ANNEX III – EMC MANUFACTURER’S DECLARATION
Electromagnetic Compatibility (EMC) Statement for Home Healthcare Environment the Desktop
Device System has been evaluated to international standard IEC 60601-1-2 “General requirements
for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -
Requirements and tests”.
1.
The Medical Electrical Equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided here within the
accompanying documents.
2.
Portable and mobile RF communication can affect the Medical Electrical Equipment. See below
recommended separation distances between portable and mobile RF communication equipment
and the desktop device system.
3.
Wireless communications equipment such as wireless home network devices, mobile phones,
cordless telephones and their base stations, walkie-talkies can affect this equipment and should be
kept at least a distance d=3.3 m way from the system.
Rated Maximum Output Power of
Transmitter (W)
Separation Distance* according to
Frequency of Transmitter (m)
0.01
0.23
0.1
0.73
1
2.3
10
7.3
100
23
NOTE - The distance calculated from 800 MHZ to 2,5 GHZ
Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may
result in incorrect operation of the unit.
This system should not be used adjacent to or stacked with other equipment. If adjacent or stacked
use is necessary, observe the device function to verify normal operation in the configuration in which
it will be used.