19
Overload Protection:
Fuse 1: F 400mA/690V
Fuse 2: F 10A/690V
Max. Input Current:
10A (For inputs > 2A : measurement duration
< 10 secs, interval > 15 minutes)
Frequency Range:
40Hz ~ 400Hz
Response:
Average, calibrated in rms of sine wave
Max. Voltage Drop:
400mV
Note:
The 10A range is specified from 20% to 100% of range.
Frequency
Range
Resolution
Accuracy
10Hz
0.01Hz
± (1.0% + 3)
100Hz
0.1Hz
± (0.8% + 3)
1kHz
0.001kHz
10kHz
0.01kHz
100kHz
0.1kHz
Input Voltage:
1Vrms -20Vrms
Duty Cycle
Range
Resolution
Accuracy
5% ~ 99%
0.1%
1 Hz ~ 10kHz: ± (2% + 5)
> 10kHz: not specified
Input Voltage:
3Vp-p ~ 10Vp-p
Overload Protection:
600V DC/AC rms
Resistance
Range
Resolution
Accuracy
400
ȍ
0.1
ȍ
± (1.0% + 5)
4k
ȍ
0.001k
ȍ
± (1.0% + 3)
40k
ȍ
0.01k
ȍ
400k
ȍ
0.1k
ȍ
4M
ȍ
0.001M
ȍ
40M
ȍ
0.01M
ȍ
± (1.8% + 5)
Capacitance
Range
Resolution
Accuracy
40nF
0.01nF
± (4%+20)(REL)
400nF
0.1nF
± (3%+5)(REL)
4µF
0.001µF
± (4% + 5)
40µF
0.01µF
100µF
0.1µF
± (8% + 5)
For the ranges
)
400nF, use Relative mode (REL) to subtract the stray capacitance of the
test leads and the Meter.
For the range of 100µF, wait about 30 secs for the reading to settle.
Содержание 076 A