Manual – Starflow QSD Model 6527B and 6537A
Unidata Manual - 6527B and 6537A Starflow QSD User Manual 17 01 2020.docx
Page 19
6.0 CONFIGURATION - US
IN
G THE SONIC CONFIGURATION SOFTWARE
The Starflow QSD as supplied is configured for unidirectional flow (positive flow measurement).
Changes to the instruments configuration can be done using extended SDI-12 commands but this
method is sometimes confusing and inconvenient. A much easier way is to use the Unidata Sonic
Instrument Configuration Software
The “Sonic” instrument configuration Software can be supplied with the Starflow QSD and is a stand-
alone utility that is copied to the operator’s office and / or laptop computer. This allows a user friendly
user interface that is able to communicate directly to the QSD instrument and used to customise the
Instruments internal setup.
To be able to use this Starflow QSD utility the operator must also purchase a 6603WFTDI
programming cable. This cable is simply plugged into a computer USB port and the Sonic Instrument
Configuration Software is run.
The following screenshots are used to explain the functionality of the software and the various
parameters that the user is able to change.
As with all configuration changes, care must be taken to mitigate the ramifications of these changes.
Always make note of the original settings before making adjustments so that a recovery path is
maintained.
6.1 Setup – Connection to Starflow QSD
Wire up the Starflow QSD to the terminal block at the end of the cable ensuring cable colours are
followed correctly.
Use the table below for these connections.
Starflow QSD / Wire Colour
USB Cable End
White
White (SDI-12)
Brown
Brown (+12V Power)
Green
Green (Ground)
The USB interface provides power to the QSD instrument so you do not need to provide external
power to the instrument.
Plug the USB end of the cable into your computer. If it is the first time the interface cable is used the
computer may need to load device drivers over the internet. Please ensure you have internet access
during this time. Patience may be needed as this process may take up to five minutes. Once the
drivers have been installed you are ready to run the “Sonic” software.
Once the “Sonic” software is run the following screen will appear.
Click the “Connect” button on top left of screen and the following will show.