1. General Specifications:

   

 Maximum voltage including transient overvoltage between

any terminals and grounding:500V rms

    Display: Maximum display 1999

    Auto Polarity Display

    Overloading                  : Display OL or - OL

    Low Battery Indication  : Display

    Measurement Speed    : Updates 3 times/second

    Measuremnet Deviation: When the conductor being meaured 

                                            

is not placed in a correct position 

                                            

during AC current measurement,

                                            

current measurement, it will cause

                                            ±1% reading deviation.

                                                          

    

To obtain accurate reading, measure only one conductor

at each time.

    When current measurement has been completed, disconnect

the connection between the conductor under test and the 

jaw, and remove the conductor away from the transformer

jaw of the Meter.

    The buzzer beeps once when pressing any buttons at any

rotary switch positions except at 2/20A positions if the button

is valid. If the button is not valid, it does not beep. At 2/20A

rotary switch position, the buzzer is set not to beep.

    The buzzer beep 5 times continuously on around 1 minute 

before entering the sleep mode. When it is just before entering

the sleep mode, it will have one long beep to warn you.

    

The LCD displays OL indicating the circuit being tested

is open.

     When continuity testing has been completed, disconnect

    

In a circuit, a good diode should still produce a forward

voltage drop reading of 0.5V to 0.8; however, the reverse

voltage drop reading can vary depending on the resistance

of other pathways between the probe tips.

    Connect the test leads to the proper terminals as said

above to avoid error display.

  

  The LCD will display OL indicating either open circuit or

wrong polarity connection.

    The unit of diode is volt(V), displaying the forward voltage

drop readings.

    Measuring the tested object that is already removed from

the in-line circuit can help to obtain a more accutate reading.

    When diode testing has been completed, disconnect the

connection between the testing leads and the circuit under

test and remove testing leads from the input terminals.

    If Ω reading with shorted test leads is not≤0.5Ω, check

for loose test leads, wrong function selected, or enabled

data hold function.

    

The LCD displays OL indicating open-circuit or the teste

resistor value is higher than the maximum range of the Meter

    Resistance measurement is default to auto range mode.

    

Measuring the tested object that is already removed

from the in-line circuit can help to obtain a more accu-rate

reading.

    

When resistance measurement has been completed,

disconnect the connection between the testing leads and

the circuit under test and remove testing leads from the

input terminals.

 (See Figure 8)

 (See Figure 7)

F. Measuring Temperature

G. Measuring AC Current

D.Testing Diodes

 (See Figure 6)

Warning

  To avoid damages to the Meter or to the devices under

test, disconnect circuit power and discharge all the

high-voltage capacitors before testing diodes.

    

Use the diode test to check diodes, transistors, and other

semiconductor devices.  The diode test sends a current 

through the semicondutor junction, then measure the

voltage drop across the junction.  A good silicon junction

drops between 0.5V and 0.8

V.

     To test the diode out of a circuit, connect the Meter as

follows:
 1.  Insert the red test lead into the               terminal and

V

      the black test lead into the 

COM

 terminal.

 2.  Set the rotary switch to            and press 

SELECT

 button  

          

to select       measurement mode.

 red      black

(

 

Figure

 

6)

then 

Specifications

 Maintenance

This section provides basic

 

maintenance information including 

battery replacement instruction.

Warning

A. General Service

B. Replacing the Battery 

(S

ee F

ig

u

r

e 10

)

Do not attempt to repair or service your Meter unless

you are qualified to do so and have the relevant calibration,

performance test, and service information. To avoid 

electrical shock or damage to the Meter,do not get water

nside the case.

Periodically wipe the case

 

with a damp cloth and mild

      detergent. Do not use abrasives or solvents.

T

o clean the terminals with cotton bar with detergent,

       as dirt or moisture in the t

e

rminals can affect readings.

T

urn the Meter power off when it is not in use.

T

ake out the battery when it is not using for a long time.

Do not use or store the Meter in a place of humidity,

       high temperature, explosive, inflammable and strong

       magnetic field.

(

 

Figure

 

10)

Screw

Warning

To avoid false readings, which could lead to possible

electric shock or personal injury, replace the battery

as soon as the battery indicator “     ” appears.
Make sure the transformer jaw and the tets leads are

disconected from the circuit being tested before

opening the case bottom.

T

o replace the battery:.

 1.T

urn the Meter off and remove all the connections

      from the input terminals

 2. T

urn the Meter’s case top down.

 3. Remove the screw from the battery compartment,

     and separate the battery compartment from the case

     bottom.

 4. Remove the old battery from the battery compartment.

 5. Replace the battery with 2pcs of new 1.5V (AAA)

     battery

.

 6. Rejoin the case bottom and the battery compartment,

     and reinstall the screw

.

** END **

This operating manual is subject to change without notice.

F

Opening

 Tested

H

Display sinewave RMS( AVG response)

sinewave RMS (AVG response)

 RMS value

 (See Figure 9)

2.

3.

REV.1

DATE:2015-12-04

P/N:110401105587X 

 

3.  For forward voltage drop readings on any semiconductor

component, place the red test lead on the component,s

anode and the black test lead on the component,s cathode.

Note

    To avoid damages to the Meter or to the devices under

test, disconnect circuit power and dicharge all the 

high-voltage capacitors before measuring continuity.

    To test for continuity,connect the Meter as follows:

  

1. Insert the red lead into the               terminal and the

 black test lead into the

 COM

 terminal.

 

 2. Set the rotary switch to               and press 

SELECT

 button to select measurement mode.

 

 3. The buzzer sounds if the resistance of a circuit under 

test is less than 50 Ω.

  4. The buzzer may or may not sounds if the resistance of

a circuit under test is between 50 Ω to 120 Ω.

  5. The buzzer does not sound if the resistance of a circuit 

under test is higher than 120Ω.

Note

the connection between the testing leads and the circuit under

test and remove testing leads from the input terminals.

release the Meter slowly until the transformer jaw is completely 

closed, Make sure the conductor to be tested is placed at 

the center of the transfor- mer jaw, otherwise it will casue 

deviation.

 

4. The measured value shows on the display,it is a effective

value of sine wave (mean value response).

Note

Sleep Mode

    

To preserve battery life, the Meter automatically turns off

if you do not turn the ratary switch or press any button for

around 15 minutes.

    

The Meter can be activated by turning the rotary switch

or pressing any button with the following conditions:

 1) When the Meter enters Sleep Mode at temperature

functions of Model: UT202, the Meter cannot be activated

by turning the rotary switch to AC current ranges.

 2) Pressing any button must be according to

“ The  effectiveness

 of Functional Buttons” section.

 

3) The Hold function will be cancelled if the Meter is activated

 

 by pressing the 

HOLD

 button.

    

To disable the Sleep Mode function, Press and hold

 

HOLD

 button while turning on the Meter.