17
standard threads. When used with the handpiece, it is easy to damage and slip the teeth, and even cause
irreparable damage to the product. Please use the original tip.
17) When the operator uses a different series of tips, it needs to adjust the working mode accordingly to avoid
breaking the tip.
18)
Don’t
move or flip the device during use.
4.
Cleaning and sterilization
4.1. Handpiece
4.1.1. When using alcohol for sterilization, use 70%~80% (volume ratio) ethanol to soak a piece of clean dry gauze,
and then use this gauze to wipe the surface to be sterilized twice for 3 minutes. Dry or wipe the residual
alcohol with a clean, dry cloth.
4.1.2.
Sterilization
1)Please sterilize at 135℃ temperature and 0.22MPa for 15 minutes by high pressure sterilizer.
2)Pull out the handpiece and remove the tip after each use.
3)Wrap the handpiece with a disinfectant towel or bag.
4)After sterilization, the handpiece should be naturally cooled before use again to avoid scalding.
NOTICE
:
1)
Before sterilization, use compressed air to remove the cleaning fluid remaining in the handpiece.
2)
When sterilization, ensure to remove the tip from the handpiece, and avoid mixing it with other
instruments for sterilization.
3)
When sterilization, please always pay attention to whether the handpiece is damaged externally. It
is strictly forbidden to apply any protective oil on the surface of the handpiece.
4)
Sterilizable parts can be sterilized at least 250 times
5)
It is strictly forbidden to sterilize the handpiece in the following ways:
Put the handpiece in the solution and cook.
https://stomshop.pro