6HOLD FUNCTION
55
・
Timing Chart
t1
:
Time from the short-circuiting of the hold input (OFF
→
ON) to the display of the
peak hold value.
t2
:
Time until the A/D conversion of the analog value.
t3
:
Time from the input of the hold input (ON
→
OFF) to the reset of the analog peak
hold value.
t4
:
The minimum tracking (resetting) time required for resetting the display of the
held value.
Time
t1
t2
t3
t4
Hold
input
Display and
BCD output
t1
:
MAX 25mS
t2
:
MAX 10mS
t3
:
MAX 25mS
t4
:
MIN 25mS
Indicated value