Dealing With Errors 

When the Cap Track encounters an error with a specific tube, it will light the tube red. Upon finishing its cycle 

(or the “STOP” button being pressed) it will prompt for you to resolve the issues. This brings you to the error 

resolution screen, which asks you to address or at least acknowledge any errors from this run. 

 

 

A

: This box displays the current error being dealt with. You 

must perform some action on each error. 

B

: This area will change depending on the current error type 

being resolved. 

C

: This table contains a list of errors which still need to be 

dealt with. This is the order that they will be processed, as 

well as the current order on the ring from left to right 

(counter-clockwise). 

Error Types 

Barcode Read Errors

 are generated when the system failed to read the barcode on a given sample. To resolve 

this error, tap the text box labeled barcode and enter the barcode manually. Once entered, you can press the 

Resolve button. You have the option to skip this error, which will mean the tube will not be loaded into the rack, 

and can be run through the machine again. 

 

Cap Application Errors

 are generated when the system has failed to get a cap on the sample after repeated 

attempts. To resolve this error, manually place a cap on the sample, and press the resolve button. Otherwise, 

you have the option to skip this error, which will mean the tube will not be loaded into the rack, and can be run 

through the machine again. 

 

Tube Scanner Errors

 are generated when the sensor network mis-reads the height and cap state of a given 

sample. Ensure that the tube is clean, and nothing is stuck to the top of the sample that could interfere with the 

Cap Track’s sensors. Your only option is to acknowledge this error. If the error has occurred on several 

samples, contact Lab Improvements at [email protected]

 

Tube Seating Errors

 are generated when the system repeatedly fails to grip a given sample. This error can 

only be acknowledged. Ensure that nothing is trapped in the bottom of the ring nest in question. If the error has 

occurred on several samples, contact Lab Improvements at [email protected]

 

 

 

 

Содержание Lab Improvements Cap Track

Страница 1: ...Cap Track CT1 User Manual 1 800 423 2767 info universalmedicalinc com...

Страница 2: ...cation 3 Initial Setup Procedure 3 Operation 4 Dealing With Errors 5 Error Types 5 Sample Recall 6 Adding Racks 6 Expired Samples 7 Manager Password 7 Rack Expiration Time 7 Change Language 7 Change T...

Страница 3: ...z Certifications CSA UL Languages English French Spanish Features Compact benchtop size and weight allows for convenient placement within the lab Simple and intuitive 7 Touchscreen interface allows fo...

Страница 4: ...being depressed 5 Open the black lid and pour in 1 bag of Lab Improvements caps provided separately Close the lid 6 Press the power button located on the front of the machine 7 Allow 10 15 seconds fo...

Страница 5: ...ed with the rack 6 Wait for the machine to complete it s cycle The machine will inform you of any system errors that may occur during operation Should a fatal error occur contact Lab Improvements at s...

Страница 6: ...ion to skip this error which will mean the tube will not be loaded into the rack and can be run through the machine again Cap Application Errors are generated when the system has failed to get a cap o...

Страница 7: ...different samples F An image of the sample s cap will appear in this box Adding Racks The Add a Rack screen can be found under DATABASE and is used to add new racks to the system A The Rack s RFID ta...

Страница 8: ...nge Password Rack Expiration Time To change the cutoff time for Expired Racks go to the SETTINGS screen and tap on the RACK CUTOFF box This will prompt a manager password Once the password has been en...

Страница 9: ...es quality motors and electrical components which are selected for high reliability and should not need servicing 5 Remove accessories before moving All cables samples and caps must be removed before...

Отзывы: