3. Setting up
3.1 Installation
Installation must be done at a place of low moisture, low dust particle, and also
level. The layout arrangement should be considered for setting up, overhaul, checking,
or cleaning the pump. When setting up the pump in any systems, pay attention to
temperature in atmosphere. Mount rubber isolator to absorb vibration from the pump.
On the environmental condition, refer to 「0.4.2 Store of Product」.
Do not tilt, lay its side, or operate up side down. It may cause damage.
The pump must be placed on a level place in an upright position as shown
in Fig-1, 2.
3.2 Oil charging
Remove the oil-filling plug, and then fill the oil (one time charge) to the pump or
recommended oil (SMR-100) up to specified quantity through the oil filling port. In
first oil charging, pour the oil up to maximum scale of oil gauge. Then, attach the
exhaust port. (see Fig-3)
Oil quantity must be checked or adjusted to keep the oil quantity is always in the
range of oil gauge during operating the pump.
Improper oil quantity may cause the performance deterioration of the pump, and also
may cause damage. When oil quantity is out of gauge scale ultimate pressure become
higher, and it may cause bubbling sound like "poko poko".