Chapter 11: Maintenance
00.053.204, Revision A
SonixTouch Q+ Service Manual
11-8
11.3 TRANSDUCER MAINTENANCE
Transducers require proper handling, care, and cleaning. Improper transducer care will cause damage to the transducer.
Transducer care includes daily inspection, daily cleaning, and disinfection whenever necessary. To clean a transducer:
All transducers may be exposed to many potentially damaging situations during use and cleaning. Regularly check them for
signs of damage:
• Cracks on the handle
• Cracks on the nose piece
• Cuts or gouges on the lens material
• Swelling of the lens material
• Cracks or other signs of damage to the connector
• Bent or damaged pins on the connector
• Integrity and flexibility of the cable.
11.3.1 Usage Guidelines
Ultrasonix recommends inspecting the transducers prior to each use:
• Ensure the transducers are always clean before they are used. There must be no ultrasound gel (from previous
imaging), any debris, films or unusual odors present.
• Ensure there are no cracks or other damage to the transducers before they are used. Inspect the transducer
surfaces for cracks and feel for cracks with finger tips as well.
Where any transducer (including, but not limited to, an intracavity transducer) is used in a clinical application of a
semi-critical nature (including, but not limited to intraoperative, transrectal, transvaginal, transesophageal, etc.),
ensure the transducer is covered with the appropriate STERILE transducer cover/sheath which has received
regulatory clearance for use. Refer to
Third Party Accessories
in Appendix B of the relevant
Extended User
Manual
for the recommended transducer cover/sheath.
11.3.2 Ultrasound Coupling Gels
The following ultrasound coupling gel is recommended for use with Ultrasonix transducers:
Table 11-2: Recommended Ultrasound Coupling Gels
Do not use lotions.
If there are any questions, contact Ultrasonix Medical Corporation.
Warnings:
DO NOT
drop the transducer. Dropping or striking a transducer against a hard surface can damage transducer elements, the
acoustic lens and electrical safety features.
DO NOT
use a cracked or broken transducer. Damaged transducers present a danger of electric shock.
DO NOT
pinch or kink the transducer cable. If the transducer housing becomes cracked or broken or if there are cuts or
openings in the cable, the electrical safety features of the transducer could be compromised.
DO NOT
immerse the external transducer beyond the first seam in any liquid. The transducer is not watertight. Immersing it
beyond the first seam can destroy its electrical safety features.
Gel Name
Manufacturer
Address
Telephone/Fax
Internet/E-Mail
Aquasonic 100
Parker Laboratories, Inc.
286 Eldridge Road
Fairfield, NJ, 07004
USA
Tel: 800.631.8888
Fax: 973.276.9510
www.parkerlabs.com
[email protected]
Caution:
Failure to use the recommended gel may cause damage and void transducer warranties.
Содержание SonixTouch Q+
Страница 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Страница 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Страница 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Страница 147: ......