Power Switch: 

ON: Power on and probe work;     

OFF: Power off and probe do not work. 

 

Power indicator: When the power switch is on, the indicator is lighted green.   

 

Battery Low Voltage Indicator: when the voltage of the battery is lower than 6.5V, the power 

indicator will light red, reminding the users to change battery. 

 

Range Switch: press the button switch to change between H and L Range.                                                                 

 

Range Indicator: indicating the current Range choice. H level: 100A; L level: 10A. 

 

Auto Zero button: to make accurate measurement, users need to do zero degaussing before 

measuring to avoid the influence of environment factors like earth magnetic field and temperature 

shift.   

 

Auto Zero Indicator: when Auto Zero, the indicator is lighted on and in green and will be turned off 

when the Auto Zero is finished.   

 

Clamp: the clamp used for current measurement; the max diameter for the cable is 12mm.   

 

External Power Supply Plug: powered by external DC supply. 

 

Battery Box: standard 9V alkaline battery is used. Please turn off the switch and disconnect from 

the conductor under test & oscilloscope when changing battery.   

 

 

Accessories   

 

 

        

Output Cable: 1 meter

 

 

 

 
 

Power Supply Adaptor:DC12V/1.2A 

Содержание UT-P4100A

Страница 1: ...www uni trend com Low Frequency ACDC Current Probe UT P4100A 100A 600kHz UT P4100B 100A 2MHz...

Страница 2: ...oltage range and pollution rate while the probe s test terminal is connected to the other terminals Please do not surpass the safe voltage range if the fundamental insulation of the test terminal cann...

Страница 3: ...The probe s upper and lower contact surfaces is made by precise polishing process Please protect this part or the probe could be malfunction When the battery voltage is lower than 6 5V the probe will...

Страница 4: ...atalog 4 1 Summary 5 2 Introduction about probes and accessories 5 3 Electronic Characteristics 7 4 Operating instruction 9 5 Mechanic Characteristics 10 6 Environment characteristics 10 7 Maintenance...

Страница 5: ...value The auto zero function make it easy for the users to operate and it also has power and low voltage indicator as well as overload sound alert The probe can be powered by battery or external power...

Страница 6: ...on to make accurate measurement users need to do zero degaussing before measuring to avoid the influence of environment factors like earth magnetic field and temperature shift Auto Zero Indicator when...

Страница 7: ...V A Figure 4 Rise time UT P4100A 583ns UT P4100B 175ns Max operation current Please refer to voltage and current rating table Max operation voltage Please refer to voltage and current rating table Max...

Страница 8: ...600 AC peak to peak value 200 20 1200 RMS CAT III 70 7 7 07 300 300 RMS CAT II 70 7 7 07 600 600 RMS CAT I 70 7 7 07 600 600 Figure 1 UT P4100A Typical Gain VS Frequency curve Figure 2 UT P4100B Typic...

Страница 9: ...ope from voltage to current display If user s oscilloscope doesn t have such function user have to manually set the corresponding attenuation ratio for instance if the probe is H level 0 01V A the osc...

Страница 10: ...on the current probe clamp If the direction of the current fits the tip output will be positive On the contrary the result is negative Adjust the vertical sensitivity of the oscilloscope to get a sta...

Страница 11: ...ction contact with water Handle Please use clean cloth or sponge to clean the handle Please use little alcohol instead of water to erase the dirt and parch it To maintain the performance of the produc...

Страница 12: ...12 9 Packing List Packaging Name Quantity Current probe 1 9V battery 1 DC12V 1 2A adaptor 1 BNC output cable 1 Tool box 1 Instruction book 1 Warranty card 1...

Отзывы: