5.FUNCTION SETTINGS

57

5-4. Analog Filter

This function stabilizes the display by removing low frequency noise from the load cell signal as
well as averaging analog signals. 
Selectable frequencies are: 2, 4, 6 or 8Hz. A higher frequency will make a more quick response but
sometimes read a noise in.
Select the most suitable value for the weighing.

Operation

1) Select setting mode 1.

2) Select cutoff frequency. (8-Analog Filter)

5-5. Filter in Stable Condition

When indicated value is stable, this function automatically insert the Digital Filter for restraining
the instability. Select whether inserts it or not.
Regarding the definition of stability, refer to "5-6.Motion Detection (MD)" on P.58.

Operation

1) Select setting mode 4.

2) Select whether or not to insert the filter. (7-External Function Selection)

F

CNG/ENT

CNG/ENT

1

UPPER

Analog Filter

3 : 8Hz
2 : 6Hz
1 : 4Hz
0 : 2Hz

8

UNDER

CNG/ENT

CNG/ENT

Load cell

Analog

Filter

A/D

Digital

Filter

Digital

Filter

F701-C

OFF

Stab OFF or filter is not 

Setting counts

Stable

("5-3.Digital Filter" 

on P.56)

inserted in stable condition.

(128 Count fixed)

Setting frequency

("5-4.Analog Filter" 

on P.57)

converter

F

CNG/ENT

CNG/ENT

4

SP1

Filter in stable condition

1 : Not insert
0 : Insert (128 times)

ZERO

7

OVER

CNG/ENT

CNG/ENT

Содержание F701-C

Страница 1: ...F701 C WEIGHING CONTROLLER 01 Apr 2015 Rev 1 16 OPERATION MANUAL ...

Страница 2: ...ntroller used safely notes we would like you to surely follow divide into and and are indicated by the following documents Notes indicated here are the serious contents related to safely Please use F701 C after understanding the contents well For safety reasons please read the following safety precautions thoroughly WARNING CAUTION WARNING This sign forewarns the presence of hazards that could res...

Страница 3: ...he attached AC cable is designed for domestic use in Japan and its rating is 125V AC 7A For use at voltages exceeding the rating and for overseas use have a separate AC cable prepared Before performing the following make sure that no power is applied Attachment detachment of connectors of options etc Wiring connection of cables to the signal input output terminals Connection of the earth cable For...

Страница 4: ...1 C as it is incorporated in a control panel etc Do not install in the following environments Locations where temperature or humidity exceeds specifications Locations subjected to drastic temperature fluctuations or icing and condensing Outdoors or locations above 2 000m Locations exposed to direct sunlight Locations subject to dust accumulation Locations with poor ventilation Locations with a lot...

Страница 5: ...avy contamination wipe off the contamination with a cloth after dipping it into a diluted neutral detergent and wringing it well and then wipe with a soft dry cloth When the F701 C is shipped spacers made of corrugated cardboard are used as cushioning materials Though it is factory designed so that shocks can sufficiently be absorbed breakage may result if shocks are applied when the spacers are r...

Страница 6: ...the Connector 15 2 5 3 Equivalent Circuit Input 16 2 5 4 Equivalent Circuit Output 16 3 METHODS OF SETTING 17 3 1 Setting Procedure 17 3 1 1 Method of Selecting a Setting Mode 17 3 1 2 Method of Entering a Setting Value 19 3 2 Setting Mode 23 3 2 1 Setting Mode 0 24 3 2 2 Setting Mode 1 26 3 2 3 Setting Mode 2 28 3 2 4 Setting Mode 3 31 3 2 5 Setting Mode 4 33 3 2 6 Setting Mode 5 36 3 2 7 Setting...

Страница 7: ...re Subtraction 63 5 12 One touch Tare Subtraction Reset 64 5 13 Digital Tare Subtraction 65 5 14 Restriction of the Tare Subtraction Function 66 5 15 Switching Gross Weight Net Weight Display 67 5 16 Reversing Symbol at Discharging Control 68 5 17 Invalidation of Function Keys 68 6 DISCHARGING CONTROL MODE 69 6 1 Feeding Weighing and Discharging Weighing 69 6 1 1 Feeding Weighing 69 6 1 2 Discharg...

Страница 8: ... STATISTICAL AND ACCUMULATION FUNCTION SETTING AND OPERATION 102 7 1 Data Display 103 7 2 Accumulation Command 103 7 3 Accumulation Clear 105 7 4 Auto Accumulation Command 107 8 SYSTEM MODE 109 8 1 LOCK soft 109 8 2 Password 109 8 3 Self Check and Memory Clear 110 9 EXTERNAL INPUT OUTPUT SIGNALS CONTROL CONNECTOR 111 9 1 Control Connector Pin Assignment 111 9 2 Equivalent Circuit Input 112 9 3 Equ...

Страница 9: ...mission Format 128 10 2 6 Communication Format 129 10 3 BCD Parallel Data Output Interface 137 10 3 1 Control Connector Pin Assignment 137 10 3 2 Equivalent Circuit Input 137 10 3 3 Equivalent Circuit Output 138 10 3 4 BCD Data Output 138 10 3 5 Polarity Output MINUS 138 10 3 6 Over Status Output OVER 139 10 3 7 Print Command Output P C 139 10 3 8 Data Strobe STROBE 139 10 3 9 BCD Data Update Rate...

Страница 10: ... SPECIFICATIONS 170 17 1 Analog Section 170 17 2 Display Section 171 17 3 Setting Section 171 17 4 External Input Output 172 17 5 Interface 172 17 6 General Performance 174 17 7 Attachment 175 18 THE LIST OF INITIAL SETTING VALUE 176 18 1 Setting Mode 0 176 18 2 Setting Mode 1 176 18 3 Setting Mode 2 177 18 4 Setting Mode 3 177 18 5 Setting Mode 4 177 18 6 Setting Mode 5 178 18 7 Setting Mode 8 17...

Страница 11: ...CONTENTS X M E M O ...

Страница 12: ... display shows error and weight value alternately 2 Over scale and error display Over scale sequence errors and calibration errors are displayed Please refer to 11 OVER SCALE ERROR on P 157 3 Setting value display Various final discharge setting values and setting values for adjustment such as Final and Set Point 2 are displayed Status display Main display Setting keys Unit indication Function key...

Страница 13: ...tion on P 56 1 Accumulation Count Accumulation Value 2 Latest Accumulation Data Accumulation Value 3 Accumulation Count Latest Accumulation Data Final 4 Final Over Under 5 Near Zero Upper Limit Lower Limit 6 None Accumulation Count Accumulation Value Latest Accumulation Data Accumulation Value Accumulation Count Final Latest Accumulation Data Final Under Over Near Zero Lower Limit Upper Limit ...

Страница 14: ...r zero value set in the setting mode 0 3 Upper Limit Displays the upper limit value set in the setting mode 0 1 Lower Limit Displays the lower limit value set in the setting mode 0 2 Unit indication The unit can select from six types t kg g N lb and None Status display SP1 Lights when the Set Point 1 signal is ON SP2 Lights when the Set Point 2 signal is ON SP3 Lights when the Set Point 3 signal i...

Страница 15: ...re weight is displayed and flashes When the Tare weight display with the key is valid under Restriction on the Tare Subtraction Function in setting mode 4 To go back to the weight display press the key again Numerical keys to make settings Function key to switch the setting mode Change Enter key to confirm setting items and setting values Function keys Key to perform one touch Tare subtraction on ...

Страница 16: ...ss Net In setting mode 0 Pressing on Gross weight display when is OFF switches to Net weight display and pressing on Net weight display when is ON switches to Gross weight display However the display cannot be switched with this key if the switching Gross weight Net weight display is set to external input mode under External Function Selection in setting mode 4 Use as an ESC key for setting operat...

Страница 17: ...conducted It may be better to remove depending on the environment of the installation location Control connector This is a connector to input external signals and output control signals The Input Output circuit and internal circuit are photocoupler insulated electrically The applicable connector is the following accessory manufactured by Fujitsu Component or an equivalent Connector FCN 361J024 AU ...

Страница 18: ...h as printer remote display or data converter LOCK switch LOCK switch for avoiding changes of setting value it prohibits to change setting value while the switch is ON Load cell input terminal block This is a terminal block to connect load cell s The applicable terminal block is Osada manufactured ETB42 07P Option space One of the following options can be mounted BCD parallel data output interface...

Страница 19: ...V and the maximum current is 120mA to which up to four 350Ω load cells can be connected in parallel Load cell terminal block pin assignments Pin No Signal 6 wire Signal 4 wire 1 SIG SIG 2 SIG SIG 3 EXC EXC Connect 3 to 4 4 S 5 EXC EXC Connect 5 to 6 6 S 7 SHIELD SHIELD Load cell terminal block ...

Страница 20: ...iver etc 5 Insert the wire into the hole so as not to loosen the end 6 Tighten the screw with the screwdriver 7 Lightly pull the wire to check that it is clamped securely Connectable wires are 0 21 3 31mm2 AWG12 24 Recommendable tightening torque is 0 5Nm 8 Insert the wire connected plug into the F701 C body and tighten the screws two 5mm Hole Turn counterclockwise Turn clockwise Turn clockwise Sc...

Страница 21: ... may occur because excessive voltage is applied to the load cell For connection use the accessory jumper lines IN OUT IN EXC S S EXC SIG SIG SHIELD OUT 3 4 6 5 1 2 7 Load cell IN OUT IN EXC S S EXC SIG SIG SHIELD OUT 3 4 6 5 1 2 7 Load cell These jumpers MUST be connected The load cell excitation voltage of the F701 C is 10V Heating or breakage may occur unless the load cells maximum excitation vo...

Страница 22: ...l or data sheet and check signal names and colors in order to connect the cables correctly R R R R R R EXC SIG SIG EXC EXC SIG EXC SIG SHIELD S S When connecting several load cells in parallel load cell capacity should be higher than expected load to compensate for mechanical shock or eccentric loading R R The group of n parallel load cells viewed from this device side can be regarded as a unit lo...

Страница 23: ... tighten the screws 2 3 Connection of the Guard Ground The grounding terminal is for prevention of electric shocks Use an approx 0 75mm2 thick wire and be sure to ground block cover Within 6mm AC power cord Frame ground Connect with no power applied because it may cause an electric shock The attached AC cable is designed for domestic use in Japan and its rating is 125V AC 7A For use at voltages ex...

Страница 24: ...nnected Twist the bare wire to fit the terminal hole 2 Insert the supplied screwdriver into the upper hole and lift upward 3 Insert the twisted wires into the lower hole 4 Pull the screwdriver out from the upper hole 5 Make sure cable is clamped securely and does not come out with a slight tug 5 to 6mm Cable can be from 24 14AWG 0 2 2 5mm2 It is not necessary to solder the cable wires or to fix a ...

Страница 25: ...ase refer to 6 11 Weight Error Sequence Error on P 101 6 8 Complete Signal Output Mode Complete Output Time Judging Time Comparison Inhibit Time Output Selection 2 on P 93 9 5 7 Accumulation Error on P 119 SP1 turns ON when Weight value Final SP1 SP2 turns ON when Weight value Final SP2 SP3 turns ON when Weight value Final CPS A1 COM B1 COM A2 In G N B2 In Input Selection 1 1 A3 In D Z ON B3 In In...

Страница 26: ...with the other case and fit the cases 3 Tighten the M2 8 pan head machine screws two Tighten the M2 10 pan head machine screws two Be aware that washers should be set to the M2 10 pan head machine screws two Connector Pan head machine screw M2 8 short two Nut M2 four Pan head machine screw M2 10 long two Washer two Case two Screw two ...

Страница 27: ... the signal input circuit The external element is required to withstand Ic 10mA Leakage from the external element is required to be 100μA or below switch IC 6mA TTL open collector ON when IN is HI Spark killer Load Spark killer DC Power Varistor Load AC power Relay Vext Output data Tr 0 OFF 1 ON F701 C Inside Vcc 12V COM Vceo 30V max Ic 120mA max Use external power source up to DC30V for driving r...

Страница 28: ...mode is described as follows Example For selecting setting mode 3 This operation can be performed by the following procedure 1 Press the key when the weight value is displayed 2 Press the key F CNG ENT CNG ENT 3 NEAR Z kg Normally the state is in setting mode 0 The mode number is not displayed for mode 0 Setting mode number Present weight value F kg Setting mode number The display changes CNG ENT ...

Страница 29: ...ode number The input number blinks CNG ENT Present weight value The selected setting mode number Setting mode number and weight value are displayed Setting mode display kg By pressing the key when the setting mode number is displayed you can go back to normal display Setting mode 0 GROSS NET ESC ...

Страница 30: ...ing value is input press the key to enter the setting value 1 UPPER CNG ENT CNG ENT 1 UPPER kg Present setting value The setting mode number setting Setting mode number item number and present setting Setting item number value are displayed CNG ENT kg Present setting value The highest digit of the setting Setting mode number value blinks Setting item number 0 5 SP2 0 0 0 kg Setting value Every tim...

Страница 31: ...g mode number item number and present setting Setting item number value are displayed kg CNG ENT Present setting value The highest digit of the setting Setting mode number value blinks Setting item number kg 0 0 3 NEAR Z 0 F Setting value Every time a number is pressed the blinking Setting mode number digit moves to the lower one Setting item number Since the highest digit starts blinking again af...

Страница 32: ...til the digit you want to set blinks 7 OVER CNG ENT CNG ENT ZERO ZERO ZERO 7 OVER Present setting value Setting mode number Setting item number The setting mode number setting item number and present setting value are displayed kg CNG ENT Present setting value Setting mode number Setting item number The highest digit of the setting value blinks kg ZERO ZERO ZERO ZERO kg Not defined Setting mode nu...

Страница 33: ...ne Setting mode number Setting item number Since the blinking digit moves every time the key is pressed setting can be redone again and again ZERO kg CNG ENT kg Present weight value Setting mode number The display returns to the setting mode display By pressing the key when the setting item number is displayed while changing a setting value after selecting an item you can exit the item The display...

Страница 34: ...Range P 30 Gravitational Acceleration acceleration P 32 1 UPPER 2 LOWER 3 NEAR Z 4 SP1 5 SP2 6 CPS 7 OVER 8 UNDER 9 FINAL ZERO When each setting mode is selected the 1 9 keys function as setting item selection keys Setting mode 4 Setting mode 5 Setting mode 8 Setting mode 9 D A Output Mode P 33 Input Selection P 36 Average Weight P 38 Span Calibration P 53 D A Zero Output Weight P 33 Output Select...

Страница 35: ...it For details please refer to 6 6 on P 90 Near Zero For details please refer to 6 6 on P 90 Set Point 1 For details please refer to 6 4 on P 86 Set Point 2 For details please refer to 6 4 on P 86 Compensation For details please refer to 6 4 on P 86 0 99999 1 UPPER 0 99999 2 LOWER 0 99999 3 NEAR Z 0 99999 4 SP1 0 99999 5 SP2 9999 9999 If negative setting value display the sign 6 CPS ...

Страница 36: ... METHODS OF SETTING 25 Over For details please refer to 6 4 on P 86 Under For details please refer to 6 4 on P 86 Final For details please refer to 6 4 on P 86 0 999 7 OVER 0 999 8 UNDER 0 99999 9 FINAL ...

Страница 37: ...8 on P 93 Complete Output Time For details please refer to 6 8 on P 93 Adjust Feeding Time Effective when selecting sequence mode For details please refer to 6 9 on P 96 Auto Zero Times Effective when selecting sequence mode For details please refer to 6 9 on P 96 Judging Times Effective when selecting sequence mode For details please refer to 6 9 on P 96 sec 0 00 9 99 1 UPPER sec 0 00 9 99 2 LOWE...

Страница 38: ...uto Free Fall Compensation Regulation For details please refer to 6 3 on P 83 Analog Filter For details please refer to 5 4 on P 57 Tare Weight For details please refer to 5 13 on P 65 0 99999 7 OVER 0 3 8 UNDER 0 99999 9 FINAL ...

Страница 39: ... Not defined Complete signal output mode 2 ON for the complete output time from when the judging timer has expired or from when the weight value becomes stable after the SP3 signal turns ON 1 ON for the complete output time from when the weight value becomes stable after the judging timer has expired 0 ON for the complete output time from when the judging timer has expired Over Under comparison mo...

Страница 40: ...se refer to 6 3 on P 83 For details please refer to 6 3 on P 83 3 NEAR Z 1 Negative Compensation Effect 0 Negative Compensation Inhibit For details please refer to 6 5 on P 89 1 Sequence mode 0 Simple comparison mode Sequence control 1 At start time weight value check ON 0 OFF 1 Adjust feeding ON 0 OFF Sequence control 1 At start time Near Zero check ON 0 OFF For details please refer to 6 9 on P 9...

Страница 41: ...8 Zero Tracking Period For details please refer to 5 7 on P 60 Zero Tracking Range For details please refer to 5 7 on P 60 TARE key 1 Valid 0 Invalid GROSS NET key 1 Valid 0 Invalid TARE RESET key 1 Valid 0 Invalid ZERO key 1 Valid 0 Invalid 5 SP2 Digital Filter 0 256 6 CPS Period sec Range Set the stability detection parameter 0 0 9 9 0 99 7 OVER Period sec 0 0 9 9 8 UNDER Range 0 9999 9 FINAL ...

Страница 42: ...4 6 on P 48 Capacity For details please refer to 4 4 4 on P 47 Min Scale Division For details please refer to 4 4 5 on P 47 Net Over For details please refer to 6 10 on P 100 Gross Over For details please refer to 6 10 on P 100 DZ Regulation Value For details please refer to 5 10 on P 62 0 99999 1 UPPER 0 99999 2 LOWER 1 100 3 NEAR Z 0 99999 4 SP1 0 99999 5 SP2 0 9999 6 CPS ...

Страница 43: ...ease refer to 4 4 8 on P 50 7 OVER Subdisplay selection 5 Near Zero Upper Limit Lower Limit 4 Final Over Under 3 Accumulation Count Latest Accumulation Data Final 2 Latest Accumulation Data Accumulation Value 1 Accumulation Count Accumulation Value 0 None For details please refer to 5 2 on P 56 Unit 5 N 4 lb 3 kg 2 g 1 t 0 None For details please refer to 4 4 2 on P 46 3 2 1 0 When using the F701 ...

Страница 44: ...ull Scale For details please refer to 10 4 3 on P 143 RS 485 I F For details please refer to 10 5 3 on P 147 Test mode 2 20mA fixed output 1 4mA fixed output 0 Tied to the weight value Output mode 1 Net weight 0 Gross weight 1 UPPER 0 99999 2 LOWER 0 99999 3 NEAR Z Baud rate 5 38400bps 4 19200bps 3 9600bps 2 4800bps 1 2400bps 0 1200bps Terminator 1 CR LF 0 CR Character length 1 8bit 0 7bit Stop bi...

Страница 45: ...8400bps 4 19200bps 3 9600bps 2 4800bps 1 2400bps 0 1200bps Communication mode 7 Communication mode 6 6 Communication mode 5 5 Communication mode 4 4 Communication mode 3 3 Communication mode 2 2 Communication mode 1 1 Communication mode 0 CR LF 0 Communication mode 0 CR Character length 1 8bit 0 7bit Stop bit 1 2bit 0 1bit Parity bit 2 Even 1 Odd 0 None 6 CPS ...

Страница 46: ... Switching Gross weight Net weight 1 External input mode 0 Internal key mode 1 Checker mode 0 Stable mode Motion Detection mode For details please refer to 5 6 on P 58 For details please refer to 5 15 on P 67 For details please refer to 5 16 on P 68 For details please refer to 5 5 on P 57 For details please refer to 10 3 9 on P 140 display 7 OVER 8 UNDER LOCK1 1 ON 0 OFF LOCK2 1 ON 0 OFF One touch...

Страница 47: ...d 4 Stop 3 Start 2 Feed Discharge 1 Judge 0 HOLD 6 Accumulation Clear 5 Accumulation Command 4 Stop 3 Start 2 Feed Discharge 1 Judge 0 HOLD 6 Accumulation Clear 5 Accumulation Command 4 Stop 3 Start 2 Feed Discharge 1 Judge 0 HOLD Output selection 3 B11 pin 1 Accumulation Error 0 RUN Output selection 1 B9 pin 2 Weight error or sequence error 1 Sequence error 0 Weight error Output selection 2 B10 p...

Страница 48: ...ber of occupied station Transmission speed CC Link I F Station No Transmission speed 4 10M 3 5M 2 2 5M 1 625k 0 156k Not defined Number of occupied station 2 Occupies 4 stations 1 Occupies 2 stations 0 Occupies 1 station 3 NEAR Z 01 64 4 SP1 ...

Страница 49: ... Value Min Value General Standard Deviation Sample Standard Deviation Accumulation Count Latest Accumulation Data Max Min Average Weight Max Value Min Value General Standard Deviation Sample Standard Deviation Accumulation Count n Latest Accumulation Data 0 99999 1 UPPER 0 99999 2 LOWER 0 99999 3 NEAR Z 0 99999 4 SP1 0 99999 5 SP2 0 10000 6 CPS 0 99999 7 OVER ...

Страница 50: ... CC Link interface σn Σ n i 1 n χi χ 2 Use all the data of the finite population and find the standard deviation of the population σn 1 Σ n i 1 n 1 χi χ 2 Use the sample data among the population and the standard deviation of the population Data taking conditions When auto free fall compensation is OFF Data is taken when judgment is made When auto free fall compensation is ON Data is taken when fr...

Страница 51: ...0 040 0 030 0 012 0 015 4 80 240 20 080 20 060 20 080 20 040 0 040 0 016 0 018 5 100 260 20 020 20 052 20 080 20 020 0 060 0 021 0 024 6 120 260 20 000 20 043 20 080 20 000 0 080 0 027 0 030 7 140 270 20 010 20 039 20 080 20 000 0 080 0 028 0 030 8 160 250 19 980 20 031 20 080 19 980 0 100 0 033 0 035 9 180 360 20 110 20 040 20 110 19 980 0 130 0 039 0 042 10 200 370 20 010 20 037 20 110 19 980 0 ...

Страница 52: ... F701 C accurately displays 100 00kg when an actual load or weight of 100kg is applied to the load cell balance section of the weighing apparatus to which the F701 C is connected This operation is called Span Calibration Connect F701 C to the load cell The display is incorrect 100kg 100kg After calibration The F701 C and load cell function as a weighing device ...

Страница 53: ... scale Register the minimum unit scale division of the scale Register the value of load balance weight for load cell Calibrate the initial zero point Place the load balance weight on the load cell and register the span value If it is necessary re calibrate initial zero point Disable changing setting value related to calibration Turn the calibration LOCK ON to prohibit the calibration for avoiding ...

Страница 54: ...OCK which inhibits the calibration The displayed unit is set Register the position of decimal point Register the maximum capacity of the scale If the registered value exceeds by 9 scale divisions display shows over scale Register the minimum unit scale division of the scale Set the weight value corresponding to the output value to be entered in 1 Register the rated output value mV V of load cell I...

Страница 55: ...of the load cell set the Capacity to the same value as the rated value of the load cell When connecting several load cells in parallel it is possible to occur some differences between input and output value due to voltage drop caused by connection or material of lines In this case register actual input value to perform accurate calibration Request ...

Страница 56: ...el Release both of locks when the calibration is performed Operation 1 LOCK OFF on the rear panel 2 Select setting mode 4 3 Set LOCK2 to OFF 8 Setting Value LOCK LOCK is released through above procedure After the calibration is finished LOCK ON to protect the calibration value LOCK OFF 4 SP1 F CNG ENT CNG ENT LOCK2 1 ON 0 OFF 8 UNDER CNG ENT CNG ENT ZERO Concerning LOCK and setting values to be pr...

Страница 57: ...ce should be selected from 0 0 0 0 00 or 0 000 This selection results in a common decimal place Operation 1 Select setting mode 3 2 Select a desired decimal place 7 Function Selection F CNG ENT CNG ENT 3 NEAR Z Unit 5 N 4 lb 3 kg 2 g 1 t 0 None ZERO 7 OVER CNG ENT ZERO ZERO CNG ENT ZERO F CNG ENT CNG ENT 3 NEAR Z Decimal place ZERO 7 OVER CNG ENT CNG ENT ZERO 3 2 1 0 In the F701 C the decimal plac...

Страница 58: ... Select setting mode 3 2 Input the Capacity 2 Capacity 4 4 5 Minimum Scale Division Register the minimum unit scale division of the scale Input range 1 100 Operation 1 Select setting mode 3 2 Input the Min Scale Division 3 Min Scale Division F CNG ENT CNG ENT 3 NEAR Z Capacity 0 99999 2 LOWER CNG ENT CNG ENT F CNG ENT CNG ENT 3 NEAR Z Min Scale Division 1 100 CNG ENT CNG ENT 3 NEAR Z ...

Страница 59: ...ng is not needed Find the area number 01 16 of the area where actual load calibration is carried out from the Gravitational Acceleration correction table on the next page set the number and then perform actual load calibration Next find the actual installation area from the table and input the area number Now the difference from the calibration location in gravitational acceleration is corrected A...

Страница 60: ...janeiro 9 788m s2 Bangkok 9 783m s2 Rome 9 803m s2 Birmingham 9 813m s2 San Francisco 9 800m s2 Brusseles 9 811m s2 Singapore 9 781m s2 Buenos Aires 9 797m s2 Stockholm 9 818m s2 Calcutta 9 788m s2 Sydney 9 797m s2 Capetown 9 796m s2 Taichung 9 789m s2 Chicago 9 803m s2 Tainan 9 788m s2 Copenhagen 9 815m s2 Taipei 9 790m s2 Cyprus 9 797m s2 Tokyo 9 798m s2 Djakarta 9 781m s2 Vancouver BC 9 809m s2...

Страница 61: ...vision selects ON OFF Operation 1 Select setting mode 3 2 Select 1 4 scale division display ON OFF 7 Function Selection ZERO F CNG ENT CNG ENT 3 NEAR Z 1 4 scale division display 1 ON 0 OFF ZERO 7 OVER CNG ENT CNG ENT ZERO ZERO n n 1 1 4 scale Min Scale Division ON 1scale divisio ZERO ON ZERO When the 1 4 scale division setting is OFF lamp only works at the zero point ZERO 1 1 4 scale 0 1 ON ZERO ...

Страница 62: ...eted If a calibration error is displayed redo Zero Calibration corresponding to the following error messages Please refer to 11 OVER SCALE ERROR on P 157 Initial dead load is above zero adjustment range Remove any excess load from load cell or scale If is still displayed connect a resistor between EXC and SIG load cell connections This should shift the zero point STAB 9 FINAL F CNG ENT CNG ENT kg ...

Страница 63: ... a temperature coefficient of 50ppm or more recommended value of about 5ppm Resistance Strain Calculated value Approx value μ STRAIN mV V 875 kΩ 437 kΩ 291 kΩ 219 kΩ 175 kΩ 146 kΩ 125 kΩ 109 kΩ 97 kΩ 87 3 kΩ 79 4 kΩ 72 7 kΩ 67 1 kΩ 62 3 kΩ 58 2 kΩ 54 5 kΩ 51 3 kΩ 48 4 kΩ 45 9 kΩ 43 6 kΩ 41 5 kΩ 39 6 kΩ 37 9 kΩ 36 3 kΩ 34 8 kΩ 866 kΩ 442 kΩ 294 kΩ 221 kΩ 174 kΩ 147 kΩ 124 kΩ 110 kΩ 97 6 kΩ 86 6 kΩ ...

Страница 64: ... that is on Correct calibration can not be completed if signal is unstable Operation 1 Select setting mode 9 Calibration mode 2 Perform Span Calibration 1 Span Calibration 3 When the weight value display becomes equal to the Balance Weight Value Span Calibration is completed If any error message appears refer to 11 OVER SCALE ERROR on P 157 STAB F CNG ENT CNG ENT 9 FINAL In span calibration displa...

Страница 65: ...the output value to be entered in 4 1 Balance Weight Value 3 Select setting mode 9 4 Input the Output value of the load cell 2 Equivalent Calibration If any error message appears refer to 11 OVER SCALE ERROR on P 157 If Zero Calibration again appears redo Zero Calibration and then perform Equivalent Calibration again is displayed when calibrating by resolution that exceeds 1 10000 F CNG ENT CNG EN...

Страница 66: ...CPU processing speed are not changed The available display frequencies are 25 13 6 or 3 times sec 25 times sec is recommended for normal operation If the display flickers select a lower frequency Operation 1 Select setting mode 3 2 Select Display frequency 7 Function Selection F CNG ENT CNG ENT 3 NEAR Z Display frequency 3 25 times sec 2 13 times sec 1 6 times sec 0 3 times sec ZERO 7 OVER CNG ENT...

Страница 67: ...value by calculating the average frequency of the data converted from analog to digital The frequency of the moving average selected from OFF 0 256 times A higher frequency will make a more stable display with slower response A lower frequency will have quicker response but more unstable display Select the most suitable value for the weighing Operation 1 Select setting mode 2 2 Input the frequency...

Страница 68: ... the Digital Filter for restraining the instability Select whether inserts it or not Regarding the definition of stability refer to 5 6 Motion Detection MD on P 58 Operation 1 Select setting mode 4 2 Select whether or not to insert the filter 7 External Function Selection F CNG ENT CNG ENT 1 UPPER Analog Filter 3 8Hz 2 6Hz 1 4Hz 0 2Hz 8 UNDER CNG ENT CNG ENT Load cell Analog Filter A D Digital Fil...

Страница 69: ...e compared with a set range at each A D conversion D1 represents the difference between current weight value and the value 1 second previous If one of them exceeds the range the Stab turns OFF at once Checker mode D1 D3 shown below are compared with a set range at each A D conversion If one of them exceeds the range the Stab turns OFF at once D1 represents the difference between current weight val...

Страница 70: ...ng the value setting value minimum scale division with the range of weight change Operation Input range 0 99 1 Select setting mode 2 2 Input the Period and Range 7 Motion Detection Analog Filter A D Digital Filter Digital Filter Comparator OFF Stab OFF Cutoff frequency selectable 2 4 6 8 Hz Counts selectable 0 256 Times Stab ON 128 Times fixed Set under Analog Filter in setting mode 1 and Digital ...

Страница 71: ...usts the zero point every set period when the shifting of the zero point is within the set point The Tracking Period must be set between 0 0 and 9 9 seconds The Tracking Range digit must be set from 0 9999 using a 1 4 scale division as a unit 0002 0 5 divisions 0012 3 divisions Zero Tracking does not work when the period is set 0 0 or the range is set to 00 Boundary of zero track Tracking period W...

Страница 72: ...g A3pin D Z ON and COM on the control connector Gross weight is immediately zeroed Net weight changes according to the following expression Net weight Gross weight Tare weight In a case where 0 is not indicated when performing digital zero operation or when flashes the following causes are considered ZALM ZERO CNG ENT Cause Measure Digital zero operation was performed when the DZ Regulation Value ...

Страница 73: ...range of zero point adjustment a gap from the registered Zero Calibration value by digital zero or zero tracking If the digital zero operation is performed or zero tracking operates where the DZ Regulation Value is exceeded flashes to give a warning of the problem Input range 0 9999 Operation 1 Select setting mode 3 2 Input the DZ Regulation Value 6 DZ Regulation Value ZALM ZALM F CNG ENT CNG ENT ...

Страница 74: ... the control connector Net weight is immediately zeroed and lights If Net weight is not zeroed by performing the operation of Tare subtraction the following causes are presumable TARE STAB TARE TARE Cause Measure The indicated value is Gross weight Press the key to change the indicated value into Net weight If is on the indicated value is Net weight The indicated value is not stable By setting Per...

Страница 75: ...COM on the control connector Tare subtraction is immediately reset and Net weight and Gross weight are equalized goes out If Net weight and Gross weight are not equalized by resetting Tare subtraction the following cause is presumable TARE RESET TARE Cause Measure The digital tare subtraction function is valid Set the value of Tare Weight in setting mode 1 to 0 or set digital tare subtraction ON O...

Страница 76: ...e Tare Weight 9 Tare Weight TARE F CNG ENT CNG ENT 2 LOWER 1 Digital tare subtraction ON 0 Digital tare subtraction OFF ZERO CNG ENT CNG ENT ZERO ZERO 3 NEAR Z ZERO F CNG ENT CNG ENT 1 UPPER Tare Weight 0 99999 9 FINAL CNG ENT CNG ENT Tare subtraction and digital tare subtraction functions independently When is pressed tare subtraction will be executed at once then Net weight becomes 0 while digit...

Страница 77: ... Valid Invalid expansion Prohibition of Tare Weight and digital tare subtraction ON OFF when one touch tare subtraction is working Select valid for inhibiting Tare Weight and digital tare subtraction ON OFF when tare subtraction is in operation Operation a b c d 1 Select setting mode 4 2 Put restrictions on the above four items 9 Restriction of the Tare Subtraction Function 0 0 F CNG ENT CNG ENT 4...

Страница 78: ... invalid Operation 1 Select setting mode 4 2 Select from Internal key mode External input mode 7 External Function Selection Switching by external signal A2pin In internal key mode Net weight display results on the ON edge OFF ON of the external input Gross weight display results on the OFF edge ON OFF of the external input In external key mode Net weight display results at ON and Gross weight dis...

Страница 79: ...ioning due to key operation a TARE key Select from Valid Invalid b TARE RESET key Select from Valid Invalid c ZERO key Select from Valid Invalid d GROSS NET key Select from Valid Invalid Operation a b c d 1 Select setting mode 2 2 Select whether each key is Valid or Invalid 5 Function Key Invalid F CNG ENT CNG ENT 4 SP1 At discharging weighing time 1 Net weight is displayed with the sign not rever...

Страница 80: ... selectable 6 1 Feeding Weighing and Discharging Weighing 6 1 1 Feeding Weighing A method controlling a weight while feeding raw materials into a measuring tank hopper Example of feeding weighing In this example system raw materials are to be fed from the material tank to the weighing tank At first the feeding valves are to be fully opened to feed and closed in the order of Set Point 1 Set Point 2...

Страница 81: ... tank is closed 4 When the weighed value reaches Final Compensation Set Point 3 output signal turns ON and if the judging timer is set the timer starts The feeding valves are completely closed 5 After the judging timer expires Over Under judgment is made If the weighed value exceeds the range of the Over Under setting value the over signal or under signal turns ON 6 Raw materials are discharged fr...

Страница 82: ...6 DISCHARGING CONTROL MODE 71 0 NET Over Under TIME Judgment Go Over or Under Near Zero Final SP1 Final SP2 Final CPS Final Tare subtraction SP2 SP1 SP3 Comparison Judging Timer Inhibit Timer Near Zero ...

Страница 83: ...y of raw materials are to be discharged from the weighing tank to a container At first the discharge valves of the weighing tank are to be fully opened to discharge raw materials and the discharge valves are to be closed in the order of Set Point 1 Set Point 2 and Set Point 2 Set Point 3 at the time of Final Set Point 1 and Final Set Point 2 respectively The discharge valves are to be completely c...

Страница 84: ...2 output signal turns ON If the comparison inhibit timer is set the timer starts The discharge valve Set Point 2 is closed 6 When the weighed value reaches Final Compensation Set Point 3 output signal turns ON and if the judging timer is set the timer starts The discharge valves are completely closed 7 After the judging timer expires Over Under judgment is made If the weighed value exceeds the ran...

Страница 85: ...dgment Go Over or Under Final CPS Lower Limit Final SP2 Final SP1 Raw materials are fed until Gross for Upper Lower limit comparison and Net for Final Set Point 2 and Near Zero comparisons In discharging weighing Net is compared negatively the upper limit signal is output Tare subtraction SP2 SP1 SP3 Comparison Judging Timer Lower Limit Inhibit Timer ...

Страница 86: ...etting mode 2 2 Select Discharging control mode 1 Weighing Function 1 Discharging control mode by external signal Switch with the pin set at Feed Discharge out of B2 to B5 pin on the control connector OFF Feeding control ON Discharging control F CNG ENT CNG ENT 2 LOWER Discharging control mode 2 External selection 1 Discharging control 0 Feeding control 1 UPPER CNG ENT CNG ENT OFF ON Undecided int...

Страница 87: ...nditions of final discharging setting items In simple comparison control the next weighing is judged to be possible when the weight falls short of 25 of the Final after completion of the previous weighing Near Zero Final SP1 Final SP2 Final CPS Final 0 NET Over Under TIME Near Zero Stable Complete Complete Timer OFF ON ON ON ON ON OFF Over Go Under ON t3 t2 t1 t1 ON SP2 SP1 SP3 Comparison Judging ...

Страница 88: ... value Final Under Select the weight value for Near Zero comparison from Gross weight Net weight Weighing Function 1 in setting mode 2 Select the comparison weight value for Set Point 1 Set Point 2 and Set Point 3 output signals and Over Go and Under judgment signals from Gross weight Net weight Weighing Function 1 in setting mode 2 6 2 2 Sequential Control This is a control method which compares ...

Страница 89: ... in setting mode 2 is ignored but comparisons are made when the completion output is ON and the weight value is held Upper Lower limit comparison is always made in defiance of the setting of Upper Lower limit comparison mode Weighing Function 2 in setting mode 2 0 NET Over Under TIME Start Stop Complete Judgment Over Go or Under ON ON OFF Complete Timer t3 ON ON ON ON t2 t1 t1 HOLD Near Zero Final...

Страница 90: ...alue Final CPS Under turns ON when Weight value Final Under Over turns ON when Weight value Final Over Go turns ON when Final Over Weight value Final Under Select the weight value for Near Zero comparison from Gross weight Net weight Weighing Function 1 in setting mode 2 Select the comparison weight value for Set Point 1 Set Point 2 and Set Point 3 output signals and Over Go and Under judgment sig...

Страница 91: ...to ON The completion signal output timing depends on the complete signal output mode setting Weighing Function 2 in setting mode 2 The Over Under comparison timing is such that the Over Under comparison mode setting Weighing Function 2 in setting mode 2 is ignored but comparisons are made when the completion output is ON and the weight value is held Upper Lower limit comparison is always made in d...

Страница 92: ...in setting mode 1 About the stop signal When the stop signal turns ON during weight cycle a sequence error will result and all the signals of Set Point 1 output Set Point 2 output and Set Point 3 output are turned OFF If the start signal is turned ON when the stop signal is ON it will become a sequence error and will not start the weighing START STOP SP1 Sequence error SP2 SP3 ON OFF ON ON ON ON O...

Страница 93: ...t the time of a sequence error 6 2 3 Sequence Mode Setting Simple comparison Sequence control Operation 1 Select setting mode 2 2 Select from Simple comparison mode Sequence mode 4 Sequence Mode ON ON OFF OFF OFF OFF START STOP SP1 Sequence error SP2 SP3 F CNG ENT CNG ENT 2 LOWER 1 Sequence mode 0 Simple comparison mode 4 SP1 CNG ENT CNG ENT ...

Страница 94: ...ient then add the result of Compensation value F701 C can regulate the value of D for minimizing the errors When D is within the range of Final AFFC Actual weight Final AFFC F701 C processes auto free fall compensation And when compensation feeding is ON in Sequence Mode F701 C samples the weight value before compensation feeding D1 D2 D3 DA A C Value added to Compensation value The auto free fall...

Страница 95: ...0 1 0 530 0 003 6 20 000 0 000 2 0 530 0 003 7 20 010 0 010 3 0 530 0 003 8 20 110 0 110 3 0 530 0 003 9 20 010 0 010 4 0 0 530 0 003 0 040 4 0 010 0 010 2 4 0 005 CPS Value 10 19 880 0 120 1 0 535 0 002 11 19 990 0 010 1 0 535 0 002 12 20 010 0 010 2 0 535 0 002 13 20 000 0 000 3 0 535 0 002 14 19 980 0 020 4 0 0 535 0 002 0 020 4 0 005 0 005 2 4 0 003 CPS Value 0 532 0 001 Setting compensation c...

Страница 96: ...ion Input range 1 9 d AFFC coefficient Setting for avoiding fluctuations of the compensation value be computed by multiplying the compensation and the coefficient of not more than 1 Operation Select from 1 3 4 2 4 or 1 4 a 1 Select setting mode 1 2 Input the AFFC 7 Auto Free Fall Compensation Regulation b c d 1 Select setting mode 2 2 Set AFFC ON OFF average count of AFFC and AFFC coefficient 3 We...

Страница 97: ...er Under TIME OFF ON ON ON ON OFF ON Near Zero Final SP1 Final SP2 Final CPS Final Near Zero Over Go Under SP2 SP1 SP3 Setting Discharging Fixed Value Setting Display Judging Formula Near Zero NZ Weight value Set value of Near Zero Set Point 1 SP1 Weight value Final SP1 Set Point 2 SP2 Weight value Final SP2 Set Point 3 SP3 Weight value Final CPS Under Weight value Final Under Over Weight value Fi...

Страница 98: ...ve compensation is effect e Over Input range 0 999 f Under Input range 0 999 Not for using Set Point 1 and Set Point 2 make the settings of Set Point 1 and Set Point 2 equal to the Final Operation a 1 Select setting mode 0 2 Input the Final 9 Final b 1 Select setting mode 0 2 Input the Set Point 2 5 Set Point 2 F CNG ENT CNG ENT 0 Final 0 99999 9 FINAL CNG ENT CNG ENT F CNG ENT CNG ENT 0 Set Point...

Страница 99: ...ompensation e 1 Select setting mode 0 2 Input the Over 7 Over f 1 Select setting mode 0 2 Input the Under 8 Under F CNG ENT CNG ENT 0 Set Point 1 0 99999 CNG ENT CNG ENT 4 SP1 F CNG ENT CNG ENT 0 Compensation 9999 9999 6 CPS CNG ENT CNG ENT F CNG ENT CNG ENT 0 Over 0 999 7 OVER CNG ENT CNG ENT F CNG ENT CNG ENT 0 Under 0 999 CNG ENT CNG ENT 8 UNDER ...

Страница 100: ... Inhibit Not including the negative value of the input range of compensation Input range is 0 9999 Effect Including the negative value of the input range of compensation Input range is 9999 9999 Operation 1 Select setting mode 2 2 Input the Negative Compensation 3 Weighing Function 3 F CNG ENT CNG ENT 2 LOWER 1 Effect 0 Inhibit CNG ENT CNG ENT 3 NEAR Z Negative Compensation ...

Страница 101: ...9999 c Lower Limit Weight value Lower Limit ON Input range 0 99999 Operation a 1 Select setting mode 0 2 Input the Near Zero 3 Near Zero b 1 Select setting mode 0 2 Input the Upper Limit 1 Upper Limit c 1 Select setting mode 0 2 Input the Lower Limit 2 Lower Limit F CNG ENT CNG ENT 0 Near Zero 0 99999 CNG ENT CNG ENT 3 NEAR Z F CNG ENT CNG ENT 0 Upper Limit 0 99999 CNG ENT CNG ENT 1 UPPER F CNG EN...

Страница 102: ...et weight d Over Under comparison Select from Gross weight Net weight Comparison OFF e Over Under comparison mode Select from Comparison Regularly External Judging ON Complete Signal ON Complete ON Hold Operation a c d 1 Select setting mode 2 2 Set Upper Lower limit comparison Near Zero comparison and Final and Over Under compari son 1 Weighing Function 1 F CNG ENT CNG ENT 2 LOWER Final and Over U...

Страница 103: ...ade when the complete output is ON 1 Comparison is made when the external judgment input is ON 0 Comparison regularly comparison mode 1 Comparison is made when the external judgment input is ON 0 Comparison regularly ZERO 2 LOWER CNG ENT CNG ENT ZERO OFF ON Undecided interval Judge within 50ms Be aware that no judgment is made if there is no judgment input set by the input selection when the over ...

Страница 104: ...ain period of time to eliminate false control or judgement with the vibration caused by opening or closing of valve Complete Output Time Setting the period of complete signal Control connector B10pin output when the weighing is completed 0 Final SP1 Final SP2 Final CPS Comparison Inhibit Time Judging Time 0 NET TIME ON ON Final CPS Final Near Zero Stab Complete Complete time Complete Output Time O...

Страница 105: ... 1 Select setting mode1 2 Input the Complete Output Time 3 Complete Output Time F CNG ENT CNG ENT 2 LOWER Complete signal output mode 2 ON for the Complete Output Time from when the judging timer has expired or from when the weight value becomes stable after the SP3 signal turns ON 1 ON for the Complete Output Time from when the weight value becomes stable after the judging timer has expired 0 ON ...

Страница 106: ...PPER Judging Time sec 0 00 9 99 2 LOWER CNG ENT CNG ENT F CNG ENT CNG ENT 1 UPPER Comparison Inhibit Time sec 0 00 9 99 CNG ENT CNG ENT 1 UPPER F CNG ENT CNG ENT 5 SP2 Output Selection 2 B10 pin 1 Complete output 0 Go output ZERO CNG ENT CNG ENT 2 LOWER Unless the weight value once drops down to 25 or less of the Final after the complete output the complete output will not turn on after the next w...

Страница 107: ...ting judging every time 02 Conducting an judging for every twice 03 Conducting an judging for every three times 99 Conducting an one judging in 99 times Complete Judging AFFC Judging Times 2 times Average count of AFFC 2 times The auto free fall compensation function stores compensation samples by using the Over Under judgment signals If the number of Judging Times is set to 00 no judgment the F70...

Страница 108: ...F Weighing will normally start if the Near Zero is ON is displayed if the Near Zero is OFF Refer to 6 6 Near Zero Upper Limit Lower Limit on P 90 concerning Near Zero setting 00 Without auto 0 01 Conduct an auto zero every time 02 Conduct an auto zero for every twice 03 Conduct an auto zero for every three times 99 Conducting one auto zero in 99 times Start Auto zero AZ times 3 times Auto zero is ...

Страница 109: ... Set Point 1 Compensation Over Under on P 86 concerning Set Point 1 setting e Adjust feeding Select from ON OFF f Adjust feeding time Input range 0 00 9 99 Operation a 1 Select setting mode 1 2 Input the Judging Times 6 Judging Times b 1 Select setting mode 1 2 Input the Auto Zero Times 5 Auto Zero Times F CNG ENT CNG ENT 1 UPPER Judging Times 0 99 6 CPS CNG ENT CNG ENT F CNG ENT CNG ENT 1 UPPER A...

Страница 110: ...de f 1 Select setting mode 1 2 Input the Adjust Feeding Time 4 Adjust Feeding Time F CNG ENT CNG ENT 2 LOWER Sequence control Sequence control 1 Adjust feeding ON 0 OFF 1 At start time weight value check ON 0 OFF 1 At start time Near Zero check ON 0 OFF ZERO 4 SP1 CNG ENT CNG ENT F CNG ENT CNG ENT 1 UPPER 4 SP1 Adjust Feeding Time sec 0 00 9 99 CNG ENT CNG ENT ...

Страница 111: ...ut range 0 99999 Operation a 1 Select setting mode 3 2 Input the Net Over 4 Net Over b 1 Select setting mode 3 2 Input the Gross Over 5 Gross Over Conditional formula Display Net Over Net weight Net Over set value Gross Over Gross weight Gross Over set value F CNG ENT CNG ENT 3 NEAR Z CNG ENT CNG ENT Net Over 0 99999 4 SP1 F CNG ENT CNG ENT 3 NEAR Z Gross Over 0 99999 CNG ENT CNG ENT 5 SP2 ...

Страница 112: ...ror When the display is Err the output turns ON Weight error and sequence error output signals are selectable by setting Operation 1 Select setting mode 5 2 Set Output selection 1 2 Output Selection F CNG ENT CNG ENT 5 SP2 Output Selection 1 B9 pin ZERO CNG ENT CNG ENT 2 LOWER ZERO 2 Weight error or sequence error 1 Sequence error 0 Weight error ...

Страница 113: ...be performed to clear the accumulated and statistical data While the number of accumulations is 10000 an accumulation error results The accumulation error can be outputted from B11 pin out of the external input output signals by setting Even while the accumulation error is given other weighing operations and the print command SI F etc are not affected Operations other than those relating to statis...

Страница 114: ...ulation Command Weight values can be accumulated by the external input output signal or through the RS 232C RS 485 option CC Link option when desired Accumulation command by the external input output Set one of the B2 to 5 pins out of the external input output signals to accumulation command An immediate accumulation is executed at the ON edge of an input to the set pin F CNG ENT CNG ENT 8 UNDER 1...

Страница 115: ...lation is executed by ON edge When occupied 4 or 2stations When 1 station is occupied RY0008 is used F CNG ENT CNG ENT 5 SP2 CNG ENT CNG ENT 1 UPPER Input selection 1 B2 pin Input selection 3 B4 pin 6 Accumulation Clear 5 Accumulation Command 4 Stop 3 Start 2 Feed Discharge 1 Judge 0 HOLD Input selection 2 B3 pin Input selection 4 B5 pin 6 Accumulation Clear 5 Accumulation Command 4 Stop 3 Start 2...

Страница 116: ...cleared by the password input or external input output signal or through the RS 232C RS 485 option CC Link option Accumulation clear by the external input output Set one of the B2 to 5 pins out of the external input output signals to accumulation clear Immediate accumulation clearing is executed at the ON edge of an input to the set pin F CNG ENT CNG ENT 9 FINAL 9 FINAL CNG ENT CNG ENT 1 2 UPPER L...

Страница 117: ...lation clearing is executed by ON edge When occupied 4 or 2stations When 1 station is occupied RY0009 is used F CNG ENT CNG ENT 5 SP2 CNG ENT CNG ENT 1 UPPER Input selection 1 B2 pin Input selection 3 B4 pin 6 Accumulation Clear 5 Accumulation Command 4 Stop 3 Start 2 Feed Discharge 1 Judge 0 HOLD Input selection 2 B3 pin Input selection 4 B5 pin 6 Accumulation Clear 5 Accumulation Command 4 Stop ...

Страница 118: ... accumulation value and statistical data are updated and also an auto print command is sent onto the SI F format However in the case where the result of weighing is negative or in the case of over scale no accumulation is made if no judgment is set Auto accumulation command timing 1 Simple comparison command 0 Near Zero Final Setpoint 1 Final Setpoint 2 Final CPS Final Weighing Value over under Ti...

Страница 119: ...re Start SP1 SP2 SP3 Complete ON ON OFF ON ON ON ON ON Judging Time Accumulation Regarding Simple Comparison Control and Sequence Control refer to 6 2 Simple Comparison Control and Sequence Control on P 76 Final Setpoint 1 Final Setpoint 2 Final CPS Final Weighing Value F CNG ENT CNG ENT 2 LOWER Accumulation command selection 1 Auto accumulation command ON 0 Auto accumulation command OFF ZERO CNG ...

Страница 120: ...from ON OFF Operation 1 Select setting mode 4 2 Select LOCK1 and LOCK2 ON OFF 8 Setting Value LOCK 8 2 Password Use for clearing statistical data in setting mode 8 Operation 1 Select setting mode 9 2 Input the password 1235 9 Pass Word CNG ENT F CNG ENT 4 SP1 LOCK2 1 ON 0 OFF LOCK1 1 ON 0 OFF 8 UNDER CNG ENT CNG ENT F CNG ENT CNG ENT 9 FINAL 9 FINAL CNG ENT CNG ENT 1 2 UPPER LOWER 5 SP2 3 NEAR Z ...

Страница 121: ... not correct it is faulty Ask our agency or us for repair Memory clear While pressing the and keys turn ON the power so that the memory SRAM setting values and work areas is cleared and the self check program is executed ALL CLr is displayed However the settings and calibrated values stored in the NOV RAM nonvolatile RAM remain unchanged Refer to 18 THE LIST OF INITIAL SETTING VALUE on P 176 Descr...

Страница 122: ...ce Error on P 101 6 8 Complete Signal Output Mode Complete Output Time Judging Time Comparison Inhibit Time Output Selection 2 on P 93 9 5 7 Accumulation Error on P 119 SP1 turns ON when Weight value Final SP1 SP2 turns ON when Weight value Final SP2 SP3 turns ON when Weight value Final CPS Connector FCN 361J024 AU Cover FCN 360C024 B A1 COM B1 COM A2 In G N B2 In Input Selection 1 1 A3 In D Z ON ...

Страница 123: ...ternal voltage to the signal input circuit The external element is required to withstand Ic 10mA Leakage from the external element is required to be 100μA or below Switch IC 6mA TTL Open Collector ON when IN is HI Spark killer Load Spark killer DC Power Varistor Load AC power Relay Vext Output data Tr 0 OFF 1 ON F701 C Inside Vcc 12V COM Vceo 30V max Ic 120mA max Use external power source up to DC...

Страница 124: ... ON OFF display switches to Gross weight In the case of level input G N setting external input mode Net weight display results at ON and Gross weight display results at OFF 9 4 2 Digital Zero D Z ON Edge Input When external input is ON edge OFF ON it brings the Gross weight to zero This function works within the range of DZ Regulation Value If out of this range zero will not result but flashes The...

Страница 125: ... subtraction can be selected from all range or 0 Tare Capacity During Tare subtraction lights 9 4 4 Tare Subtraction Reset TARE OFF Edge Input Reset the Tare by external input edge OFF ON This does not apply to Tare Weight The key operation also results in the same movement when TARE RESET key is valid under the Function Key Invalid setting in setting mode 2 9 4 5 Hold HOLD Level Input The weight ...

Страница 126: ...neffective in Sequence Mode 9 4 7 Feed Discharge Level Input Discharging control mode Weighing Function 1 in setting mode 2 is effective in external selection For the method of setting refer to 6 1 3 Discharging Mode on P 75 Feeding control is activated when it is OFF Discharging control is activated when it is ON 9 4 8 Input Signals Used in Sequence Mode For the method of setting refer to 6 2 3 S...

Страница 127: ...N edge OFF ON of the external input The weight value gross or net set in the over under comparison mode is accumulated 9 4 10 Accumulation Clear Edge Input Immediate clearing of accumulated data and statistical data is executed at the ON edge OFF ON of the external input OFF ON Pulse width 50msec or more OFF ON Pulse width 50msec or more ...

Страница 128: ...l result When HOLD is selected on two or more pins the weight value is held if at least one signal is ON Operation 1 Select setting mode 5 2 Set the Input selection items 1 Input Selection F CNG ENT CNG ENT 5 SP2 CNG ENT CNG ENT 1 UPPER Input selection 1 B2 pin Input selection 3 B4 pin 6 Accumulation Clear 5 Accumulation Command 4 Stop 3 Start 2 Feed Discharge 1 Judge 0 HOLD Input selection 2 B3 p...

Страница 129: ...lue Upper Limit For the method of setting refer to 6 6 Near Zero Upper Limit Lower Limit on P 90 6 7 Upper Lower Limit Comparison Upper Lower Limit Comparison Mode Near Zero Comparison Final and Comparison Over Under Comparison Mode on P 91 9 5 3 Stable This output turns ON when the weight value is stable Refer to 5 6 Motion Detection MD on P 58 for details 9 5 4 Weight Error When the display is L...

Страница 130: ...nal of output selection 3 by setting Operation 1 Select setting mode 5 2 Set Output selection 3 2 Output Selection 9 5 7 Accumulation Error The output turns ON when the number of accumulations is 10000 The accumulation error can be outputted externally as the signal of output selection 3 by setting For the method of setting refer to 9 5 6 RUN on P 119 For the accumulation error refer to 7 STATISTI...

Страница 131: ...f the start signal OFF ON Set Point 1 Set Point 2 and Set Point 3 turn ON SP1 turns OFF when Weight value Final SP1 SP2 turns OFF when Weight value Final SP2 SP3 turns OFF when Weight value Final CPS For the method of setting refer to 6 4 Final Set Point 2 Set Point 1 Compensation Over Under on P 86 6 7 Upper Lower Limit Comparison Upper Lower Limit Comparison Mode Near Zero Comparison Final and C...

Страница 132: ...gnal ON and hold it while ignoring the setting of Over Under comparison mode Weighing Function 2 in setting mode 2 only when judge ON The under signal turns ON when Weight value Final Under The over signal turns ON when Weight value Final Over The go signal turns ON when Final Over Weight value Final Under The go output signal is used for complete output by setting Either go or complete is selecta...

Страница 133: ...mplete Output Time in setting mode 1 Without Over Under judgment Conditions The setting of Final and Over Under comparison is comparison OFF The number of Judging Times is 0 The complete output turns ON in defiance of the setting of complete signal output mode Weighing Function 2 in setting mode 2 on the OFF edge ON OFF of the set point 3 signal The output ON time depends on the Complete Output Ti...

Страница 134: ...and Go The complete and go are outputted from output selection 2 of B10 pin RUN and accumulation error The RUN and accumulation error are outputted from output selection 3 of B11 pin Operation 1 Select setting mode 5 2 Set the Output selection 1 2 and 3 2 Output Selection F CNG ENT CNG ENT 5 SP2 CNG ENT CNG ENT 2 LOWER Output selection 3 B11 pin 1 Accumulation Error 0 RUN Output selection 1 B9 pin...

Страница 135: ...or the connection of remote displays LD557A LD514 M252B Each display can select the data Gross weight Net weight Tare weight individually 10 1 2 Auto Print Command The F701 C gives an auto print command to any device connected by SI F interface In sequence mode the auto print command is output when the judgment signal Go Over Under turns ON Therefore be aware that the auto print command is not out...

Страница 136: ...iagram is for connecting a personal computer as a DTE Data Terminal Equipment device If it is a DCE Data Circuit terminating Equipment device connect pin to pin DTR to DTR DSR to DSR etc Cables should be prepared after checking connector type and pin assignments of the connected device Adaptable connector 9 pin D SUB connector JAE DE 09SN OMRON XM3D 0921 etc 1 6 2 In RXD 7 Out RTS 3 Out TXD 8 In C...

Страница 137: ...puter The terminator can be selected CR or CR LF Communication mode 1 F701 C sends Gross weight continuously Command is invalid at this mode F CNG ENT CNG ENT 4 SP1 Baud rate 5 38400bps 4 19200bps 3 9600bps 2 4800bps 1 2400bps 0 1200bps Communication mode Character length 1 8bit 0 7bit Stop bit 1 2bit 0 1bit Parity bit 2 Even 1 Odd 0 None 7 Communication mode 6 6 Communication mode 5 5 Communicati...

Страница 138: ...weight at the timing of auto print command Command is invalid at this mode Communication mode 6 F701 C sends Gross and Net weight at the timing of auto print command Command is invalid at this mode Transmission format 2 20byte Transmission format 3 23byte Transmission format 1 20byte Transmission format 2 20byte Transmission format 3 23byte Please refer to each page about following items Transmiss...

Страница 139: ... Net weight 5 digit Sign gross weight or Gross weight 5 digit 5 SOH STX ETX BCC Sign net weight or For BCC calculation Each ASCII code of SOH STX and ETX SOH 01 STX 02 ETX 03 BCC Block check character It is calculated as the XOR of the transmitted character Hexadecimal 1 O Overload LOAD OFL S Stable M Unstable H Hold Priority H O S or M 4 N Near zero OFF Z Near zero ON 2 A Zero tracking OFF T Zero...

Страница 140: ... digit weight value decimal point Reading out the status 1 7 digit Host F701 C R A 1 0 0 0 0 CR LF R A CR Host F701 C R B 1 0 0 0 0 CR LF R B CR Host F701 C R C 1 0 0 0 0 CR LF R C CR 1 ON 0 OFF Rear panel LOCK Weight Not defined Weight value hold Weight value Zero alarm Weight value stable Tare subtracted 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF 0 Gross 1 Net 0 OFF 1 ON display Host F701 C R D 0 0 0 0 0 ...

Страница 141: ...utput signal 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF SP1 output signal SP2 output signal SP3 output signal 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF Host F701 C R E 0 0 0 0 0 0 0 CR LF R E CR 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF Near zero output signal Lower limit output signal Upper limit output signal Not defined Host F701 C R F 0 0 0 0 0 0 0 CR LF R F CR ...

Страница 142: ...OFL2 1 OFL1 0 Weighting value Over flow Sequence error 0 or 1 5 Calibration error 0 or 2 9 Same as the error display of F701 C Same as the error display of F701 C Host F701 C R G 0 0 0 0 0 0 0 CR LF R G CR Host F701 C R H 9 9 9 9 CR LF R H CR 9 Host F701 C R I 9 9 9 9 9 9 9 CR LF R I CR 9 9 Host F701 C R CR R J 0 0 1 0 0 LF J 0 CR 0 Fixed value Host F701 C R CR R J CR LF J Up to 256 pieces of data...

Страница 143: ...ing value No Setting value LF No return data Host F701 C W W CR CR Setting value No Setting value Setting value No LF Host F701 C C A N 0 CR A Calibrating 0 Normally completed 1 9 Error same as the error display of F701 C CR LF Host F701 C C B N 0 CR B Calibrating 0 Normally completed 1 9 Error same as the error display of F701 C CR LF Zero Calibration Span Calibration Before sending this command ...

Страница 144: ... zero reset Accumulation Command Accumulation Clear Host F701 C C C No return data CR Host F701 C C D No return data CR Host F701 C C E No return data CR Host F701 C C F No return data CR Host F701 C C G No return data CR Host F701 C C H No return data CR Host F701 C C I No return data CR Host F701 C C J No return data CR ...

Страница 145: ...㼀㼍㼞㼑㻌㼃㼑㼕㼓㼔㼠 䠳 㻞 㻡 䠟䠮 䠨䠢 㻭㼡㼠㼛㻌㼆㼑㼞㼛㻌㼀㼕㼙㼑㼟 䠳 㻞 㻢 㻜 㻜 㻜 䠟䠮 䠨䠢 㻶㼡㼐㼓㼕㼚㼓㻌㼀㼕㼙㼑㼟 䠳 㻞 㻣 㻜 㻜 㻜 䠟䠮 䠨䠢 㻯㼛㼙㼜㼘㼑㼠㼑㻌㻻㼡㼠㼜㼡㼠㻌㼀㼕㼙㼑 䠳 㻞 㻤 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔...

Страница 146: ...㼠㼕㼛㼚㻌㼂㼍㼘㼡㼑 䠳 㻠 㻣 㻜 䠟䠮 䠨䠢 㻳㼞㼍㼢㼕㼠㼍㼠㼕㼛㼚㼍㼘㻌㻭㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㻔㼍㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㼕㼚㼜㼡㼠㻕 䠳 㻠 㻤 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏...

Страница 147: ... 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪞㪼㫅㪼㫉㪸㫃㩷㪪㫋㪸㫅㪻㪸㫉㪻㩷㪛㪼㫍㫀㪸㫋㫀㫆㫅 䌗 㪏 㪊 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪪㪸㫄㫇㫃㪼㩷㪪㫋㪸㫅㪻㪸㫉㪻㩷㪛㪼㫍㫀㪸㫋㫀㫆㫅 䌗 㪏 㪋 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪤㪸㫏㪅㩷㪄㩷㪤㫀㫅㪅 䌗 㪏 㪌 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪘㪺㪺㫌㫄㫌㫃㪸㫋㫀㫆㫅㩷㪚㫆㫌㫅㫋 䌗 㪏 㪍 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪣㪸㫋㪼㫊㫋㩷㪘㪺㪺㫌㫄㫌㫃㪸㫋㫀㫆㫅㩷㪛㪸㫋㪸 䌗 㪏 㪎 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉...

Страница 148: ...t such as a relay or a switch or a noncontact such as a transistor or an open collector TTL Adaptable connector DDK 57 30360 or equivalent one 1 COM 19 COM 2 Out 1 20 Out 20000 3 Out 2 21 Out 40000 4 Out 4 22 Out 80000 5 Out 8 23 Out MINUS 6 Out 10 24 Out OVER 7 Out 20 25 Out P C 8 Out 40 26 Out STROBE 9 Out 80 27 In Data hold 10 Out 100 28 In Logic switching 11 Out 200 29 In Output selection 1 12...

Страница 149: ... Output MINUS The polarity of weight values output as BCD data is output as follows When positive 0 When negative 1 F701 C Inside COM Vext Vceo 30V max Ic 120mA max Internal transistor status Output data Negative Positive 0 OFF ON 1 ON OFF Output pin level Output data Negative Positive 0 H L 1 L H Through logic switching pin28 Outside Vcc 12V digit data 8 4 2 1 0 0 0 0 0 1 0 0 0 1 2 0 0 1 0 3 0 0 ...

Страница 150: ...t in synchronization with the BCD data For reading data use the pulse end edge The data update rate is normally 200 times sec which is the same as the operating speed of the main unit but data may not be read with some connected external devices because the speed is too fast In such a case set the data update rate longer reduce the number of times Weight value Conditional expression Net weight Net...

Страница 151: ...put Select weight values to be output as BCD data F CNG ENT CNG ENT 4 SP1 Data update rate 7 1 times sec STROBE range Approx 500msec 6 2 times sec STROBE range Approx 250msec 5 5 times sec STROBE range Approx 100msec 4 10 times sec STROBE range Approx 50msec 3 20 times sec STROBE range Approx 25msec 2 50 times sec STROBE range Approx 10msec 1 100 times sec STROBE range Approx 5msec 0 200 times sec...

Страница 152: ...otentiometer For fine adjustment of zero Trimmer adjusts output to 4 00mA for zero Zero can be adjusted through the front panel as well The terminals for current output of 4 20mA Connect an instrument whose load resistance is less than 350 Ω Current output terminals The trimmer is for fine adjustment Temperature drift may increase if you have severely changed the value that had been set when you f...

Страница 153: ...equipment load resistance 350Ω or less between CUR G of F701 C External equipment Load resistance 350Ω or less SIG GND Inside Outside CUR F701 C G Do not apply voltage or current to the current output terminal externally by mistake Doing so will cause breakage CAUTION ...

Страница 154: ...ut section Output connector Cage clamp type terminal block two piece type Potentiometer A potentiometer to readjust fine adjust ZERO 4mA and GAIN 20mA is placed on the rear panel The adjustable range is a few percentage 4mA and 20mA are factory adjusted 10 4 3 Method of Adjusting the D A Zero and Gain The D A converter of the F701 C obtains analog output by setting the weight value to output 4mA o...

Страница 155: ... c 1 Select setting mode 4 2 Input the D A Full Scale 3 D A Full Scale F CNG ENT CNG ENT 4 SP1 Test mode 2 20mA fixed output 1 4mA fixed output 0 Tied to the weight value Output mode 1 Net weight 0 Gross weight 1 UPPER CNG ENT CNG ENT F CNG ENT CNG ENT 4 SP1 D A Zero Output Weight 0 99999 Input the weight value to output 4mA The default value is 00000 2 LOWER CNG ENT CNG ENT F CNG ENT CNG ENT 4 SP...

Страница 156: ...1 10000 with respect to 4 20mA That is the minimum unit of current is as follows 20 4mA 1 10000 1 6μA Also the minimum unit of weight value is as follows D A Full Scale setting value 1 10000 Current mA 20 4 D A Zero Output Weight setting Resolution 1 10000 D A Full Scale setting Net 500 2000 2500 Net Current mA 480 500 1000 1500 2500 2520 3 84 4 00 8 00 12 00 20 00 20 16 Zero Full scale ...

Страница 157: ...as internal terminal resistor By turning ON the Rt switch it is not necessary to install external resistance The terminal SG is a grand terminal used on the circuit for protecting the circuit When the main body of F701 C and the device connected to F701 C are grounded by D type ground there is usually no need to use the terminal SG However confirm the specifications of the devise connected before ...

Страница 158: ...B SD A B A B SD RD A B A B SD RD A B A B SD RD A B A B SD RD Rt Twisted para cable F701 C F701 C F701 C F701 C Rt 120ΩRt can be installed or removed according to the switch The switch is available on the option sheet metal section at the back of the main unit Rt OFF Rt OFF Rt OFF Rt ON CNG ENT F CNG ENT 4 SP1 Baud rate 5 38400bps 4 19200bps 3 9600bps 2 4800bps 1 2400bps 0 1200bps Terminator 1 CR L...

Страница 159: ...ion so that normal communications cannot be carried out 4 ID number setting Operation 1 Select setting mode 4 2 Input the ID Number 5 ID Number When the ID number is 0000 transmittable and receivable conditions are brought about at power on therefore set ID numbers other than 0000 for connecting two or more F701 Cs If the ID number setting is other than 0000 other formats R W C etc are invalid unt...

Страница 160: ... digit weight value decimal point Reading out the status 1 7 digit Host F701 C R A 1 0 0 0 0 CR LF R A CR Host F701 C R B 1 0 0 0 0 CR LF R B CR Host F701 C R C 1 0 0 0 0 CR LF R C CR 1 ON 0 OFF Rear panel LOCK Weight Not defined Weight value hold Weight value Zero alarm Weight value stable Tare subtracted 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF 0 Gross 1 Net 0 OFF 1 ON display Host F701 C R D 0 0 0 0 0 ...

Страница 161: ...utput signal 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF SP1 output signal SP2 output signal SP3 output signal 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF Host F701 C R E 0 0 0 0 0 0 0 CR LF R E CR 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF Near zero output signal Lower limit output signal Upper limit output signal Not defined Host F701 C R F 0 0 0 0 0 0 0 CR LF R F CR ...

Страница 162: ...OFL2 1 OFL1 0 Weighting value Over flow Sequence error 0 or 1 5 Calibration error 0 or 2 9 Same as the error display of F701 C Same as the error display of F701 C Host F701 C R G 0 0 0 0 0 0 0 CR LF R G CR Host F701 C R H 9 9 9 9 CR LF R H CR 9 Host F701 C R I 9 9 9 9 9 9 9 CR LF R I CR 9 9 Host F701 C R CR R J 0 0 1 0 0 LF J 0 CR 0 Host F701 C R CR R J CR LF J Fixed value Measurement value Up to ...

Страница 163: ...ing value No Setting value LF No return data Host F701 C W W CR CR Setting value No Setting value Setting value No LF Host F701 C C A N 0 CR A Calibrating 0 Normally completed 1 9 Error same as the error display of F701 C CR LF Host F701 C C B N 0 CR B Calibrating 0 Normally completed 1 9 Error same as the error display of F701 C CR LF Zero Calibration Span Calibration Before sending this command ...

Страница 164: ... zero reset Accumulation Command Accumulation Clear Host F701 C C C No return data CR Host F701 C C D No return data CR Host F701 C C E No return data CR Host F701 C C F No return data CR Host F701 C C G No return data CR Host F701 C C H No return data CR Host F701 C C I No return data CR Host F701 C C J No return data CR ...

Страница 165: ...㼀㼍㼞㼑㻌㼃㼑㼕㼓㼔㼠 䠳 㻞 㻡 䠟䠮 䠨䠢 㻭㼡㼠㼛㻌㼆㼑㼞㼛㻌㼀㼕㼙㼑㼟 䠳 㻞 㻢 㻜 㻜 㻜 䠟䠮 䠨䠢 㻶㼡㼐㼓㼕㼚㼓㻌㼀㼕㼙㼑㼟 䠳 㻞 㻣 㻜 㻜 㻜 䠟䠮 䠨䠢 㻯㼛㼙㼜㼘㼑㼠㼑㻌㻻㼡㼠㼜㼡㼠㻌㼀㼕㼙㼑 䠳 㻞 㻤 㻜 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻝㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔...

Страница 166: ...㼠㼕㼛㼚㻌㼂㼍㼘㼡㼑 䠳 㻠 㻣 㻜 䠟䠮 䠨䠢 㻳㼞㼍㼢㼕㼠㼍㼠㼕㼛㼚㼍㼘㻌㻭㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㻔㼍㼏㼏㼑㼘㼑㼞㼍㼠㼕㼛㼚㻌㼕㼚㼜㼡㼠㻕 䠳 㻠 㻤 㻜 䠟䠮 䠨䠢 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏㼍㼚㻌㼚㼛㼠㻌㼣㼞㼕㼠㼑㻌㼕㼚䠅 㻔㼃㼔㼑㼚㻌㻸㻻㻯㻷㻞㻌㼍㼚㼐㻌㻸㻻㻯㻷㻌㻿㼃㻌㼏...

Страница 167: ... 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪞㪼㫅㪼㫉㪸㫃㩷㪪㫋㪸㫅㪻㪸㫉㪻㩷㪛㪼㫍㫀㪸㫋㫀㫆㫅 䌗 㪏 㪊 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪪㪸㫄㫇㫃㪼㩷㪪㫋㪸㫅㪻㪸㫉㪻㩷㪛㪼㫍㫀㪸㫋㫀㫆㫅 䌗 㪏 㪋 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪤㪸㫏㪅㩷㪄㩷㪤㫀㫅㪅 䌗 㪏 㪌 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪘㪺㪺㫌㫄㫌㫃㪸㫋㫀㫆㫅㩷㪚㫆㫌㫅㫋 䌗 㪏 㪍 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㪣㪸㫋㪼㫊㫋㩷㪘㪺㪺㫌㫄㫌㫃㪸㫋㫀㫆㫅㩷㪛㪸㫋㪸 䌗 㪏 㪎 䌃䌒 䌌䌆 㩿㪩㪼㪸㪻㩷㫆㫌㫋㩷㫆㫅㫃㫐㪀 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉㩷㪺㪸㫅㩷㫅㫆㫋㩷㫎㫉㫀㫋㪼㩷㫀㫅䋩 㩿㪮㪿㪼㫅㩷㪣㪦㪚㪢㪉...

Страница 168: ...cale division Gross weight Gross Over set value Error item Alarm message When start signal turns on the stop signal turns ON During weigh cycle the stop signal turns ON Zero Alarm turns on according to set times When the start of weighing Near Zero output signal turns OFF At start Near Zero confirmation on is selected for sequence control When the start of weighing the value SP1 At start weight va...

Страница 169: ...ero adjustment range The initial Tare is minus The input balance weight is beyonde the Capacity The balance weight is 00000 The load cell output dose not reach the span adjustment range The load cell output is miunus The load cell output is beyond the span adjustment range The weight value is not stable and calibration stopped ...

Страница 170: ... cable which is connected to the F701 C and the load cell And whether the load cell connector on the rear panel is open or not when the load cell input termi nal block is open not connected with the load cell will display also Net weight Net Over set value Net weight exceeds the Net Over set value Decrease the signal coming from the load cell unitl the normal display returns or reset the Net Over ...

Страница 171: ... Near Zero setting value and the Near Zero comparison target Next check the following for occurrence A start was made before complete discharge The start ON timing was too early Discharge is in a jam etc Set the stop signal from OFF to ON to resolve the sequence error Sequence error It is displayed if the Set Point 1 output SP1 signal is ON when weighing starts when it is set to check the weight v...

Страница 172: ...acy of the controller Use a resistance having a temperature coefficient of 50ppm or more recom mended value of about 5ppm Resistor Strain Ideal Approx μ STRAIN mV V 875 KΩ 866 KΩ 200 0 1 437 KΩ 442 KΩ 400 0 2 291 KΩ 294 KΩ 600 0 3 219 KΩ 221 KΩ 800 0 4 175 KΩ 174 KΩ 1000 0 5 146 KΩ 147 KΩ 1200 0 6 125 KΩ 124 KΩ 1400 0 7 109 KΩ 110 KΩ 1600 0 8 97 KΩ 97 6 KΩ 1800 0 9 87 3 KΩ 86 6 KΩ 2000 1 0 79 4 KΩ...

Страница 173: ...ion again The relationship between Capacity and Balance Weight Value Balance Weight Value must be between 50 to 100 of Capacity in order to do Span Calibration correctly Calibration error The set Balance Weight Value is 00000 Set adequate value to Balance Weight Value Calibration error The load cell output dose not reach the span adjustment range of the F701 C Check how load is applied to load cel...

Страница 174: ...cell check load cell has sufficient output mV V to reach span range Do Span Calibration again Calibration error The load is not stable enough for correct Calibration Adjust the stable period and stable range of Motion Detection so display lights during Calibration Do Calibration again STAB The span adjustable range of the F701 C is 0 3 3 2 mV V Since display to the Capacity is guaranteed in Span C...

Страница 175: ...in some cases While is lit do not turn off the power of the F701 C Setting values have been changed when the power is turned on The voltage of the lithium battery for memory backup may have dropped The setting values stored in the SRAM refer to 18 THE LIST OF INITIAL SETTING VALUE on P 176 are backed up by the lithium battery If the power of the F701 C is turned off with the voltage of the lithium...

Страница 176: ... occurs in self checking Checksum error It is displayed if a checksum error occurs in self checking or at power on NOV RAM check error It is displayed if a NOV RAM check error occurs in self checking In the case of self check error display the F701 C is faulty Request repair from your distributor or us ...

Страница 177: ...g it 3 Remove the connector 4 Remove the screw fixing the nylon clamp and replace the lithium battery with a new one Replacement battery CR14250SE We deal with it Specify the lithium battery for the F701 C 5 After connection of the connector of the new lithium battery make a cold start While pressing the and keys at the same time turn on the power Initial values are written into the SRAM as settin...

Страница 178: ...V 13V 12V VCC ISOLATION Indicates the insulation by Photo coupler Monitoring RAM 128Kbyte CONTOROL CONTOROL BCD DATA OUTPUT RS 232C RS 485 D A CONVERTER CPU 16bit RESET WD NOV RAM 4096bit OPTION LOCK SW LITHIUM BATTERY 8 IN 12 OUT 24 OUT 8 IN SI F character display tube BASE 10V EXCITATION 5V DISPLAY KEY INPUT SI F AC4V For fluorescent 4 20mA ADJUSTMENT AMPLIFIER VOLTAGE VOLTAGE CONVERSION 24bit I...

Страница 179: ...15 DIMENSIONS 168 15 DIMENSIONS Unit mm ...

Страница 180: ... and insert the controller into the panel 3 Insert the mounting rails into both sides from the back of the controller 4 Securely fix the mounting hardware on both sides with the attached M4 screws Panel cutout size 138W 68H mm Unit mm 138 68 Mounting panel thickness 1 6mm or more For transportation after panel mounting consideration should be given so as not to give excessive shocks or vibration C...

Страница 181: ...stment circuit Fine adjustment Automatic adjustment by digital operation Minimum input sensitivity 0 3μV count Accuracy Non linearity within 0 01 FS Typ 0 005 FS ordinary Zero drift within 0 2μV RTI Typ 0 15μV Gain drift within 15ppm Typ 5ppm Analog filter Bessel type low pass filter 12dB oct Selectable from 2 4 6 and 8Hz A D converter Speed 200 times sec Resolution 24bit binary Minimum indication...

Страница 182: ...ect from 3 6 13 or 25 times sec The system speed is 200 times sec Status display SP1 SP2 SP3 NEAR Z HOLD ZALM STAB TARE NET ZERO 17 3 Setting Section Setting method Setting by keyboard operation embossed membrane switches with key click sound External setting also allowable through the RS 232C Memory of set values Initial values NOV RAM nonvolatile RAM Other values Lithium battery backed up C MOS ...

Страница 183: ... Tare OFF Hold Judge Feed Discharge 8 points Start Stop Accumulation Command Accumulation Clear ON when short circuited with the COM terminal by contact relay switch etc or non contact transistor open collector output etc Output signals Near Zero SP1 SP2 SP3 Under Over Upper Limit Lower Limit 12 points Stab Weight Error Error Go Complete RUN Accumulation Error Transistor open collector output Emit...

Страница 184: ...t values are output as converted to analog signals of current Zero output weight value and full scale weight value can be set D A converter Current output 4 20mA D A conversion speed 200 times sec Resolution 1 10000 Over range 5 of the full scale Current 3 2mA 20 8mA Zero drift Within 0 3μV Gain drift Within 30ppm Non linearity Within 0 02 FS Not including the drift of the analog input section RS ...

Страница 185: ...tage AC100 240V 10 15 free power source 50 60Hz Power consumption Approx 15VA Rush current 20A 5mSec 100V AC mean load state ordinary temperature at cold start time 40A 5mSec 200V AC mean load state ordinary temperature at cold start time Operating conditions Temperature Operation 10 40 Storage 20 85 Humidity 85 RH or less non condensing Dimension 144 W 72 H 160 D mm Projections excluded Panel cut...

Страница 186: ... option 1 Operation manual 1 Jumper line 2 About the power cable The power cable attached to this product as standard equipment can be used in the AC100V power supply in Japan Official ratings voltage AC125V Please use the power cable authorized in the country when you use this product outside Japan Our company sells following resistance pressure cable AC250V European standard product separately P...

Страница 187: ...AM Display Setting cannot be changed 18 1 Setting Mode 0 18 2 Setting Mode 1 0 F CNG ENT CNG ENT Setting item Initial value LOCK 1 SRAM LOCK 2 NOV RAM LOCK SW NOV RAM Display 1 Upper Limit 000 00 2 Lower Limit 000 00 3 Near Zero 000 00 4 Set Point 1 000 00 5 Set Point 2 000 00 6 Compensation 00 00 7 Over 0 00 8 Under 0 00 9 Final 000 00 F CNG ENT CNG ENT 1 UPPER Setting item Initial value LOCK 1 S...

Страница 188: ...nge 0000 F CNG ENT CNG ENT 3 NEAR Z Setting item Initial value LOCK 1 SRAM LOCK 2 NOV RAM LOCK SW NOV RAM Display 1 Balance Weight Value 100 00 2 Capacity 100 00 3 Min Scale Division 0 01 4 Net Over 999 99 5 Gross Over 999 99 6 DZ Regulation Value 02 00 7 Function Selection 13213 8 Gravitational Acceleration area number input 09 9 Gravitational Acceleration acceleration input 9 798 F CNG ENT CNG E...

Страница 189: ...G ENT CNG ENT 8 UNDER Setting item Initial value LOCK 1 SRAM LOCK 2 NOV RAM LOCK SW NOV RAM Display 1 Average Weight 000 00 2 Max Value 000 00 3 Min Value 000 00 4 General Standard Deviation 000 00 5 Sample Standard Deviation 000 00 6 Count of Data 00000 7 Latest Accumulation Data 000 00 8 Max Min 000 00 9 Optional Board 0000 F CNG ENT CNG ENT 9 FINAL Setting item Initial value LOCK 1 SRAM LOCK 2 ...

Отзывы: