Calibration

During calibration, the sensor must be inserted in the supplied calibration 
chamber. In this there is a sponge, which should be kept moist.
The sensor should be powered for at least 5 minutes before calibration.
(in case of a new or recently regenerated sensor, or if the sensor has not 

been used for more than 2 weeks, 10-15 minutes are recommended)

During this time and during calibration, the sensor must be in the 
supplied calibration chamber!

After this waiting period, the calibration is started by pressing the button 

on the transmitter for at least 3 seconds.
If the calibration is successful, the LED flashes green 3 times.
If the LED is flashing solid green, the calibration was unsuccessful.
By briefly pressing the button this can be quitted. The transmitter then 
continues to use the previous valid calibration values.

After successful calibration, the loop current is approximately 12.3 mA 

when the sensor is in the calibration chamber and the relative value 
output has been selected.

5

Содержание OxyScan Transmitter waterproof

Страница 1: ...sor electrochemical Always store the sensor in its calibration chamber when not in use The membrane thin be careful atthe sensor tip can be cleaned carefully with a moistened cotton ball A constantly...

Страница 2: ...When closing the cable clamps be sure to fix the lock nut otherwise the clamp will rotate and its bonding will come off This allows water to penetrate into the housing Screw the housing cover in the...

Страница 3: ...l operation the LED on the transmitter is shining with light intensity blue or green depending on the measuring parameter Green relative value sat Blue absolute value mg l The output of the oxygen val...

Страница 4: ...DO value too high sensor defective or incorrectly wired short circuit Flashing red DO value too low sensor defective or incorrectly wired Flashing blue NTC error short circuit or not connected The sma...

Страница 5: ...e sensor must be in the supplied calibration chamber After this waiting period the calibration is started by pressing the button on the transmitter for at least 3 seconds If the calibration is success...

Страница 6: ...on optionally mg l or saturation 4 row clamp Normal 4 20 mA Alarm signal 24 mA Power supply 6 32 VDC min 24 mA Storage and operating temperature of the sensor 0 70 C Measuring ranges Parameter Measuri...

Страница 7: ...er use and or malfunction of the device We provide a 12 month legal guarantee on the entire measuring system In the guarantee case please return the sensor or measuring instrument with the test certif...

Страница 8: ...8...

Отзывы: