  Alternating Current (AC) 

  Direct Current (DC) 

 

Double Insulated 

 

V.

 Technical Specifications 

1. Basic Operating Conditions 

1) Ambient temperature: (23±5)°C 
2) Ambient humidity: (45~75)% RH 
3) Waveform of measured signal: Sinusoidal wave,  =0.02 
4) Frequency of measured signal: (50±0.2)HZ 
5) Location of measured current-carrying conductor at the clamp jaws: At center 
6) Current amplitude when measuring phase frequency, phase sequence, power 
and power factor: 5A±0.2A 
7) Voltage amplitude when measuring phase frequency, phase sequence, power 
and power factor: 220V±20V 
8) Electromagnetic interference to external-reference frequency: Shall be avoided. 

2. Rated Operating Conditions 

1) Ambient temperature: (-15~+45)°C 
2) Ambient humidity: (0~90)% RH 
3) Operating altitude: 

1500 m 

4) Waveform of measured signal: Sinusoidal wave,  =0.05 
5) Frequency of measured signal: 45~65HZ 
6) Current amplitude when measuring phase frequency, phase sequence, power 
and power factor: 20mA~1000A 
7) Voltage amplitude when measuring phase frequency, phase sequence, power 
and power factor: 20V~600V 
8) Location of measured current-carrying conductor at the clamp jaws:  Any 
location 

3. General Specifications 

Functions 

Measure 3-phase AC voltage, current, phase between 

voltages, phase between currents, phase between voltage 

and current, frequency, phase sequence, active power, 

reactive power, apparent power, power factor, and current 

vector sum; identify transformer connection section, 

inductive circuit, and capacitive circuits; and  test 

secondary loop and bus differential protection system, 

display the phase relation between differentially-protected 

CTs of each group, check if the wiring of electric energy 

meter is correct, repair equipment of measured line, etc. 

Содержание UT268B

Страница 1: ...P N 110401111944X UT268B...

Страница 2: ...erating Conditions 4 2 Rated Operating Conditions 4 3 General Specifications 4 4 Performance Specifications 6 VI External Structure 7 VII Operating Instructions 8 1 Power On Off 8 2 Data Hold Cancella...

Страница 3: ...n is correct when performing phase test Remove the test lead from measured conductor before disconnecting the test lead with the voltammeter Do not make contact with the terminal by hand to avoid elec...

Страница 4: ...in electric power petrochemical industry metallurgy railway industrial and mining enterprises scientific and research institute metrology service electric energy charging system relay protection syste...

Страница 5: ...veform of measured signal Sinusoidal wave 0 05 5 Frequency of measured signal 45 65HZ 6 Current amplitude when measuring phase frequency phase sequence power and power factor 20mA 1000A 7 Voltage ampl...

Страница 6: ...er 0 1W Reactive power 0 1VAR Apparent power 0 1VA Power factor 0 001 Current vector sum 1mA Phase sequence If the phase sequence of U1 U2 and U3 or I1 I2 and I3 is positive then the cursor flashes fr...

Страница 7: ...0V 9 99V 0 01V 1 5 rdg 3dgt AC 10 0V 99 9V 0 1V 1 5 rdg 3dgt AC 100V 600V 1V 1 5 rdg 3dgt Current AC 0 0mA 9 9mA 0 1mA 1 5 rdg 3dgt AC 10 0mA 99 9mA 0 1mA 1 5 rdg 3dgt AC 100mA 999mA 1mA 1 5 rdg 3dgt...

Страница 8: ...UT268B...

Страница 9: ...sable data hold When the voltammeter holds data it performs automatic numbering stores the data held currently and displays the numbering of groups i e Save 002 500 groups of data can be stored at mos...

Страница 10: ......

Страница 11: ...sts Single phase testing Connect the measured voltage lines L and N with U1 YELLOW and COM BLACK terminals correspondingly then the current clamp I1 clamps the measured line L U2 GREEN COM BLACK and I...

Страница 12: ......

Страница 13: ......

Страница 14: ......

Страница 15: ...ely Charger USB cable 1 set User manual 1 pcs The content of this user manual cannot be used as a reason for using the product for special purposes The company is not responsible for other losses caus...

Отзывы: