28
Safety Standards:
IEC / EN 61010-1
IEC / EN 61010-2-032
CAT. III 600V, Pollution Degree 2
Electromagnetic Compatibility Standards (EMC):
EN 61326-1
Application Standards:
IEC 61557-
13: Class 2, ≤ 30A/m
@ In: 3.5 to 600 mA, Fn: 40 to 1k Hz
Valid at best resolution in 6 mA, 60 mA and 600 mA
range.
When measuring under the following condition specified
in IEC 61557-13, add additional
specification to accuracy specification.
Intrinsic uncertainty
or influence quantity
Designa
-tion
code
Additional Spec.
Position
E1
±1% of reading
Supply voltage
E2
No additional error
Temperature
E3
0.1 x (spec.accuracy)
/ °C (<18 °C or >28
°C)
Distorted waveform
E9
No additional error
External low
frequency magnetic
field 15 Hz to 400 Hz
according to
IEC 61000-4-8
E11
Add ±10μA error
per 1uT (magnetic
field)
Load current
E12
Add ±6μA error
per 1A (load current)
Touch current caused
by common mode
voltage
E13
No additional error
Frequency
E14
No additional error
Repeatability
E15
No additional error