7 Unihedron Device Manager
7.6 Calibration tab
The Calibration tab is used to show and set the calibration data for the selected SQM device.
Figure 7.33: Calibration tab (initial)
Press the “Get Calibration Info” button so that the updated calibration data is brought in from the SQM on to the
Actual values boxes as show below.
Figure 7.34: Calibration tab (populated)
Use the following steps to change the calibration values:
1. Enter the new calibration data on the left side entry box (Desired values column).
2. Press the associated “Set” button.
3. The value is sent to the SQM, and then confirmed in the right side box (Actual values column).
Note that temperatures set in to the SQM use their own resolution and may not be reflected as the same value entered.
For example. 24.7
°
C might read back as 24.8
°
C.