US-5011
.
10
Note: If the weight unit is lb: oz, U1U2= “lb oz” in above item
①②③
.
④
Normal weight is displayed, current weight unit is kg or lb, decimal point position is set by C4:
<LF><p>w1w2w3w4w5w6<dp>w7w8u1u2<CR><LF>H1H2H3<CR><ETX>
⑤
Normal weight is displayed, current weight unit is lb:oz,
<LF><p>w1w2w3w4w5w6lb<sp>w7w8<o><z><CR>H1H2H3<CR><ETX>
Or
<LF><p>w1w2w3w4w5lb<sp> w6w7<dp>w8oz<CR>H1H2H3<CR><ETX>
(2).
Command: S<CR> (53h 0dh)
Response: <LF> H1H2H3<CR><ETX>
(3).
Command: Z<CR> (5ah 0dh)
Response: <LF>H1H2H3<CR><ETX>
Zero function is activated, and then it returns to current scale status, similar to pressing the ZERO key. If
ZERO function cannot be activated, it will return to current scale status.
(4).
Command: T<CR> (54h 0dh)
Response: <LF> H1H2H3<CR><ETX>
TARE function is activated, and then returns scale status, similar to pressing the TARE key. If TARE
function cannot be activated, it will return to current scale status.
(5).
Command: U<CR> (55h 0dh)
Response: <LF>u1u2<CR><LF> H1H2H3<CR><ETX>
Changes units of measure and return scale status with new units, similar to pressing the UNIT key. The
new measure unit should be allowed to use as a C5 setting. If the weight unit is lb:oz, U1U2= “lb oz”
(6).
Command: L<CR> (4ch 0dh)
Response: <LF> H1H2H3<CR><ETX>
If HOLD function is enabled, go to or exit from HOLD mode, similar to pressing the HOLD key.
(7).
Command: X<CR> (58h 0dh)
Response: NONE
Power off the scale, similar to pressing and holding the ZERO/ON/OFF key for 4 seconds.
(8).
Command: all others
Response: <LF>? <CR><ETX>
Unrecognized command
e)
Additional Commands and Responses for Scale Base Application:
(1).
Command: F<CR> (46h 0dh) --- to restore factory calibration data
Response: <LF>OK H1H2H3<CR><ETX>
(2).
Command: O<CR> (4Fh 0dh) --- zero point calibration
Response: <LF>OK H1H2H3<CR><ETX> --- if zero calibration is sufficient
<LF> H1H2H3<CR><ETX> ---- if zero calibration resulted in an error
(3).
Command: H<CR> (48h 0dh) --- weight calibration
Response: <LF>OK H1H2H3<CR><ETX> --- if weight calibration is sufficient
<LF> H1H2H3<CR><ETX> --- if weight calibration resulted in an error