TECHNICAL SPECIFICATION

z

Model No. : UT-1401

PAGE : 2

FDA 21 CFR part 11 - GAMP-4 compatible.

LEDs. S

one

Microcontroller based indicator with inbuilt Temperature & DP measurement in single enclosure.
Incorporates highly stable built-in or DP sensor. (with different unit like mmwc/ pa/ kpa/ mmhg)
3 Digit, 0.56" bright LED display for displaying DP
Process parameter Hi and Low

tatus indication For

channel.

User selectable set limits for process value violation.
Inbuilt buzzer to audible process value violation.
Password protection for calibration and programming mode.
Software base Calibration .
MODBUS RTU multi drop RS-485 serial communication for PLC, SCADA, etc.
Programmable devise address, baud rate, (max no. of device is 1-256).
Remote programmable from master device.
User programmable offset correction.
Compact, All Stainless Steel or FLP enclosure with 35 mm depth for installation in prefab panels.
Brick and modular wall mounting and flush mounting.
FDA21 CFR part 11 compliance data logging software available
Ex. Flameproof model available for Hazardous applications.

Technical Specification

Flame

Proof

Material of constructions

Cable entery

Gas group

Enclose certification

Display glass

Degree protection

I IIA IIB as per IS 2148/1981

Enclosure is certifies by C.M.R.I. dhanbad

Clear glass thognet type A as per Is 2206

IP 65 as per IS 2147

2 x 3/4 “ ET

Aluminum alloy

Technical data

Nos. of input

Input

Indication

Range

Accuracy

Transmitter power supply

Comunication

Programmable

Parameters

Other

Supply voltage Power consumption

24 v DC @ 4 watts maximum

Operating ambient temperature and humidity

Temperatures input

1 nos .( DP)

- 0~50 °C

- Below 90%, non condensing

Mounting

Bricks and modules wall Flush panel mounting.

DP input

Integral Or Remote (external)

Integral Or Remote (external)

24V/5V DC @ 30 mA max, isolated from I/p & O/P.

DP

Channel-1 : 3 digit Seven segments 0.56” RED display.

3-LEDs for visual alarms, buzzer for audible alarm, LED for 485comm.

Display Resolution

Status indication

Read /write parameter

Read: Procces value ,decimal point. Write: Hi Low set point.

Programmable parameter

Modbus RTU RS- 485 Isolated communications.

Baud rate , Parity , Node address , stop bit.

Communication Protocol

Full range as per input selection.

Programming of parameters

Through 4 tactile keys (with password protections)

No. of slave ID

Total 1-256 unit.

Alarm logic

Full Scale HI OR Full scale Low

MOC

Full S S 316 plate and box , back side wall mounting M S covers.

Set point / High scale / Low scale

-10 to 10 mm / -25 to 25mm / -100-100 mm /-200 to 200 mm.....

mm wc : 0.1 / PA : 1 / Hpa :0.01

1 % on FS in all unit

DP

DP

Response time

1 sec

+91 94273 01436

Sales:

www.utplindia.in

Website:

Sure for service with hi-end

versal

nology

Uni

Tech

Sure for service with hi-end

versal

nology

Uni

Tech

E-mail :

[email protected]

Add.-78/1/Z/3, Makarpura GIDC, Makarpura, Vadodara Ph : +91 7046223333

Содержание UT-1401

Страница 1: ...94273 01436 Sales www utplindia in Website Sure for service with hi end versal nology Uni Tech Sure for service with hi end versal nology Uni Tech E mail sales utplindia in Add 78 1 Z 3 Makarpura GID...

Страница 2: ...1 is specially designed to indicate the Room pressure between two different areas The large 4 digit 0 4 high LED display is easy to read even from a distance Two alarms can be selected for above or be...

Страница 3: ...N z z Press to change Set point 1 in Normal Mode and increase selected digit or value in Programming Mode Press to change digit selection or selected value in Programming Mode Press to change Set poin...

Страница 4: ...2206 IP 65 as per IS 2147 2 x 3 4 ET Aluminum alloy Technical data Nos of input Input Indication Range Accuracy Transmitter power supply Comunication Programmable Parameters Other Supply voltage Power...

Страница 5: ...WIRING CONSIDERATION SHIELDED TWISTED PAIR WIRE IS USED FOR ALL ANALOG I P PROCESS SENSORS RTD THERMOCOUPLE DC mV mA VOLT DC LINEAR INPUT AND COMMUNICATION FAILURE TO COMPLY WITH THESE INSTRUCTION MAY...

Страница 6: ...e PROGRAMMING FLOW CHART z z 0000 91 94273 01436 Sales www utplindia in Website Sure for service with hi end versal nology Uni Tech Sure for service with hi end versal nology Uni Tech E mail sales utp...

Страница 7: ...Action TSP2 L oN L oN Change the type by or key One decimal 123 4 0001 Two decimal 12 34 0002 Three decimal 1 234 0003 Active in High H oN Active in Low L oN Active same as above Pressure Unit PASCAL...

Страница 8: ...point Set Point 1 SP1 0300 Increase digit by key decrease by key Activation value for Relay 1 Hys1 0002 Increase digit by key decrease by key Hysteresis 1 Value Difference between Activation and De a...

Страница 9: ...on Range Default Value ofst 0000 Offset to Process Value Scroll the digit by key increase by key This value Increase OR Decrease the Display PV Hys1 0002 Increase digit by key decrease by key Hysteres...

Страница 10: ...ad continuity All LED s and display not light No power to controller Control circuit failure Check power lines connection Check Control Supply Voltages Replace keypads OPEn ERROR MESSAGE AND TROUBLESH...

Страница 11: ...4 SETPOINT 2 40004 READ WRITE DEFAULT COMMUNICATION DETAILS z z BAUD RATE PARITY DATABIT NODE ADDRESS STOP BIT 9600 NONE 8 1 1 91 94273 01436 Sales www utplindia in Website Sure for service with hi en...