7.  Practical tips 

 
Run any new tubing in. New tubing is very stiff and requires a lot of torque from the motor. Once the tubing has 
been "softened" you will be able to run at much higher speeds without problems. We suggest 100 revolutions to 
run a tube in. 
 
Adjust the run current to the speed of rotor. If you want to run the pump very slowly, for example at the min 
speed of 50 steps/sec, reduce the Run Current to 500mA. Too much current causes the motor to shudder and 
the processor to overheat. In most applications you will not need a higher supply current than 1500mA. You can 
set a different Run Current for each protocol you run. 
 
When a stepper motor slips, a high pitch noise can be heard and the motor stalls. Any of the variants can cause 
a motor to slip. The optimal settings of  the acceleration, the initial speed and the final speed are all crucial for 
the best motor run. 
 
Setting a high speed with too high initial speed will cause the motor to slip. 
 
 
 
In very short rotations, for example 100 steps, the motor might not reach the requested speed. 
 

Содержание Stepper Control Board

Страница 1: ...nue London NW3 4AU England Co Reg No 4694961 VAT No 858 3837 74 Uno International Ltd Stepper Control Board Manual 1 21 Stepper Control Board Manual ISS 8 June 2013 USA Clark Solutions 10 Brent Drive...

Страница 2: ...2 21 Table of contents 1 General description 3 2 Required hardware 3 3 Connection between motor and controll board 3 4 Software installation 4 5 Hardware installation 4 6 Software explanation 5 7 Pra...

Страница 3: ...he control board you will require the following hardware 1 Power supply unit PSU 24V 2 5A 2 USB cable A to B 3 PC with Windows XP Vista or Windows 7 installed NET platform 2 0 minimum 3 Connection bet...

Страница 4: ...will find the setup exe file Using this file install the BoxerPumps Controller application onto your computer Plug the PSU plug into the controller board and make sure mains power to the power supply...

Страница 5: ...s this button The PCB has also a socket for a panic stop actuator 1 Motor configuration 1 Right click Motors and choose Configure A new dialogue box pops up 2 Press Add to add a new motor 3 Choose a s...

Страница 6: ...e and configure motors for other addresses on the board 2 Gravimetric pump calibration procedure 1 Calibrate the pump using the media you are intending to use the pump for 2 Use precision scales with...

Страница 7: ...pense side into a receiving vessel 7 Prime the pump Put both tube sides suction dispense into the reagent reservoir Right click your chosen to prime and choose Run Continuously Press the stop button a...

Страница 8: ...the motor 10 full revolution 2000 steps 11 Carry out 10 dispenses and calculate the mean value 12 Go back to Motor Configuration of the motor you are calibrating and choose Pump as description of the...

Страница 9: ...ated automatically such as l revolution if you chose Pump 4 Enter the value you established during the calibration process whether l revolution mm step or steps revolution and press OK 5 From now on w...

Страница 10: ...ify each function separately You can delete any of the functions by right click and choosing Delete You can specify Repeat Number of repeats Delay Period of delay between actions in seconds Backstep V...

Страница 11: ...following Right click in the chain pane and choose Add Repeat Specify the number of repeats drag the commands you wish to repeat onto the Repeat command Example You have written the following command...

Страница 12: ...uno Uno International Ltd Stepper Control Board Manual 12 21 Now drag the sub menu into a position in the protocol where you want these actions to take place...

Страница 13: ...g or from the File drop down menu choose Scheduling Choose the chain you would like to schedule Define the start time and the intervals in hours and minutes you would like the chain to run Press Add t...

Страница 14: ...e Run Current to 500mA Too much current causes the motor to shudder and the processor to overheat In most applications you will not need a higher supply current than 1500mA You can set a different Run...

Страница 15: ...run any number of different motor types sequentially You can not run the motors simultaneously 8 1 Daisy chaining cable In order to enable the software to talk to all pumps you need a Daisy Chain Cab...

Страница 16: ...computer Each pump has to have a power supply A daisy cable must connect the pumps together Here is a typical installation of 2 daisy chained pumps USB cable to the PC 24V DC 2 5A Power supply cable...

Страница 17: ...USB cable must be set on the controller software as no 1 The DIP switch setting on this pump must be as per following picture This is the master motor Picture shows setting for Master DIP switch addre...

Страница 18: ...erence is that now as part of each protocol you have to specify which of the motors should carry out the task Motor 1 Master Motor 2 Slave Motor 3 Slave 3 and so on 1 Configure a new motor and set it...

Страница 19: ...l Ltd Stepper Control Board Manual 19 21 A simple protocol made from 4 different motors could look like this You can add all the sub functions as done previously and a finished protocol for 4 motors c...

Страница 20: ...20 21 Once this is done you can add many chains in series Protocol 1 Protocol 2 Protocol 4 Protocol 2 Protocol 7 Protocol 3 Protocol 2 Protocol 4 Protocol 7 Protocol 3 Protocol 2 Protocol 2 Protocol 4...

Страница 21: ...r authorized by Uno International Ltd or 5 theft vandalism fire water or other peril Product may not be returned without proper authorisation from Uno International Ltd Cost of transportation removal...