Overpressure
Overpressure can occur when the water pressure from the
tap is more than 6 bar.
To check the water pressure of the line, proceed as follows:
1. Stop the HydroPower RO by pressing the STOP button.
`
The manometer indicates the line pressure.
`
As soon as the pump is switched on, the pressure
increases by 4 bar. If the inlet pressure is too high,
connect a valve to the water inlet or tap and regulate
the water.
3
The water pressure is checked and adjusted.
6.3
Troubleshooting