27

3. Application

Hang onto nail, hook or other objects, or attach to iron objects such as power 
distribution cabinet, photovoltaic combiner box, automobile engine hood, etc. 
As shown below:

Iron

XVI. Maintenance and Repair

The multimeter is an advanced intelligent precise instrument. The maintenance 
must be carried out by authorized personnel with performance calibration 
skills and repair knowledges.

The multimeter is designed with auto calibration technology, all parts and 
components (except for specified replaceable ones) cannot be replaced 
without authorization, so as to avoid specification deviation.

1. General maintenance

● Please clean the multimeter with wet cloth and mild cleaning agent only, 
   do not wipe the product surface with chemical solvent.
● If any problem with the multimeter is found, please stop use and send it 
   for maintenance.
● The calibration and maintenance must be performed by qualified repair 
   personnel or designated repair department.
● Power off the multimeter when not use. Remove the battery if not use for 
   a long time.
● Do not keep in environments with high humidity, high temperature and 
   strong electromagnetic fields.

2. Replace battery or fuse (Figure 11)

 Warning:

(1) When the symbol “      ” shows on the LCD, please replace the battery in 
     

otherwise the measurement accuracy can be affected. Battery 

time, 

     

1.5V AAA x3 (4.5V)

specification: 

(2) If the display does not function at all when measuring current, please 
     check if the built-in fuse is blown out. Please check the fuse as per the 
     instructions below:
     Set the rotary switch to Ω position, then short-circuit V terminal and A 
     terminal. If the displayed resistance is about 0.0Ω, the fuse is good, 
     otherwise damaged.
     If the fuse is blown out, please replace it by a new one same in original 
     specifications.
     Fuse specification: 11A/1000V (energy: 30KA) Φ10.3×38 

Operating steps:

 

    

Set the rotary switch to OFF position, then remove the test lead from input 

    terminal.
 

    

Battery replacement: Loosen the screw at the battery cover by screw driver, 

    remove the battery cover, and replace the battery. Please install the battery 
    according to the correct polarity.
 

    

Loosen the two screws at the rear cover, remove the rear cover, and replace 

    the blown fuse.

Screw

Support and battery cover

Battery

Fuse

Multimeter

UT117C User Manual

UT117C User Manual

Figure 11

Содержание UT117C

Страница 1: ...UT117C P N 110401111679X High Precision True RMS Digital Multimeter User Manual...

Страница 2: ...end guarantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misuse...

Страница 3: ...d auto power off and Bluetooth II Features True RMS to ensure accurate measurement of non linearity load The ability to measure 20A 10 seconds of transient measurement LoZ function Low impedance to pr...

Страница 4: ...asured devices and discharge all capacitors completely Before measuring current please check if the fuse of the multimeter is good and switch off the measured current before connecting the multimeter...

Страница 5: ...IEC 61010 1 CAT 600V VII External Structure 1 NCV sensing end 2 LCD display 3 Functional buttons Used to select measurement functions 4 Rotary switch 5 A terminal 6 COM terminal 7 V terminal 8 Holder...

Страница 6: ...our times at a frequency of 2Hz and the auto off function is disabled To restore auto range and auto off function please exit MAX MIN mode If the auto off function is disabled manually the multimeter...

Страница 7: ...ction Composite sinusoidal signal generated by inverter or variable frequency motor can be measured through the LPF function as shown in Figure 6 below Long press SELECT again to exit LPF function 5 W...

Страница 8: ...ent Do not input voltage over 30V DC AC to avoid personal injury Disconnect the test lead with measured circuit after all measurement operations are completed 5 Resistance measurement Figure 6 1 Set t...

Страница 9: ...asurement under ACA mode of current probe the input current amplitude shall be greater than 10 of full range The frequency response is 45Hz 400Hz For range at 600 0A 1mV 1A AC DC The current probe mea...

Страница 10: ...V AC please observe if the conductor of measured electric field is insulated so as to avoid personal injury Even the displayed result is that voltage is not present it does not mean voltage is present...

Страница 11: ...ution Accuracy a of reading b digit Overload protection Input impedance About 10M Range to ensure accuracy 1 100 of range Least significant digit under short circuit 2 digits 600 00mV 0 01mV 1 30 600V...

Страница 12: ...racy 1 100 of range The least significant digit at current range under open circuit is allowed to be 2 Note 10 20 A Perform measurement continuously for 10 seconds at most then stop for 15 minutes at...

Страница 13: ...allowed to be about 5 under short circuit 12 Current probe measurement ACA DCA 600 0A 0 1A 1 8 3 600V AC DC 600 0A 0 1A 0 8 2 ACA DCA Range Resolution Accuracy a of reading b digit Overload protectio...

Страница 14: ...a synchronizing and more Please refer to the use manual of iDMM2 0 for use of these functions 4 Uninstallation of the software Uninstall the software through the uninstallation function of mobile phon...

Страница 15: ...o not keep in environments with high humidity high temperature and strong electromagnetic fields 2 Replace battery or fuse Figure 11 Warning 1 When the symbol shows on the LCD please replace the batte...

Отзывы: