Low Alarm Limit Setting

In the HOLD interface, pressing the "SET" button twice will enter the low alarm limit setting interface. The user can 

use 

 

or “▲” to adjust the value. Short pressing will add or subtract 1 to the value each time, and long pressing 

“ ”

will add or subtract 10 each second.

Emissivity Setting

In the HOLD interface, pressing the "SET" button three times will enter the emissivity setting interface, and then 
pressing the "       " button can quickly select the preset values (P1-P5). If there is no wanted value among the 

preset values, the user can use "

▼" 

or “▲” to adjust at any of the preset values that is closest to the wanted one. 

Short pressing will add or subtract 0.01 to the value each time, and long pressing will add or subtract 0.1 each 
second.

Temperature Unit Setting

In the HOLD interface, pressing the "SET" button four times will enter the temperature unit setting interface. The 

user can use "

▼" 

or “▲” to convert the unit °C/°F.

Sound Alarm Setting
In the HOLD interface, pressing the "SET" button five times will enter the sound alarm setting interface. The user 
can use "▼" or “▲” to turn on/off this function.
Lock Measurement Setting
In the HOLD interface, pressing the "SET" button six times will enter the lock measurement setting interface. The 
user can use "▼" or “▲” to turn on/off this function.

When the function is turned on, pressing      button can set the lock measurement timing "00:00", and the selected 
time position will flash. The user can use "▼" or “▲” to adjust the value. If the user needs to turn off the timing 
function, set it to "00:00" and then it will jump to display "--:--".

D: S (Distance to Spot Ratio)

As the distance (D) between the thermometer and the measured target increases, the light spot diameter (S) of 
the measured area also increases.
The relationship between measurement distance and light spot diameter is shown in the figure below.

The light spot diameter represents 

90% of the energy in the circle

Field of View

When measuring, make sure that the measured target is larger than the light spot diameter . The smaller the target, the 
closer the test distance should be (refer to D: S for the detailed light spot diameter). It is recommended that the measured 
target be larger than twice the light spot diameter of the thermometer.

Good

So-So

Bad

Target

 

>

Light spot 

T=S

T < S

diameter (S)

Emissivity

Emissivity represents the material energy radiation. The emissivity of most organic materials, painted or oxidized surfaces 
is about 0.95. The user can use masking tapes or flat paints to cover the metal surface, use the high emissivity setting, 
and then wait for a period of time to make the surface temperatures of the tapes/flat paints and the covered object the 
same. At this point, the surface temperature of the tapes/flat paints is equal to the metal surface temperature. The 
following table shows the total emissivity ε of some metals and non-metals.

Aluminum
     Oxide

A3003 Alloy
     Oxide
     Crude

Brass
     Polishing

Oxide

Cuprum
     Oxide
     Electrical terminal board

Hastelloy
     Alloy

Inconel

     Oxide

     Abrasive blasting

     Electropolishing

Ferrum
     Oxide
     Rusting

Ferrum (casting)
     Oxide
     Non-Oxide
     Casting

Ferrum (forging)

Passivating

Plumbum
     Crude
     Oxide

Molybdenum
     Oxide

Nickel
     Oxide

Platinum
     Black

Steel

     Cold rolling
     Burnishing
     Polishing

Zinc
     Oxide

Non-Metals

     Asbestos
     Asphalt

     Basalt

Carbon
     Non-Oxide
     Graphite
     Carborundum

Ceramic
Clay
Concrete
Cloth

Glass
     Convex glass
     Smooth glass
     Nonex

Sheet material
Gypsum
Ice
Limestone
Paper
Plastic
Water
Soil
Wood

Measured surface

Metals

Emissivity

Maintenance

Cleaning

Blow away the fallen particles with clean compressed air, carefully wipe the lens surface with a moist swab, and 
clean the shell with a moist sponge or soft cloth. Be careful not to rinse with water or immerse it in water.

Replace Battery

电池

电池盖

Battery

Battery cover

Install or replace a 9V battery (1604A) according to the following steps:
1.Remove the battery cover
2.Install the battery (pay attention to the polarity)
3.Close the battery cover

Phenomenon

Display OL

Display -OL

Display Err (startup)

Battery symbol flashes

Laser is not working or dark

Inaccurate measurement

Cause

Measured value > maximum range

Measured value < minimum range

Exceed the minimum or 
maximum operating temperature

Low battery

Low battery

Unmatched emissivity, too far 
measurement distance, diameter 
of the measurement target<20mm

Measure

Stop measuring

Stop measuring

Place the thermometer at -5°C-50
(23°F-122°F) for 30 minutes

°C

Replace batteries

Replace batteries

Refer to Field of View, D:S and 
other instructions in this manual

Troubleshooting