Overview

UT330+ series temperature humidity meters (hereafter referred to as “Meters”) 
adopts high precision sensor and super-low power consumption microprocessor 
in its design, featuring high accuracy, manual and auto storage, USB data transfer, 
USB power supply, backlight alarm, real-time display, low power consumption, etc. 
These Meters can be used in factories, laboratories, research institutes, food and 
health, environmental protection and other fields.  

UT331+/UT332+
Digital Temperature Humidity Meters

Unpacking Inspection

Operating Manual --------------------------------- 1 piece
Battery ----------------------------------------------- 4 pieces
USB Data Cable ----------------------------------- 1 piece

Safety Information

     

Warning

A warning identifies conditions and actions that pose hazards to the user

. To avoid 

electrical shock or personal injury, follow these guidelines: 

Meter Description

Meter Structure (See Figure 1)

Description

Protective cover of sensor 
(turn to open or close)

Display screen

Power on/off

Press to display Max/Min value among 
current readings. Long press to exit the 
function (also used as     button).

Press to save data. Refer to “Store and 
Recall Data” for details

To set Meters. It is used together with  
buttons. See “Setting Meters” for details.

Short press to turn on/off the backlight, 
long press to turn on/off alarm function.

Press to freeze the displayed readings, press again to exit the function. 
(also used as     button)

Press to view data, it is used together with       buttons. See “Store and 
Recall Data” for details.

Dew-point temperature and wet bulb temperature selection key 
(also used as confirm button).

Display Screen

Description

HI: Upper limit. LO: Lower limit

Dew-point temperature

Wet-bulb temperature

Storage group number

Maximum value

Minimum value

Humidity display zone

Data hold

Humidity unit

Description

Auto power off

Low battery/USB power supply

Alarm

Humidity icon

Temperature icon

Celsius unit

Temperature display zone

Fahrenheit unit

Time display zone

No.

No.

Setting Meters

     

SETUP Options

Short press SETUP button to enter into setting mode, first enter into unit setting 
page, and then press recyclable to switch the setting state until the setting is storage 
and exit. On any setting page, the setting mode will exit and be back to measuring 
page if no any button is pressed for 15 seconds. The cycle sequence:

Measuring state

Temperature unit setup

Time interval setup

Temperature & 

Humidity Upper Limit Alarm Setup

Temperature & Humidity Lower Limit Alarm 

Setup

Sleep time setup

System time setup

Temperature & Humidity Offset setup

 

Measuring state.

     

Performing the Setups

1.

 

Setting Temperature Unit

    

Short press SETUP to access temperature unit setup mode, press  

   

(HOLD)/

   (MAX/MIN) to select 

 or 

 unit.

2.

 

Setting Time Unit

    

Short press SETUP to switch to time interval mode, “int” shows on LCD.

 

Press

          

   

 

  to change the interval for automatic data storage. Long press      to increase

   /decrease the time quickly. The interval can be set up to 59:59 and down to 00:00 

    (auto data storage disabled at 00:00).

3.

 

Upper-limit temperature & humidity setting

    

Short press SETUP for 3 times to access upper limit setup mode, “HI” shows on 

    LCD, short press ENTER/DP to switch upper-limit temperature & humidity, 
    upper-limit dew-point temperature, upper-limit wet-bulb temperature, press  
    

change the value.

to 

4.

 

Lower-limit temperature & humidity setting

    

Short press SETUP for 4 times to switch to lower limit setup mode, “LO” shows 

   on LCD, short press ENTER/DP to switch following setups: lower-limit temperature 
   & humidity, lower-limit dew-point temperature, lower-limit wet-bulb temperature, 

    press      to change the value.

5.

 

Sleep Mode Setting

    

Short press SETUP for 5 times to switch to sleep setting mode, “SLP” icon shows, 

    

press      to set up sleep time (5~60mins), long press

to increase/decrease 

      

    

quickly. When the set time is 

5mins, “OFF” icon shows and sleep 

the time 

    

disabled.

function is 

    Inspect the case before using Meters. Do not use Meters if it appears damaged. 
    Look for cracks or missing plastic. Pay particular attention to the insulation 
    around the connectors.
    Replace the batteries as soon as the battery indicator “      ” appears. 
    Do not use Meters if it operates abnormally. Protection may be impaired. When 
    in doubt, have Meters serviced.
    Do not operate Meters around explosive gas, vapor, or dust.
    When servicing Meters, use only specified replacement parts.
    Do not use Meters with any part of the case or cover removed.
    Do not clean the sensor with water. Turn to close the protective sensor cover 
    when not in use. To avoid damage to the device, please use with care. 
    Never charge the battery.
    Use the approved alkaline battery (AAA 1.5V) only.
    Pay attention to “+” and “-” marks when installing the batteries.
    Close the sensor cover and remove the batteries if Meters is not used for a long 
    time.

6.

 

System Time Setting

    Short press SETUP for 6 times to switch to system time setting mode, short 
    press ENTER/DR to switch Year/Month/Day/Hour/Minute, press       to the time 
    value, long press       to increase/decrease the time quickly. Only when the minute 
    setting is completed and short press ENTER/DP again can the system time is 
    stored.

7. Setting Temperature & Humidity Offset 

    Short press SETUP for 7 times to switch to temperature & humidity offset setting 
    mode, “t-o” shows on LCD, short press ENTER/DP to  switch temperature offset 
    setting (t-o), humidity offset setting (h-o), dew-point temperature offset setting 
    (d-o), wet bulb temperature offset setting (b-o), press       to change offset value 
    that within -6

~

6.

Using Meters

  

Turn on/off Meters

1. Long press yellow button to turn on Meters.
2. With Meters already turned on, long pressing yellow button again will turn off 
    Meters.
3. After the sleep time is set, Meters will auto turn of f if no button is active within 
    the set time.
4. To ensure an accurate result, the user is recommended to wait for a while before 
    taking the readings if the measured humidity & temperature change dramatically.

  

Data Hold

1. Press HOLD button to freeze the reading, “HOLD” appears on LCD.
2. Press HOLD again to quit the function.