1.4.2
Analog Input
Ground
Connections
To avoid errors caused by common mode voltages on analog inputs, follow the
recommended grounding guidelines in
below.
Figure 1-4. Recommended Ground Connections for Analog Inputs
Because all analog input channels in AI-201/202/207/208/225 layers are
isolated as a group, you can connect layer AGND to the ground of the signal
source and eliminate the resistors shown in
for floating differential
input signals.
Input
Configuration
Type of Input
Floating
Grounded
Typical Signal Sources:
Thermocouples
DC Voltage Sources
Instruments or sensors
with isolated outputs
Typical Signal Sources:
Instruments or sensors
with non-isolated outputs
Differential
Two resistors (10k <R< 100k) provide
return paths to ground for bias currents.
Single-Ended,
Ground
Referenced
+
_
Ret
AGnd
DNA-STP-37,
AI
nX
V
in
Sig
DNA-STP-AI-U
+
_
Ret
AGnd
DNA-STP-AI-U
AI
nX
V
in
Sig
+
_
Ret
AGnd
DNA-STP-AI-U
AI
nX
V
in
Sig
+
_
Ret
AGnd
DNA-STP-AI-U
AInX
V
in
Sig
grounds are at the same potential.
Add this connection to ensure that both
NOT RECOMMENDED