background image

11. Maintenance and Troubleshooting 

11.1  Maintenance 

Kronos does not contain any consumable items or spare parts. The exterior of Kronos can be 
cleaned with a soft cloth moisten with ethanol. Never use acetone. Replace the lids to ensure no 
contaminants or dust will fall into the modules. Avoid cleaning inside the modules due to the risk 
of optics and crystal contamination, and accidental misalignment. If absolutely necessary, ensure 
sufficient ventilation to minimize fume concentration in the modules. 

Clean optical surfaces allow for optimal reflection or transmission of laser power. Contaminants 
on the optics can reduce the efficiency of the system. Touching the polished surface of the optics 
in anyway, even with new gloves, can cause contamination. It is crucial that optics are handled at 
the ground edges only. Exercise caution when handling optics and when cleaning them. Always 
inspect the optics before and after cleaning them. Use a magnification device and a bright light to 
spot small defects and contaminants. 

Clean optics in increasing level of contact. Use non-contact methods by blowing off dust and other 
contaminants with compressed air. Be careful not to use excessive pressure that might cause the 
optics to slip or crack. 

If blowing is not sufficient to remove remnant contaminations, e.g. oils, spectroscopic-grade 
methanol is the recommended solvent for cleaning optics. Other solvents and poorer grades can 
leave residues or degrade the optical coatings. Always read the associated data sheets and MSDS 
sheets prior to using any solvent. 

Prepare a single piece of high-quality lens tissue, folded neatly, and lightly moistened with the 
solvent. Gently use the tissue for one wipe and one direction only and discard it immediately. Use a 
new tissue for subsequent passes. Reusing the tissue or reversing direction can drag contaminants 
back across the cleaned surface, causing residual streaks and potential damage. 

33

Содержание KRONOS

Страница 1: ......

Страница 2: ...direct indirect special incidental or consequential damages resulting from any defects in this documentation 2022 Ultrafast Systems All rights reserved No part of this manual may be reproduced or copi...

Страница 3: ...g the Excitation and Detection Wavelengths 15 7 4 Acquiring Data 18 8 Familiarizing Yourself with the Reference Samples 20 8 1 Base Catalysis of the cis trans Isomerization of Congo Red 20 8 2 Determi...

Страница 4: ...mental Approach 32 11 Maintenance and Troubleshooting 33 11 1 Maintenance 33 11 2 Troubleshooting 34 12 Appendix 35 13 Warranty 36 Limitation of Warranty 36 14 Notice 37 14 1 Con fi dentiality Proprie...

Страница 5: ...used for different experiments in Kronos The inset shows the full curve note the strong spike due to the fl ash lamp 19 Figure 9 Ground state absorption spectrum of Congo Red trans conf in 20 water e...

Страница 6: ...positions with respect to each other Chemical Kinetics is one of the major divisions of Physical Chemistry and is basically the quantitative study of the rates and mechanisms of chemical reactions Che...

Страница 7: ...when it is fi lled with liquid Always remove the cuvette before changing contents Never fi ll or re fi ll the cuvette when it is in the sample holder The optimal operation temperature of Kronos is at...

Страница 8: ...uirements USB cable 28 AWG shielded Power source USB connection Supplied current 500 mA Supplied voltage 5 V Operating temperature 15 to 40 C Operating relative humidity 10 to 70 non condensing Spectr...

Страница 9: ...Detector Speci fi cations Detector type Silicon PIN Active area 3 6 x 3 6 mm 13 mm2 Wavelength range 350 to 1100 nm Rise time 14 ns Bias voltage 10 V 8...

Страница 10: ...phing software e g Microsoft Excel OriginLab Origin etc 4 Unpacking and Inspection The Kronos spectrometer is carefully packaged at the factory to minimize the possibility of damage during shipping In...

Страница 11: ...h allows rapid replacements of standard square fi lters and cuvettes The detection fi lters are mounted in a customized holder Figure 2c shows the simpli fi ed optical layout of the Kronos Figure 2 3D...

Страница 12: ...pump source Sample 10mm 10mm cuvette holder L1 L2 Lenses PD Silicon photodiode Exc fi lter Holder for 25mm 25mm 2mm fi lters for excitation wavelength selectivity Det fi lter Holder for 0 5 diameter f...

Страница 13: ...djust the range slider to achieve su ffi cient signal to noise Absorption Emission Select either Absorption or Emission experiment modes Run Starts measurement Save Save data in CSV format fi rst colu...

Страница 14: ...atically the Y axis range will adjust to display the signal only showing the large spike around time 0 The default X axis setting is Auto Scale To change this right click anywhere on the plot and choo...

Страница 15: ...ault cursor mode on the graph is y axis zoom ELEMENT DESCRIPTION Axis Autoscale mode OFF and ON Autoscale axis once Pan graph Select between graph zoom modes Select X and Y region to zoom in Select X...

Страница 16: ...plays a value This will be I0 i e the baseline intensity before any pump induced changes In absorption mode during data acquisition the Kronos will wait for approximately 2 sec between fl ashes to all...

Страница 17: ...Figure 4 Spectra of Kronos light sources a Xenon fl ash lamp for the pump b LED for the probe 16...

Страница 18: ...dark blue fi lter 300 450 nm 700 nm has high transmittance around 350 nm and allows the Xe lamp to function as a UV blue light source If you wish to pump with a separate wavelength region e g 500 nm...

Страница 19: ...5 nm Any incident light within this bandwidth has high transmittance Figure 6 Transmission spectrum of 600 nm detection dielectric interference fi lter 7 4 Acquiring Data Place a fi lled cuvette into...

Страница 20: ...etic interference in the circuitry and is intrinsic to the equipment Since the temporal resolution of the Kronos is around 100 s you can safely reject the artifact and simply use the datapoints after...

Страница 21: ...m of a solution of CR in its trans ground state in 20 water ethanol is shown in Figure 9 The dye absorbs strongly throughout the visible range and peaks at 510 nm Experiment Base catalysis of cis tran...

Страница 22: ...ion and measure its rate on timescales that cannot be achieved by traditional mixing methods Using both acids and bases which function as catalysts this experiment allows the user to fi nd the bimolec...

Страница 23: ...ht causing the detector to register a drop in transmission Then the excited cis state returns to the trans state ground state over many milliseconds Adjust averaging as desired then measure the other...

Страница 24: ...ion for reference Piard Jonathan In fl uence of the Solvent on the Thermal Back Reaction of One Spiropyran Journal of Chemical Education 2014 91 2105 2111 6 NO BIPS undergoes a photochemical ring open...

Страница 25: ...the kinetics of the back reaction will be measured for a range of temperatures in order to determine the activation energy E of the back reaction The literature reported value for E 62 5 kJ mol Figur...

Страница 26: ...time pro fi le is shown in Figure 15 for a temperature of 55 C Also shown in this plot is the fi tted exponential decay function red line calculated using OriginLab Origin a commercial graphing softwa...

Страница 27: ...Figure 15 Example decay and exponential fi t for 55 C measurement Figure 16 Actual data example of ln as a function of 1000 T 26...

Страница 28: ...and the time window 9 3 Finding the Signal 1 Place the appropriate excitation and detection fi lters for your sample A prior UV vis absorption measurement would help determine which are appropriate 2...

Страница 29: ...oscopists will choose different instruments to perform such experiments The most basic optical spectroscopic techniques are steady state measurements like absorption or fl uorescence spectroscopy Whil...

Страница 30: ...ses or when the signal is maximum This manual uses the former de fi nition 10 2 Photochemistry Materials in this section have been adapted from A Qualitative Theory of Molecular Organic Photochemistry...

Страница 31: ...or rate constant k is 8 This equation tells us that since the concentrations of the reactants decrease during the reaction then the rate but not the rate constant decreases with time elapsed Thus the...

Страница 32: ...ction is fi rst order in A and overall fi rst order Orders higher than 2 fractional and zero orders are found in special cases These are not considered here because they are not relevant to Kronos NOT...

Страница 33: ...product The blue line is drawn at the point of 50 reaction half of the initial concentration of A has been converted into B By inspection this occurs after an elapsed time of about 28 milliseconds Thi...

Страница 34: ...cleaning them Always inspect the optics before and after cleaning them Use a magni fi cation device and a bright light to spot small defects and contaminants Clean optics in increasing level of conta...

Страница 35: ...act your local representative or Ultrafast Systems if no light is detected 2 Try to average across more scans if the baseline intensity falls within the recommended range 3 Ensure the Range is set to...

Страница 36: ...are In this window you can change the COM port and the default graphing unit from current to A Changing the latter will instruct the software to automatically perform calculations to convert the light...

Страница 37: ...ity Repairs will be made and the instrument returned freight prepaid Repaired Products are warranted for the remainder of the original warranty period Limitation of Warranty The foregoing warranty sha...

Страница 38: ...trade secrets of Ultrafast Systems and shall devote its best efforts to ensure that all its personnel protect the Ultrafast Systems Programs as trade secrets of Ultrafast Systems Customer shall not a...

Страница 39: ......

Отзывы: