28
29
Press PEAK button to return to general measuring mode if under PEAK
mode
The meter will automatically return to general measuring mode after data
delete
Attention! For the sake of your safety, disconnect the meter with the
conductor and do not keep it connected for a long time.
Notes:
(3) PEAK Measurement
Peak Measuring: to obtain maximum current value. The meter can
automatically compare changing currents, hold the existing maximum
value of current and still keep it after disconnecting the meter to the
conductor. Such operation is often used when direct reading on LCD data
is not available.
Press PEAK button to switch from general measuring mode to PEAK
Mode, indicated by PEAK light. Then the meter will display and
automatically hold the maximum current value.
If under other modes, you need to return to measuring mode before
starting PEAK measurement.
Press PEAK to exit and go back to general measurement mode,
indicated by
End
during the process.
(4) Data Hold
Press HOLD button to maintain data on LCD under general measuring
mode, indicated by
HOLD
icon, repress to cancel the operation and
switch into general measuring mode, then
HOLD
icon will disappear.
(5) Data Save
Under measuring mode, the meter will automatically number the current
data when pressing HOLD button to maintain data.
MEM
icon
flashes during this process. The maximum storage is 99 sets and
achieved when
FULL
icon continuously flashes to indicate you should