MAINTENANCE 

 

 

Содержание DM-1100

Страница 1: ...CAREFULLY BEFORE ATTEMPT TO OPERATE THE SCALE FEBRUARY 2002 Specifications subject to change without prior notice CONTENTS CAUTIONS 1 INSTALLATION 2 SPECIFICATIONS 3 KEYBOARD LAYOUT AND DESCRIPTION 4 INITIAL SETUP 5 INSTRUCTION FOR USE 6 TROUBLE SHOOTING 7 DAILY CARE AND MAINTENANCE ...

Страница 2: ...n where or when the environment condition falls beyond as those listed on SPECIFICATIONS 3 Do not attempt to open this unit or conduct any trouble shootings other than those listed on TROUBLE SHOOTING 4 Do not submerge scale into water 5 Do not use pressurized water nozzle for cleaning Do not use this scale for chemical substances or in any environment that contains chemical vapor 6 Precision meas...

Страница 3: ... Plastic platter x 1 Stainless steel cover x 1 Power adaptor x 1 Contact your dealer if any item missed FIG 1 FIG 2 1 2 Locks the plastic platter FIG 1 A to the platter holder with the screw FIG 1 B included and then apply the rubber cap FIG 1 C 1 3 Place the stainless steel platter FIG 1 D onto the plastic platter ...

Страница 4: ... jack of scale 2 SPECIFICATIONS Model No Capacity Division Accuracy Min Load DM 3000 3000g 1g 1 3000 20g DM 6000 6000g 2g 1 3000 40g DM 15K 15kg 0 005kg 1 3000 0 1g DM 30K 30kg 0 01kg 1 3000 0 2g DM 1100 1100g 0 1g 1 11000 2g DM 11K 11kg 1g 1 11000 20g Tare Range Full tare range Subtractive Display 5 x 24mm LCD display Operation Environment 0o 40o C 32o 104o F Non condensed R H Ø 85 Power source B...

Страница 5: ...ss this key to turn scale off 5 ZERO INDICATOR This indicator appears to indicate a zero weight status 6 TARE INDICATOR This indicator appears to indicate the tare function is in operation and weight displayed is net weight 7 WEIGHT UNIT INDICATOR This indicator appears to indicate the weighing unit being selected is at metric 8 LO BAT INDICATOR When this indicator appears recharge the scale immed...

Страница 6: ...SCALE In order to obtain an accurate weighing result the scale must be placed on a strong and level surface If required adjust the adjustable feet FIG 2 F underneath scale to obtain a level position Refer to the bubble indicator for leveling condition 4 2 INTERNAL SETTING 4 2 1 Select Weight Unit and Display Resolution The scale is equipped with metric and avoirdupois weight units Default setting ...

Страница 7: ... Turn scale off b Press and hold TARE then press ON ZERO c Scale displays F 1 d Press TARE three times e Scale displays F 4 f Press MODE to shift between 0 _OFF and 4 _OFF g To enable Auto Power Off Function press ON ZERO when 4 _OFF appears h To disable Auto Power Off Function press ON ZERO when 0 _OFF appears i Press ON ZERO to confirm and return to normal operation 4 3 SELECT BAUD RATE SETTING ...

Страница 8: ...and the stainless steel platter cover is placed on the plastic platter b The scale is placed on a level and strong surface c The scale is turned on d The ZERO INDICATOR is on If not press ZERO ON to set display to zero 5 2 WEIGHING a Always place an object onto the platter gently Excessive force applied to platter may cause damages to sensor inside scale b The weight of the object would be display...

Страница 9: ...iminate a positive or negative value After the ON ZERO is pressed the ZERO INDICATOR appears Refer to SPECIFICATIONS for maximum zero range 5 6 TARE OFF THE WEIGHT OF A CONTAINER When a container is used follow the below steps to tare off the weight of the container and get the net weight a Remove all loads away from platter b Make sure that the ZERO INDICATOR is on c Place the container on the pl...

Страница 10: ... cannot be turned on Check Is the scale charged Action Recharge the scale for at least 8 hours before first time use or plug in the power adaptor before power on the scale Check Is the power adaptor inserted properly into both the electricity outlet and the DC inlet of scale Action Secure both ends of the power adaptor and try again Syndrome Scale turned off automatically Check Is Auto Power Off F...

Страница 11: ...level condition Check Is the scale affected by airflow vibration or RFI Action Place the scale away from all disturbances Check Is the scale calibrated correctly Action Contact your dealer Check Is the correct weight unit employed Action Refer to 4 2 1 and 5 3 for weight unit setting and usage Syndrome Scale displays 00000 Check Is the scale overloaded Action Remove some loads from platter Syndrom...

Страница 12: ...cally Re calibrate the scale if necessary Note In some countries calibration is restricted to be done by an authorized qualified agent only Contact your dealer for more information 7 4 It is a good practice to apply the dust cover when operating the scale 7 5 The scale must be placed horizontally during transportation or long time storage 7 6 Remove platter from scale before transportation or long...

Отзывы: