17
Screen 4: Cycle Count
The second screen shows the number of cycles which the
Q-Series
ultrasonic bath has run since new.
The cycle count shows all cycles which the bath has begun and includes
those cycles which are not completed; i.e. those cycles aborted by the
operator or because the fluid level was low.
Screen 5: Low power
In its factory preset mode, the
Q-Series
ultrasonic bath will maintain the
set fluid temperature. This means that it is consuming power when sitting
idle (ie. when the sonics are not running).
The
Q-Series
can be set to save power when not in use.
Setting Low Power to ON means that after five minutes of inactivity, the
Q-Series
ultrasonic bath will enter its SLEEP mode – the screen illumination
will power down, and the internal water heater will not operate.
The
Q-Series
ultrasonic bath uses less than 10Watts of power whilst in
SLEEP mode.
By setting Low Power to OFF, the
Q-Series
screen will maintain the set fluid
temperature; i.e. when the fluid temperature falls below the specified
setting, the heaters will automatically switch on, meaning the bath is
immediately ready for use when required.
When the desired option is selected, use the MENU key to move to the next
screen.