– Exit on the Settings menu will be changed to Restart. To exit and reset select
Restart and then press
to reset the monitor.
Figure 9: Settings Menu Screen – Restart
2.9 Setting Inflation Target
By default, the BP+ will initially inflate the cuff to 180 mmHg. If the actual systolic
pressure is higher than this, then the device may automatically re-inflate
additional times to measure the systolic pressure. If you have an estimate of the
systolic pressure then you may wish to set the initial inflation target to avoid
these additional inflation cycles.
Press
the Settings menu will be displayed. Select the Inflation Target option
and press
. The Inflation Target menu will be displayed with the current
inflation target highlighted.
Press
to confirm your selection or use the arrow
keys to select another value.
You will be returned to the Settings screen. Note
that the Inflation Target will default to 180 mmHg if the BP+ is restarted.
Figure 10: Inflation Target Setup Screen
Note -
If the subject’s actual systolic pressure is significantly lower than 180
mmHg you may reduce measurement time and discomfort by reducing the initial
inflation target.