8 Setting Window
The Setting window has the following main setting items.
H
ow t
o u
se
Bla
deN
av
iSE
item
Explanation
■
Cal Wavelength
Only UIT-θLED can be set. (Not selectable for UIT-θ365.)
You can select from "385nm", "395nm", "405nm", and "450nm" depending on the
wavelength of the LED light source.
■
Cal Coefficlant
Cal Coefficlant is a function that corrects the irradiance value.
Please refer to "6.Cal Coefficlant" for the setting.
■
Mes Range
Set to the same measurement range as the measurement range of the main unit.For
details, please refer to "2.2 Preparing BladeNavi SE".
■
Sampling Rate
Sets the sampling rate for logging measurements.
Example: “512Hz” is measured at a timing of 512 times per second.
■
Sleep Setting time
Set the transition time from logging standby mode to sleep mode.
Example: When set to "5min", it sleeps when irradiance = 0 continues for 5 minutes or
more in logging standby mode.
■
Logging Delay Time
■
Continuous Logging
Set the time threshold for starting logging measurement.
In logging standby mode, logging measurement starts when irradiance is
continuously detected for the time threshold. Enter the time threshold as an integer
value from 1 to 255, and the unit is 0.1 sec.
If you check the Continuous Logging checkbox, logging measurement will start
without going through the logging standby mode.
■
Create
Maintenance
File
Not normally used.
Output of diagnostic file.
When you make an inquiry due to a problem with the main unit, please provide the
file saved by this function so that the diagnosis will proceed smoothly.
27
Copyright © 2019, USHIO INC. All Rights Reserved.
DOC1050921-03