UV Process Supply BETACOLOR S1 Скачать руководство пользователя страница 8

ADJ

1000

A

B

 

Factory slope setting of the yellow channel

 

 
1.Put the unit in 

AUTO 

mode by pressing 

B

 and holding 

for three seconds. 

2.Measure the solid process color standard.  

AUTO

DENS

, and the color indicator will appear along with the 

measured value.  

3.Raise the head and press 

and 

B

 simultaneously.  

ADJ 

will appear and the current  slope value will be displayed.  
Raise or lower this value directly by pressing 

A

 or 

B

.

  

 

4.Measure the sample again.  

ADJ DENS

 will appear and 

the density value can be changed directly by pressing 

A

 

or 

B

 after raising the head. 

 

 

8

Содержание BETACOLOR S1

Страница 1: ...the need of power consuming motors or bulky rechargeable battery packs The AC charger and internal charging circuits have been eliminated Color density measurements are made simultaneously in all four...

Страница 2: ...tting Started Open the densitometer by sliding the gray release latch forward allowing the aperture to spring forward Tap either button to turn on the unit if it has powered down Select AUTO mode with...

Страница 3: ...the paper is automatically zeroed out helping the pressman maintain optimum appearance across varying paper stocks Strongly tinted stocks can be zeroed out manually Place the aperture on the solid col...

Страница 4: ...ll four color channels each time a sample is measured The individual color channel density values can be displayed by stepping through the color channels with button B the right hand button This infor...

Страница 5: ...ro all channels Flashing AUTO will display and 0 00 will appear in the readout 3 Measure the solid target DENS will appear the density will be displayed and the color indicator will move to the proper...

Страница 6: ...ro function 1 Tap either key to activate the instrument If MAN DENS does not appear at the right of the display press and hold B for three seconds with the head raised MAN DENS 000 A B Yellow channel...

Страница 7: ...own The density of the sample as measured in the selected color channel will be displayed The other color component values of that color may be displayed by pressing B after raising the head AUTO DENS...

Страница 8: ...DENS and the color indicator will appear along with the measured value 3 Raise the head and press A and B simultaneously ADJ will appear and the current slope value will be displayed Raise or lower t...

Страница 9: ...adjusting all four channels as described above 6 Press A and B simultaneously with the head raised to exit the ADJ mode Serial Port Baud Rate Adjustment Adjust the baud rate of the unit to match the r...

Страница 10: ...unit and remove the battery cover plate Discard the old batteries and install two new ones in the same positions Reset Button If the unit fails to respond or an unusual display appears press the red...

Страница 11: ...e and date stamped data and plots can be saved to disk or printed in color The network client version supports centralized data collection Statistical analysis screen of QC01 BetaTab Data Collection S...

Страница 12: ...a Color Proofing Viewer is fitted with a 10x color corrected lens a brilliant tungsten halogen lamp and color separation filters Verify register dot quality coverage and more in every color including...

Страница 13: ...8 pin RS232 300 9600 baud Weight 10 4 oz 290 g Dimensions 3 x 5 5 x 1 5in 145 x 75 x 40mm Warranty Your densitometer is warranted to be free of manu facturing defects for two years from the date of pu...

Отзывы: