Maintenance
51
Maintenance/Replacing components
Interval
(recommended)
Half-yearly service and after
commissioning
Check and, if necessary, update firmware
version
every 6 months
Save the current configuration with INCACtrl
every 6 months
Annual service
Replace integrated filters
every 12 months
2-yearly service
Replace pump hoses
every 24 months
Replace flame arrester
1)
every 24 months
8-yearly service
Replace integrated pressure reducer
every 8 years
If necessary
Replace gas-delivering pumps
if necessary
Replace sensor, lifetime depends on sensor
type, message in display “Service [Typ] age”
or “Service [Typ] usage”
if necessary
1)
if installed