background image

 

Safety 

 

Ensure that all power sources are turned off when the machine is not in use. This encompasses electrical and 

pneumatic power. 
 

 

Read the manual for any special operational instructions for each piece of equipment. All USA Lab authored 

manuals are typically included with each device as well as posted online. 
 

 

Know how the equipment functions and understand the operating and halting processes.  

 

 

Wear the appropriate personal protective equipment for the task.  

 

 

When working on or around all equipment, avoid wearing loose clothing, jewelry, unrestrained long hair, loose 

ties, belts, scarves, or articles that may be caught in moving parts. Keep all extremities away from moving parts. 
Entanglement can cause death or severe injury.  
 

 

For new equipment, check input voltage and compare with the equipment voltage rating. DO NOT supply the 

incorrect power to any equipment for any reason whatsoever. Electrical specifications for your machine are 
printed on the machine tag. A properly grounded receptacle is required for safe operation regardless of voltage 
requirements.  
 

 

Use this equipment only for its intended purpose.  

 

 

Keep the equipment operating zone free of obstacles that could cause a person to trip or fall toward an operating 

machine. Keep fingers, hands, or any part of the body out of the machine and away from moving parts when the 
machine is operating. 
 

 

Any machine with moving parts and/or electrical components can be potentially dangerous no matter how many 

safety features it contains. Stay alert and think clearly while operating or servicing the equipment. Be aware of 
operations and personnel in your surroundings. Be attentive to indicator lights, warning lights, and/or operator 
interface screens displayed on the machine and know how to respond.  
 

 

Do not operate machinery if you are fatigued, emotionally distressed, or under the influence of drugs or alcohol.  

 

 

Know where the FIRST AID SAFETY STATION is located. 

 

 

Know where the FIRE EXTINGUISHING EQUIPMENT is located. 

 

 

Never sit or stand on the machine or on anything that might cause you to fall against the machine.  

 

 

Rotating and moving parts are dangerous. Keep clear of the operating area. Never put any foreign object into the 

operating area.  
 

 

Use proper lifting and transporting devices for heavy equipment. Some types of equipment can be extremely 

heavy. An appropriate lifting device should be used. 
 

 

Use caution when moving portable equipment. In some cases, the machinery can be heavy and/or may be top 

heavy. Portable equipment can gain momentum during transporting and must always be controlled.  

Содержание CRH Series

Страница 1: ...Recirculating Heater Manual Model CRH Series CRH 30L CRH 50L...

Страница 2: ...tion 3 1 LCD Display Operation 4 Installation 4 1 Thermal Transfer Fluids 4 2 Tubing Connections 4 3 Electrical Connection 4 4 Operating Area 4 5 Operating Instructions and Notes 5 Maintenance 5 1 Per...

Страница 3: ...cted on the best safety practices Most industry accidents are preventable through safety awareness Training It is the responsibility of the customer to ensure that all personnel who will be expected t...

Страница 4: ...cause a person to trip or fall toward an operating machine Keep fingers hands or any part of the body out of the machine and away from moving parts when the machine is operating Any machine with movin...

Страница 5: ...5 Symbols and Warnings Below are examples of commonly used symbols and what they mean Understand them and their potential consequences...

Страница 6: ...4 1 Apply a rag to wipe the parts clean after washing away stains do not use hard objects Do not use flammable corrosive or explosive substances on in or near the equipment Do not run the liquid pump...

Страница 7: ...to a heater chiller or allow you to use silicone oil with your equipment while keeping your glycol chiller Easy to use low position drain valve for convenient emptying 1 3 Technical Parameters Model C...

Страница 8: ...8 Section 2 Diagrams 2 1 Hose gaskets 2 2 Sight Glass...

Страница 9: ...9 2 3 Back Panel Only use water or water glycol in the cooling loop Install the valve on the heater outlet...

Страница 10: ...on Press set and change the set value to your desired temperature Press the pump button Press the heating button Disclaimer Parameters are not allowed to be modified by end users This is to protect th...

Страница 11: ...nickel plating Do Not Use Tap De Ionized or Reverse Osmosis Water as it can cause rusting of internal metal parts 4 2 Tubing Connections Diagram from Section 2 1 Install the ball valve onto the outle...

Страница 12: ...y operate the unit in a climate controlled facility Units returned that have signs of outside use Will automatically be determined to be improperly used and cared for When filling the unit Power on th...

Страница 13: ...Then enable the heating function When not operating the unit close the ball valve and shut off the power Operating steps Power on the unit Open the ball valve Run the pump Set the target temperature E...

Страница 14: ...similar substances during cleaning or else shock or other accidents will occur 5 1 Periodic Maintenance Fluid If using Silicone Oil degradation is not much of a concern Keep an eye on the tank level P...

Страница 15: ...15 Section 6 Technical Diagrams and Troubleshooting 6 1 Circuit Board Diagrams Main Board Display Board...

Страница 16: ...temperature probe C Replace main board Temperature display not working correctly or flashing A Incorrect power supply B Display cable loose or broken C Display board has failed A Supply the correct po...

Страница 17: ...n a heater or chiller the customer must only use distilled water Other forms of water will void the warranty Non continental United States USA Lab products are warrantied to be free of workmanship mec...

Страница 18: ...eemed defective by USA Lab In that case we will pay for return shipment and replacement shipment costs The item s must be returned in original packaging and in undamaged condition The item s must have...

Страница 19: ...not responsible for providing packaging material This manual and its contents are subject to change without notice USA Lab reserves the right of ultimate interpretation of the instruction manual Addit...