IM 50.560CA UA (B/4-03)
9
DEPOLOX
®
3
PLUS
RESIDUAL ANALYZER
connection of cables. The device may be opened and adjusted by an
electrician only.
For connection to power with two hot wires (L1, L2), replace the metal
pin on the F2 socket by a fuse corresponding to the voltage (see Figure
2.1).
The switch to adjust the voltage is inside the housing. To open up the
housing push out the two lateral lugs with a screwdriver or similar tool.
Unscrew the two screws on each side of the cover and lift off the cover of
the housing. Should the ribbon cable be pulled off when doing this, it must
be put back when putting back together. The cables must not be twisted
(if the display is not illuminated the two-core ribbon cable has been in-
serted the wrong way round).
Figure 2.1 - Rating Plate
The change in voltage setting must be marked clearly and permanently on
the rating plate as follows (example):
Line voltage changed to 230 V!
Date .... Name/firm/department
If the device fuse blows repeatedly due to a fault, the device must be
repaired or replaced. Contact USF/W&T Technical Service.
The slide switch to adjust the
voltage allows a choice be-
tween 230 V and 115 V.
When setting to 230 V or 115 V, the fuse must
be changed (fine-wire fuse 5 x 20 mm):
for 230 V: T160mA part no. UXF-92568
for 115 V: T315mA part no. UXJ-92568