US-5011 Indicator 

User/Technical Manual 

 

Content subject to change without notice 

 

Version 1.0 

Содержание US-5011

Страница 1: ...US 5011 Indicator User Technical Manual Content subject to change without notice Version 1 0 ...

Страница 2: ...US 5011 1 Issue AA ...

Страница 3: ...er or exit HOLD mode Setup or Calibration mode Shift the flashing data entry position from right to left Weighing mode Select weight unit of measure Setup or Calibration mode Increase the digit in the flashing data entry position by 1 Weighing mode Tare the weight Setup or Calibration mode Confirm the input data and continue to next step Weighing mode 3 seconds Zero the platform weight 3 seconds P...

Страница 4: ...US 5011 1 2 OVERVIEW OF CONTROLS AND FUNCTIONS Indicator Display Character Definitions ...

Страница 5: ...scale by pressing and holding the ZERO ON OFF key for 4 seconds ZERO If the display dose not shows 0 and there is nothing on the platter press the ZERO ON OFF key to zero the reading Zero range 20 full Capacity The zero function is unavailable when the displayed reading is out of the zero range and the indicator will show the error message 0 or 0 meaning the scale is over or under zero range Setti...

Страница 6: ...ecomes stable it will automatically be directed to next step If the second point cannot be calibrated correctly it will display CAL Er and return back to step2 for re calibration 4 When CAL P2 is displayed the scale will be calibrated on third calibration point xxxxxx kg or lb will be displayed The default standard weight is 100 FS Load 10 100 FS this must be equal or larger than the weight from t...

Страница 7: ...d weight unit if main unit is kg or lb then second unit is lb or kg 100 25000 Note it must be equal or less than 1 25 C1 2500 for 500kg capacity 2000 or 1000kg capacity 3000 for 1500kg capacity C3 0 1 2 Division select 0 1 1 2 2 5 2 for 1000lb 0 for 2000lb and 3000lb capacity C4 0 5 Decimal point 0 x1 1 x0 1 2 x0 01 3 x0 001 4 x0 0001 5 x10 1 for 1000lb 0 for 2000lb and 3000lb capacity C5 0 1 Main...

Страница 8: ... is NOT within the power on zero point range 0 Prompt power on zero point is over range 1 Current weight 2 Calibration zero point 3 Power off zero point power off tare weight as current tare weight 0 C11 6 Zero tracking range 0 no tracking 1 0 25d 2 0 5d 3 1d 4 1 5d 5 2d 6 3d 7 4d 8 5d 9 6d 6 C12 2 Digital filter intensity 0 very weak 1 weak 2 middle 3 strong 2 C13 5 Motion check range 0 0 25d 1 0...

Страница 9: ... 05kg 0 1 lb 0 5kg 0 5kg 1 lb 5kg 5kg 10 lb 50kg 50kg Not available Table2 use LB as calibration unit Calibration division value Display division value in different weight unit that can be used Kg lb 0 0001lb Not available 0 0001lb 0 001 lb 0 0005 kg 0 001 lb 0 01 lb 0 005 kg 0 01 lb 0 1 lb 0 05 kg 0 1 lb 1 lb 0 5 kg 1 lb 10 lb 5 kg 10 lb 0 0002 lb 0 0001 kg 0 0002 lb 0 002 lb 0 001 kg 0 002 lb 0 ...

Страница 10: ... 1 In normal weighing mode press UNIT and ZERO ON OFF until USer is shown to enter in the mode 2 Press the UNIT key to change the flashed digits press the HOLD PRINT key to shift the flashed position Press the TARE key to confirm and save the set data and enter next setting Press the ZERO ON OFF key to exit this mode 3 Summary of User Parameters Setting Parameter Option Setting setting U1 0 15 Aut...

Страница 11: ...are Serial communication function can be only activated when the scale is in normal weighing mode 2 U5 1 Press PRINT key the scale will output the current stable weight weight unit and current Status data it does not receive any data The output format is as below LF weight reading minus decimal point weight unit CR LF H1H2H3 CR ETX 3 U5 2 Press PRINT key the scale will output the data of stable gr...

Страница 12: ...two for lb and one for a space sp after lb Units of measure abbreviations are always lower case 1 If the weight is overcapacity the scale will return ten characters the field of minus sign decimal point weight data is filled by 2 If the weight is under capacity it will return ten _ characters the field of minus sign decimal point and weight data is filled by _ 3 If the zero point has an error it w...

Страница 13: ... 5 Command U CR 55h 0dh Response LF u1u2 CR LF H1H2H3 CR ETX Changes units of measure and return scale status with new units similar to pressing the UNIT key The new measure unit should be allowed to use as a C5 setting If the weight unit is lb oz U1U2 lb oz 6 Command L CR 4ch 0dh Response LF H1H2H3 CR ETX If HOLD function is enabled go to or exit from HOLD mode similar to pressing the HOLD key 7 ...

Страница 14: ...ways 1 always 1 5 always 1 always 1 always 1 6 always 0 always 1 always 0 7 parity Parity parity 7 Symbol Definitions 0 Zero point is over the setting range 0 Zero is below the setting range Ad Analog digital converter chip over max range Ad Analog digital converter chip below min range Weight signal is too large Weight signal is too small EEP E1 Config parameters incorrect no set no calibration o...

Страница 15: ...rload limit 2 The weight value cannot be displayed in the current unit of measure because it exceeds 6 digits 1 Reduce load on the scale until a weight value is displayed 2 Use a more appropriate unit of measure Weight reading below Under load limit 1 Perform zero calibration EEP E1 CONFIG parameters are not correctly set Re set CONFIG parameters as technical manual instructed EEP E2 USER paramete...