Conversion of the measured values:

Abslute value:

(Iout – 4 mA) * 1.25

( * 1 mg/l / 1 mA)
4 mA corresponds to  0 mg/l,

20 mA corresponds to 20 mg/l.

Relative value:  (Iout – 4 mA) * 12.5

( * 1 %Sat / 1 mA)
4 mA corresponds to 0 %sat,

20 mA corresponds to 200 %sat

f an error occurs, a current of 24 mA is output, the type of error is 
indicated by the LED:

Brightly red: 

DO value too high,

 sensor defective or incorrectly wired (short circuit)

Flashing red:

DO value too low, sensor defective or incorrectly wired

Flashing blue:

NTC error, short circuit or not connected

The small hole under the right cable gland must not be closed. It is 

used for pressure equalization, since the electronics contains an air 
pressure sensor. This is needed for accurate calibration and for 

calculating the saturation value.
It must not be stung with a sharp object in the hole, since the 
housing is then no longer waterproof.

4

Содержание OxyScan Transmitter waterproof

Страница 1: ...sor electrochemical Always store the sensor in its calibration chamber when not in use The membrane thin be careful atthe sensor tip can be cleaned carefully with a moistened cotton ball A constantly...

Страница 2: ...When closing the cable clamps be sure to fix the lock nut otherwise the clamp will rotate and its bonding will come off This allows water to penetrate into the housing Screw the housing cover in the...

Страница 3: ...l operation the LED on the transmitter is shining with light intensity blue or green depending on the measuring parameter Green relative value sat Blue absolute value mg l The output of the oxygen val...

Страница 4: ...DO value too high sensor defective or incorrectly wired short circuit Flashing red DO value too low sensor defective or incorrectly wired Flashing blue NTC error short circuit or not connected The sma...

Страница 5: ...e sensor must be in the supplied calibration chamber After this waiting period the calibration is started by pressing the button on the transmitter for at least 3 seconds If the calibration is success...

Страница 6: ...on optionally mg l or saturation 4 row clamp Normal 4 20 mA Alarm signal 24 mA Power supply 6 32 VDC min 24 mA Storage and operating temperature of the sensor 0 70 C Measuring ranges Parameter Measuri...

Страница 7: ...er use and or malfunction of the device We provide a 12 month legal guarantee on the entire measuring system In the guarantee case please return the sensor or measuring instrument with the test certif...

Страница 8: ...8...

Отзывы: