Guidance and manufacturer’s declaration – electromagnetic 
immunity 

The

 K-020

 is intended for use in the electromagnetic environment specified below.   

The customer or the user of the

 K-020

 should assure that it is used in such an environment. 

Immunity 
test 

IEC 60601 
test level   

Compliance 
level   

Electromagnetic environment – 
guidance 

Electrostatic 
discharge 
(ESD) 
 
IEC 61000-4-2 

± (2, 4, 6) kV 
contact   
± (2, 4, 8) kV 
air   
 

± (2, 4, 6) kV 
contact   
± (2, 4, 8) kV air   
 

Floors should be wood, concrete or ceramic 
tile. If floors are covered with synthetic 
material, the relative humidity should be at 
least 30 %. 

Electrical fast 
transient/burs

 
IEC 61000-4-4 

±2 kV for 
power               
                     
supply lines 
±1 kV for 
input/output 
                     
lines 

Not applicable 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Surge 
 
IEC 61000-4-5 

±1 kV 
differential     
                     
mode 
±2 kV common   
                     
mode 

Not applicable 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Voltage dips, 
short 
interruptions 
and 
voltage 
variations 
on power 
supply 
input lines 
 
IEC 
61000-4-11 
 

<5 % 

U

(>95 % dip in 

U

T) 

for 0,5 cycle 
 
40 % 

U

(60 % dip in 

U

T) 

for 5 cycles 
 
70 % 

U

(30 % dip in 

U

T) 

for 25 cycles 

Not applicable 

Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the 

K-020

 

requires continued operation during power 
mains interruptions, it is recommended 
that the 

K-020

 be powered from an 

uninterruptible power supply or a battery. 

Содержание UCT-CMT1

Страница 1: ...ncorrect date or breakage 6 Sudden movement of the users may lead to incorrect data 7 Make sure to keep receiver charged when using it to monitor body temperature for a long period time 8 The temperat...

Страница 2: ...ing and recording APP can monitor body temperature continuously and keep data record for your doctor reference when needed any time 2 Reminding When body temperature reaches threshold you already set...

Страница 3: ...Or 0 09 F 95 F 101 3 F 0 18 F 95 F and 101 3 F Operation environment Temperature 5 40 41 F 104 F Humidity 15 RH 85 RH Receiver iOS BLE4 0 iOS8 0 and above Android BLE4 0 Android4 3 and above Getting...

Страница 4: ...he device Press the button for 2 seconds A single LED blink in Blue 5times quickly indicates power is on Open the Baby thermometer APP and create a new user Connection will be established automaticall...

Страница 5: ...Wearing Lift the arm naturally then apply the device just below the armpit Hold arm tight against side for at least 8 minutes Note Make sure the area of the skin that contacts with the device stays f...

Страница 6: ...he device in any liquid Do not sterilize or use an autoclave on the device If you notice any deterioration or damage stop using the device immediately Troubleshooting 1 The device is not switched on P...

Страница 7: ...20 as following table Table 206 Recommended separation distances between portable and mobile RF communications equipment and K 020 as following table Table 201 Guidance and manufacturer s declaration...

Страница 8: ...s t IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines Not applicable Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV dif...

Страница 9: ...IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz Not applicable 3 V m Portable and mobile RF communications equipment should be used no closer to any part of th...

Страница 10: ...erformance is observed additional measures may be necessary such as reorienting or relocating the K 020 b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Table 206...

Страница 11: ...ufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by...

Страница 12: ...is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect...

Отзывы: