Phyaction 190
1) Turn off the Phyaction 190
2) Connect the treatment head you wish to tune
to the Phyaction 190
3) Press the blue function keys number 4 and 6
(the right most blue key and the blue key
third from the right) and keep them pressed
4) Turn on the Phyaction 190, while function
keys 4 and 6 are pressed.
When you hear the buzzer sound, release
both function keys. When you have done this
correctly, you will see the screens from figure
1, followed by figure 2 (of course the version
numbers can differ). When you release the
keys too late, you will get an error message
(error 41).
5) In the normal user screen (see figure 3) the
word 'Service' will appear in the top right
corner.
6) Keep the yellow key pressed, until the
'Service Menu' appears (see figure 4).
7) Now press function key 1 (Set head
parameters). Figure 5 appears on the
screen. If it does not appear, check if a
treatment head is connected.
8) If necessary, select the desired head with
function key 1. Then enter the correct head
parameters, using the up and down keys and
the left and right keys (function keys 3 and
4). The head parameters are on the label of
the treatment head.
9) When the parameters are entered correctly
you can press the Ok button (function key 5).
The screen as shown in figure 6 will appear.
When the Phyaction 190 detects an invalid
set of parameters, the Ok key will not be
accepted.
Ultrasound treatment head installation procedure
Valid from SW-version 3.01
In order to tune a treatment head to the Phyaction 190, it is necessary to take the following steps:
Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
Содержание PHYACTION 190
Страница 21: ......
Страница 22: ......
Страница 23: ......
Страница 24: ......
Страница 25: ......
Страница 26: ......
Страница 27: ......
Страница 28: ......
Страница 29: ......
Страница 30: ......
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 36: ......
Страница 37: ......
Страница 38: ......