background image

10

9

UT320T User Manual

UT320T User Manual

6. Operation

6.1 Replace Batteries

When using the product for the first time, please install the batteries first.

To remove the battery cover:

    Hold the metal ring on the screw with a hand or a tool, and turn the ring 

counterclockwise to unscrew the battery cover screw.

    Use a screwdriver to unscrew the battery cover.

Battery type: 2 AAA alkaline batteries

Note:

    Pay attention to the battery polarity when installing.

    After replacing the batteries, close the battery cover and tighten the screw.

6.2 Power On/Off

    When the probe is folded, press the Scan|Hold button to turn on the thermometer 

and enter the infrared (non-contact) mode.

    When the probe is unfolded, it will enter the probe (contact) mode and the 

thermometer will be turned off when the probe is folded (not in the T1/T2 mode).

    When the thermometer is turned on, the backlight will automatically turn off if no 

button press occurs for nine minutes, and then the product will automatically shut 
down if still no button press occurs for one minute.

    When the probe is unfolded, after the product shuts down automatically, press the 

Scan|Hold button to wake it up.

    In the T1/T2 mode, the temperature difference can be calculated. If the probe is 

unfolded in the T1/T2 mode, fold the probe to switch to the infrared mode, and the 

thermometer will not be turned of f. If users want to fold the probe to turn of f the 
thermometer, please exit the T1/T2 mode first.

6.3 Temperature Measurement

Infrared mode:

Press the Scan|Hold button to measure the temperature, and release it to stop 
measuring. The top left corner of the LCD displays      , and the data is held.

Probe mode:

When the probe is unfolded, the product automatically starts measuring. Insert 
the probe at least 12.7mm into the measured object, and wait for the value to 
stabilize before reading the temperature. Press the Scan|Hold button. The top left 
corner of the LCD displays      , and the data is held. Press the Scan|Hold button 
again to return to the auto measurement.

6.4 Function Settings

Lock measurement in the infrared mode:

   

When the measured data is held in the infrared mode, press the Scan|Hold button 
twice,      will show in the upper right corner of the LCD and the thermometer will 
enter the lock measurement mode. The measurement can be performed without 

pressing any buttons at this time.

   

In the lock measurement mode, press the ScanlHold button to exit the lock 
measurement, and the icon      will disappear

Max/Min:

   

When the measured data is held, press the Max/Min|ε button to step through the 

Max, Min (the last continuous measurement) and exit viewing the Max/Min.

   

If the temperature is measured when the Max/Min function is enabled, the Max 

and Min of the current continuous measurement are displayed 

Emissivity adjustment:

   

Long press the Max/Min|ε button to enter the emissivity setting interface.

   

In the setting interface, press the Up/Down button to adjust the value. Short press 

once to add or subtract 0.01. Long press to add or subtract 0.1 per second.

   

After the emissivity is set, press the Scan|Hold button to save the emissivity and 

return to the measurement interface.

   

This function can only be enabled in the infrared mode.

Temperature difference:

After the thermometer is turned on for measurement, short press the T1/T2 |

℃/℉

 

button to step through T1>T2>T1-T2>T1. When T1-T2 is displayed, press the 
Scan|Hold button to exit the T1/T2 mode (or long press the T1/T2 |℃/℉ button to 
exit in any state of the T1/T2 mode).

Temperature units:

Long press the T1/T2 |℃/℉ button to step through temperature units ℃/℉.

Содержание UT320T

Страница 1: ...P N 110401111770X UT320T 2 in 1 Thermometer User Manual...

Страница 2: ...behalf of UNI T If you need warranty service within the warranty period please contact your seller directly This warranty is the only compensation you can obtain UNI T will not be responsible for any...

Страница 3: ...should only be repaired by professional maintenance personnel using original replacement parts Replace the batteries when the low battery indicator shows to prevent incorrect measurements Please check...

Страница 4: ...ty Value display Temperature difference Infrared Probe measurement Low battery Temperature units Maximum minimum temperature Infrared sensor Single laser LCD Scan l Hold Press it to turn on the thermo...

Страница 5: ...0 or 0 01 t whichever is greater 40 t 32 32 t 572 Measuring range 40 300 40 572 Probe type Minimum measuring depth Operating time Accuracy Probe measurement NTC 12 7mm About 180h backlight on Over ra...

Страница 6: ...e is unfolded the product automatically starts measuring Insert the probe at least 12 7mm into the measured object and wait for the value to stabilize before reading the temperature Press the Scan Hol...

Страница 7: ...determine the temperature Objects of different materials have different emissivity The emissivity of most organic materials painted or oxidized surfaces is about 0 95 The user can use masking tapes or...

Страница 8: ...n this manual Repair the probe if it is damaged Cause Measured value maximum range Measured value minimum range Exceed the minimum maximum operating temperature or infrared sensor damage Low battery I...

Отзывы: