RECOMMENDATIONS 

 

Each  application  has  its  priorities.    Accordingly,  the  MiniSonic  is  particularly  flexible 

and can be adapted to extreme situations, meaning that many applications are possible.  

Each value or text field is filled in by Ultraflux before the final tests and delivery. 
Nevertheless,  it  is  advisable  to  verify  them  by  running  through  the  various  menus, 

comparing  them  with  files  given  at  the  time  of  order,  or  with  your  instructions  or 
particular choices. 
 

More particularly, geometrical descriptions specific to the site and the choice of scale 

are the responsibility of the end user. 
 

During  a  run-

through  of  the  various  menus,  it  is  advisable  to  fill  in  the  “Probes  / 

Echoes” menu first, followed by those related to the Pipe and the Flow Rate / Fluid (Q) 
to obtain a measurement to be transmitted according to the instructions entered into the 
“General” menu 
 
 

FINAL CHECKS AND TESTS

   

Once installation, connections and parameters have been set up, several possible 

situations may arise:  

 

Case 1 

– The measurement displayed and transmitted meets your expectations. 

 
Case 2 

– The measurement operates but may lack accuracy or is unstable. 

 
Case 3 

– The measurement is not reliable or the MiniSonic is displaying a Fault. 

 

 

Case 1

 : Although everything appears to be OK ( Flow Rate = OK ), check the other 

values: 

 

  The echo or gain level compared to typical values. 

- For external probes, a gain  exceeding 50/60 dB reveals certain difficulties such 

as  the  wrong  choice  of  probes,  incomplete  installation  or  coupling,  rusted  pipe, 
absorbent fluid … 

- For intrusive probes, the gain level is usually low (excepted for gas) 

    The  ESC  mode  can  decide  a  high  gain  margin.  Take  it  in  account  in  your 

diagnosis.

 

 

  The quality index Q.I. is normally close to 100 %. 

By  default,  it  indicates  several  perturbations  =  electrical  interference,  passage  of 
bubbles, highly charged fluid … 

  In this case, in-depth investigation is advisable (oscilloscope). 

 

   The  displayed  speed  of  sound  should  be  close  to  that  expected.  It  is  an  index  of 

good control of the geometry or application. 

 

 

If the flow rate can be interrupted, check the Zero 

  If necessary, carry out adjustment using the Auto-Zero menus. 

  By PC and software, it is possible to set the measurement to Zero at very low flow 

(cut-off = % of Q max). 

 

Содержание Minisonic_600-2

Страница 1: ...ual chord Minisonic_600 2 Minisonic 2000 dual chord Minisonic_2000 2 B timent TEXAS ragny Parc 9 All e Rosa Luxemburg 95610 RAGNY FRANCE T l 33 0 1 30 27 27 30 Fax 33 0 1 30 39 84 34 www ultraflux net...

Страница 2: ...0 B and 2000 B for dual pipe flow measurement configurations Speed 1 or 2 is for open channel flow velocity measurements This manual is specifically concerned with an element from the family and has b...

Страница 3: ...ersion Ex d ATEX Version 4 Wiring diagram 5 Converter Assembling Instructions Mechanical installation Electrical wiring 6 Probes installation 7 Setting and commissioning 8 Detailed Menu layout Measure...

Страница 4: ...with tapping at an oblique angle and straight ended probes Or with taps perpendicular to the pipe but with bevel end probes Good geometrical control is essential The construction of a measuring spool...

Страница 5: ...afety risks N A Only low voltage 60 V Concerning EMC EN 61000 EN 55022 et EN 50204 Concerning tightness IP EN 60529 Concerning ATEX Directive 94 4 CE EN 60079 0 General rules EN 60079 1 Ex d enclosure...

Страница 6: ...C T6 IP 67 See also the specific notice NT 219 B Each touch of the keypad has the same function as above The two glands near the front are reserved for probe cables These connections request the use o...

Страница 7: ...8 7 6 5 4 3 14 13 12 BR5 2 1 1 100 mm 4 WIRING DIAGRAM FOR MINISONIC 600 2 2000 2 Low voltage supply See also NT 218 if the GP 01 90 230 V AC 24 V DC if installed The internal cables must be kept as s...

Страница 8: ...he limits V 100 V I 0 1 A R ON 10 Ohm Each relay Re 1 terminals 7 8 or Re 2 terminals 9 10 will transmit the chosen status information in the General Parameters menu 4 20 mA analog outputs BR 2B termi...

Страница 9: ...mulation of gas planes close to vertical 30 will be avoided A horizontal plane is a good choice excepted when the conduit is vertical For intrusive probes adjust the insertion then the alignment or or...

Страница 10: ...ettings using the PC software Refer to the manual supplied with the software CD ROM The fields of the main values to be entered are listed below under Heading 8 The software offers extended configurat...

Страница 11: ...els 1 2 and also averaged values Physical measurements time of flight T and difference dT channels 1 2 Totalizers Reset possible by simultaneously pressing and Date and time Change by PC Software Hydr...

Страница 12: ...t action to take place An operation will be requested allowing a provisional code to be calculated A second press on will provide access to the existing Languages choice field 8 3 Pipe Settings menu A...

Страница 13: ...f Coefficient KH or following data to do its calculation If MANUAL entry of coefficient depending on profile Laminar 1 333 or Turbulent 1 02 to1 08 If AUTO entry of parameters Fluid viscosity Pipe rou...

Страница 14: ...Otherwise If MANUAL Gain adjustment If AUTO Margin Over amplification compared to critical gain Go If SM choice position of probes depending on geometrical measurements as distance between probes fac...

Страница 15: ...gnment If copy of TOT 1 or TOT 2 relay closing time F maxi If Threshold enter value Q Reset to Zero of TOT t1 t2 This function may be prohibited Parameter settings of two outputs 4 20 mA SA 1 SA 2 by...

Страница 16: ...actory Settings menu normally reserved to Ultraflux It permits to adjust a coefficient which optimises the compensation with Clamp on probes value 0 6 to1 6 Another parameter named Delta T band permit...

Страница 17: ...ment displayed and transmitted meets your expectations Case 2 The measurement operates but may lack accuracy or is unstable Case 3 The measurement is not reliable or the MiniSonic is displaying a Faul...

Страница 18: ...Diameter Flow Rate Assembly and Choice of probes is too low to claim any measurement accuracy Is it possible to revise these choices The measurement conditions reveal poor control of transitions from...

Страница 19: ...T 4 Echo Alternating signal peak to peak 4 Volt max 0 65 V negative threshold voltage PT 0 0 Volt Ground PT 2 Synchro Trig on Positive edge 0 to 5 Volt PT 1 Selection windows Echo must be inside PT 3...

Страница 20: ...NT 208D GB1 20 20 11 APPENDIX Calibration file example with a MiniSonic 2000 2 on a flow spool with wetted probes SM...

Отзывы: