EMC DESCRIPTIONS 

The EMC declaration according to the requirement of EN 60601-1-2 

Cautions: 

 

User must regard EMC, please install and put in service 

K-020

 according to the EMC 

information provided in the accompanying documents.

 

 

Portable  and  mobile  RF  communications  equipment  can  affect  medical  electrical 
equipment.

 

 

The performance of the EQUIPMENT and SYSTEM that was determined to be essential 
performance.

 

 

Table  201-Guidance  and  manufacturer’s  declaration-electromagnetic  emissions-for 

K-020, as following table.

 

 

Table  202-Guidance  and  manufacturer’s  declaration-electromagnetic  immunity  -for 

K-020, as following table.

 

 

Table  204  -Guidance  and  manufacturer’s  declaration-electromagnetic  immunity  -for 

K-020, as following table.

 

 

Table  206  Recommended  separation  distances  between  portable  and  mobile  RF 
communications equipment and 

K-020, as following table.

 

 

Table 201 
 

Guidance and manufacturer’s declaration – electromagnetic 
emissions 

The 

K-020 

is intended for use in the electromagnetic environment specified below. The customer 

or the user of the 

K-020 

should assure that it is used in such an environment. 

Emissions test 

Compliance 

Electromagnetic environment - guidance 

RF emissions   
CISPR 11 

 

Group 1 

The 

K-020

 uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to 
cause any interference in nearby electronic equipment. 

RF emissions   
CISPR 11 

 

Class B 

The 

K-020

 is suitable for use in all establishments, including 

domestic establishments. 

Harmonic 
emissions 
IEC 61000-3-2 

 

Not applicable 

 

Voltage 
fluctuations / 
flicker emissions   
IEC 61000-3-3 

 

Not applicable 

 

Содержание UCT-CMT1

Страница 1: ...ncorrect date or breakage 6 Sudden movement of the users may lead to incorrect data 7 Make sure to keep receiver charged when using it to monitor body temperature for a long period time 8 The temperat...

Страница 2: ...ing and recording APP can monitor body temperature continuously and keep data record for your doctor reference when needed any time 2 Reminding When body temperature reaches threshold you already set...

Страница 3: ...Or 0 09 F 95 F 101 3 F 0 18 F 95 F and 101 3 F Operation environment Temperature 5 40 41 F 104 F Humidity 15 RH 85 RH Receiver iOS BLE4 0 iOS8 0 and above Android BLE4 0 Android4 3 and above Getting...

Страница 4: ...he device Press the button for 2 seconds A single LED blink in Blue 5times quickly indicates power is on Open the Baby thermometer APP and create a new user Connection will be established automaticall...

Страница 5: ...Wearing Lift the arm naturally then apply the device just below the armpit Hold arm tight against side for at least 8 minutes Note Make sure the area of the skin that contacts with the device stays f...

Страница 6: ...he device in any liquid Do not sterilize or use an autoclave on the device If you notice any deterioration or damage stop using the device immediately Troubleshooting 1 The device is not switched on P...

Страница 7: ...20 as following table Table 206 Recommended separation distances between portable and mobile RF communications equipment and K 020 as following table Table 201 Guidance and manufacturer s declaration...

Страница 8: ...s t IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines Not applicable Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV dif...

Страница 9: ...IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz Not applicable 3 V m Portable and mobile RF communications equipment should be used no closer to any part of th...

Страница 10: ...erformance is observed additional measures may be necessary such as reorienting or relocating the K 020 b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Table 206...

Страница 11: ...ufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by...

Страница 12: ...is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect...

Отзывы: