4
CALIBRATION
32
CALIBRATION
Chapter
4
■
Calibration value selection by external signal input
With this function, four types of calibration values can be selected with external selector signals
CAL0 and CAL1 (when the calibration value selection setting is external). Set EXT 0 by pressing
the
button on the CALIBRATION screen. According to the input conditions of external
signal inputs CAL0 and CAL1, the display changes as EXT 0 to EXT 3. Normally, when there is no
input to CAL0 and CAL1 (the terminals are open), calibration value 0 is selected. When each
terminal is in the following condition, each calibration value is selected.
4-4.
Analog Input Selection
Select the output type of the device (sensor) connected to analog input terminal.
The output type can be chosen
±
10V (voltage) or
±
20mA (current).
After this setting, be sure to perform calibration.
How to set
4-5.
Unit
Set the unit of the values displayed. For settable units, see the "Unit Setting List"on page 88.
How to set
1.
Press the [SET.] button. (refer to page 33)
2.
Press the [Calibration] button. (refer to page 33)
3.
Select page 2, and
press the [Unit]
button.
CAL1
CAL0
Calibration value
Open
Open
Calibration value 0
Open
Short-circuit
Calibration value 1
Short-circuit
Open
Calibration value 2
Short-circuit
Short-circuit
Calibration value 3
B1 or B7
COM2
B11
CAL0
B12
CAL1
CAUTION
It takes one second at maximum for
the changed calibration value to
become effective. During this time,
the calibration value is indefinable.
Also, the indicated value is
accordingly indefinable.
I/O connector
SET.
Calibration
Page 1
→
→
SET.
Calibration
Page 2
→
→
Содержание F377A
Страница 1: ...18JUL2012REV 3 06 DIGITAL INDICATOR F377A OPERATION MANUAL ...
Страница 5: ...What is RoHS 4 M E M O ...
Страница 12: ...11 11 STANDARD ...
Страница 13: ...12 12 M E M O ...
Страница 84: ...7 SPECIFICATIONS 83 SPECIFICATIONS Chapter 7 Equipped with Device Net interface option Unit mm ...
Страница 94: ...93 93 EXPANSION ...
Страница 95: ...94 94 M E M O ...
Страница 97: ...Expansion Function 96 M E M O ...
Страница 152: ......