4.2.7 Manually set light calibration offset
Calibration is done at the factory, however, in the case where calibration values must be restored or set
to something else, this command allows a new calibration value to be placed into the SQM-LU.
Executing the command “
zcal5########.##x
” manually sets the light calibration offset to the
value specified in “
########.##
”. The units are
magnitudes
arcsecond
2
.
The format of the response is:
Column Example value
Description
0
z
Calibration response is being returned.
2
5
Manual Set Light Calibration Offset
4-15
00000017.60m
Value that was set into EEPROM
16-17
Carriage return (0x0d), Line feed (0x0a).
Table 8: Response for manual setting of light calibration offset
An example is:
z,5,00000017.60m
0123456789
10
1234567
Revision 1.1
Page 14
Last modified: Jun 18, 2011