4. INITIAL SETUP
4.1 PLACING THE SCALE
In order to obtain an accurate weighing result, the scale
must be placed on a strong and level surface. If required,
adjust the adjustable feet (
FIG 2 F
) underneath scale
to obtain a level position. Refer to the bubble indicator
for leveling condition.
4.2 INTERNAL SETTING
4.2.1 Display Segment Check
User can enter this function to check to see if the display
and backlight (if purchased) are functioning properly.
a.
Scale is off
b.
Press and hold
TARE
, then press
ON/ZERO
c.
Scale displays
F.1
d.
Press
TARE
once
e.
Scale displays
F.2
f.
Press
MODE
and all segments will light up to allow
user to check the condition of display
g.
Press
TARE
for next function setup or
ON/ZERO
to
restart the scale
4.2.2 Select Auto Power Off Setting
The scale is equipped with
Auto Power Off Function
. The
scale will power off automatically if it is not used
for 4 minutes. Follow the below steps to enable/disable
Auto Power Off Function.