Operation
52
9.4 General information
Fig. 9.2:
General information
Display
(example values)
Information
V 4.39R09
Version number of the software
04.02.2015 10:05:09
Current date / time
Methane: 95.01 Vol%
Volume share of methane
mV=24.56
mV signal of the thermal battery
V=1.071
Voltage signal of the density measuring cell
T1=29.04/0.24
Inlet temperature into the thermal body / heating of the
air
p=4.00
Air differential pressure
Fr=51.13
Controlled fan frequency
Stabi=0.011
Standard deviation of the measurement value over 2
minutes
NOTE
STD: During a calibration, 0.015 is typically achieved. The calibration is
concluded after that.
Содержание CWD2005 SPC
Страница 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Страница 5: ...Technical data 5 1 Technical data 1 1 Dimensions Fig 1 1 Housing dimensions Weight approx 54 kg ...
Страница 10: ...Technical data 10 ...
Страница 12: ...EU Declaration of Conformity 12 ...
Страница 18: ...Safety information 18 ...
Страница 22: ...Protective equipment 22 ...
Страница 32: ...Transport installation and acceptance 32 Fig 6 1 Wall mounting Fig 6 2 Dimensions wall mounting ...
Страница 48: ...Description of the work stations operator control elements 48 ...
Страница 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Страница 76: ...Operation 76 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 87: ...Operation 87 ...
Страница 88: ...Operation 88 ...
Страница 90: ...Decommissioning Switching off 90 ...
Страница 94: ...Maintenance 94 ...
Страница 100: ...Troubleshooting 100 ...
Страница 102: ...Service 102 ...
Страница 104: ...Disposal 104 ...
Страница 106: ...Spare parts 106 ...