Operation
53
Main menu - Options - I/O - Digital outputs
The digital output can e.g. comprise signals in combination with limit values or
operation messages.
Example:
The Wobbe signal in MJ/m³ has an alarm value at 47,000 and 50,000 MJ. The zero
position is high. The relay for the signals described in the following are switched
under the following conditions:
Process: The solenoid valve for process gas is open and the flame is burning.
Calibrate: The solenoid valve for calibration gas is open and the flame is
burning.
Service: Filter change of air filter (the frequency of the frequency converter is
> 48 Hz) or air inlet temperature > 41°C.
Filter change: Filter change of air filter (The frequency of the frequency
converter is > 48 Hz)
Fault: The air filter is heavily soiled. The differential air pressure is thus < 3.5
mbar
Operation: The flame is burning.
Operation delayed: The flame is burning and the delay time has expired.
Over temperature: The flame is too hot. The Wobbe signal is larger than 76
mV, i.e. the temperature increase in the inner pipe of the thermal battery is
more than 50°C.
Calibration deviation: In a basic calibration you can set a signal which will
trigger and alarm when the deviation to the basic calibration exceeds the
alarm value. This is possible in negative and positive direction.
Int. pressure: An alarm is triggered in case of lack of process gas or lack in
the calibration gas line.
Содержание CWD2000 EX
Страница 1: ...April 2015 V1 04 Original operating instructions Combustion calorimeter CWD2000 EX ...
Страница 5: ...Technical data 5 1 Technical data 1 1 Dimensions Fig 1 1 Dimensions of rack System overview exemplary ...
Страница 6: ...Table of contents 6 ...
Страница 9: ...Technical data 9 ...
Страница 10: ......
Страница 11: ...EC declaration of conformity 11 2 EC declaration of conformity Further information enclosed documentation ...
Страница 12: ...Table of contents 12 ...
Страница 19: ...Safety devices 19 4 7 Markings and warnings Fig 4 1 Markings and warnings exemplary 1 2 1 3 6 5 4 7 ...
Страница 20: ...Table of contents 20 2 3 ...
Страница 21: ...Safety devices 21 4 5 ...
Страница 38: ...Table of contents 38 ...
Страница 42: ...Table of contents 42 ...
Страница 44: ...Table of contents 44 Fig 9 1 Operating elements layout of display 4 2 6 7 2 1 9 3 5 8 4 1 4 3 4 4 5 1 10 ...
Страница 48: ...Table of contents 48 9 6 Menu structure ...
Страница 68: ...Table of contents 68 Main menu Options System Settings Change colour Here you can change the display colours ...
Страница 81: ...Operation 81 ...
Страница 82: ......
Страница 86: ...Table of contents 86 ...
Страница 90: ...Table of contents 90 ...
Страница 98: ...Table of contents 98 ...
Страница 100: ...Table of contents 100 ...