5
Technical data
Gas inlets
Number of measuring
points:
1
Calibration inlets:
2
Purge gas inlets:
1
Gas connections:
Clamp ring connection 6 mm
Max. gas inlet pressure:
20 mbar relative
Min. gas inlet pressure:
-100 mbar relative
Integrated fine filter:
yes
Calibration gas
Calibration interval:
manual or automatic (configurable between one
hour and up to several weeks)
Duration of calibration:
10 minutes (recommended by the manufacturer)
Gas consumption:
5 l/calibration
Power supply
Voltage:
100 - 240 V 50/60 Hz
Power consumption:
100 VA max.
Protection class:
I
Degree of protection:
IP40
Interfaces
Relay:
3
Dig. interface:
RS232
Field bus:
optional
Optional relay:
optional
Remote Control Unit:
optional
Ambient conditions
Operating temperature:
5 - 45 °C
Humidity:
0 - 95 % relative humidity
Ambient pressure:
900 - 1250 hPa (0.9 - 1.2 bar)
Storage temperature:
-20 - 60°C
Weight
Weight:
approx. 10 kg
Содержание INCA1021
Страница 1: ...Apr 2016 V1 10 Translation of the original operating instructions Process gas analyser INCA1021 ...
Страница 3: ...3 Dimensions ...
Страница 10: ...EU Declaration of Conformity 10 ...
Страница 16: ...Safety equipment 16 3 2 Markings and warnings Fig 3 1 Markings and warnings 1 Type plate 1 ...
Страница 44: ...Operation 44 ...
Страница 46: ...Decommissioning switching off 46 ...
Страница 50: ...Maintenance 50 ...
Страница 54: ...Service 54 ...
Страница 56: ...Associated documents 56 ...
Страница 58: ...Disposal 58 ...
Страница 60: ...Spare parts 60 ...
Страница 61: ...61 16 Annex ...
Страница 62: ...Annex 62 EU Declaration of Conformity Flame arrester ...
Страница 63: ...Annex 63 ...