UTR2810E Series LCR Meter
– User Manual
17
completed. It can be escape or continue the short-circuit zero clearing test.
Step 5
LCD displays information “keep fixture short”
;
If the instrument need to do short circuit zero clearing,
connect the short-circuit plate to the test fixture, or connect the test wire black clip and red clip
together;
Step 6
Press
【
ENTER
】
key to start the short circuit zero clearing. LCD displays information “short-circuit
clear is processing”
;
the instrument will automatically scan short circuit zero clearing test for each
range;
Step 7
After that, LCD displays" press ESC EXIT
;
press C
LEAR continue”. This step is completed, which
means short circuit zero clearing is completed. It can be escape or continue the open-circuit zero
clearing test again.
1
)
After the instrument the zero clearing is done, the test condition is changed (replace test fixture,
test clip, temperature and humidity environment), the instrument should do zero clearing again.
2
)
During zero clearing, press
【
ESC
】
key to escape the current test and return the test state, the
original zero clearing data remains unchanged.
3
)
During short zero clearing, it may occasionally occur “FAIL”
,
in this case, the low resistance
short circuit
(
the test wire clip or the short-circuit plate
)
may not be used or the contact is not