Lubricate air motor with KaVo spray cleaner or similar agent.
4.
Sterilize air motor in sterilization bags in a steam sterilizer (autoclave) up to a maximum temperature of 135º C for 15 minutes. Ensure that air motor
5.
is taken out of sterilizer as soon as sterilization is complete; otherwise, the device may malfunction.
Before re-using air motor, place Endo-Eze AET contra-angle handpiece (or equivalent) onto the air motor, run for a few seconds, and wipe away any
6.
excess oil and lubricant.
Lubrication of Endo-Eze AET E-Type Air Motor
Step by Step Instructions for Lubrication of Endo-Eze AET E-Type Air Motor
Insert spray cap (with nozzle) onto spray can. Remove blue insert in nozzle and replace with black cone-shaped insert.
1.
Shake spray can several times.
2.
Set the speed ring of the air motor to maximum position (position F or R).
3.
Fit the drive air tube (the smallest air intake) of motor onto the spray cap.
4.
Keep spray can upright.
5.
Spray for approximately one second.
6.
II. Using the Endo-Eze AET Air Motor and Contra-Angle
Operating Pressure
The operating pressure is measured as flow pressure.
•
Operating pressure possible with 32 psi (2.2 bar) to 43 psi (3.0 bar).
•
If only the air motor is used on the supply hose, set the pressure to 43 psi (3.0 bar).
•
If the air motor is used alternately with a turbine on the same supply hose, it is essential to ensure that the pressure set for the two drives is within
•
the prescribed range.
Operating Instructions
Connect air motor to air supply hose. Align the motor back-end tubes with holes in the supply hose connector.
1.
Push the motor onto the supply hose connector.
2.
Turn the supply hose nut and securely fasten it. If loose, air and/or water may leak, and air pressure to the motor will become insufficient.
3.
External cleaning
- manual:
- machine:
- brush the instrument under running water
- clean the exterior of the instrument using a 60% alcohol
solution
- Miele Dental Washer Disinfector G 7781 or compatible
device
Internal cleaning
- machine:
- Miele Dental Washer Disinfector G 7781 or compatible
device
External disinfection
- manual:
- machine:
- KaVo ELASTOclean 2, S&M MicroCid or similar surface
disinfectant
- Miele Dental Washer Disinfector G 7781 or compatible
device
Internal disinfection
- machine:
- Miele Dental Washer Disinfector G 7781 or compatible
device
Lubrication
- manual:
- KaVospray 2112 A, or similar lubricating agent
Sterilization
- in sterilization bag
- Steam sterilizers (autoclave class B or S)
B class: 135º C for 5 minutes
S class: 135º C for 12 minutes
Ultradent Recommendations for Cleaning, Disinfecting and Sterilizing Air Motor