Trouble shooting

 

(1)

1) No lights on power u

nit

Check the power unit is connected to the mains

power supply and that the mains switch is turned ON.

Check power unit for any blown fuses.

Low Pressure indicator is flashing

(2)

2) Low Pressure indicator is 

f

lashing (R

ED)

Check whether power was suddenly shut down.

Check that the connection between air tube and

power unit is tightly secured.

Check that all tubing connections along the mattress

are secured.

If all of above steps have been checked.

Touch “Alarm Mute” for system to be verified again.

(3)

3) Power Failu

re 

Alarm Indicator is 

f

lashing (RED)

 

and

sounding

If in use - Check the power unit is connected to

 

the 

mains power supply and that the mains switch is

 

turned ON.

If being taken out of use - Check pump has been

 

switched OFF on the panel before at power source.

 

Reconnect to power supply and turn pump ON, then

 

OFF.

Please wait a few minutes to see if the issue 

resolves. 

(4)

4)

 

The power unit is operating but the mattress

 is

 

not alternating

Ensure that the mattress inflation process is

complete.

Check that the CLP indicator on the control panel

is not illuminated. If it is turn CLP off and alternating

mode will engage. Please wait a few minutes and

check that the cells are now alternating.

(5)

5) Power unit is no

isy

• Ensure that the power unit is resting against a solid

surface.

If the problem persists, contact Ultimate Healthcare or 

your local service provider.

www.ultimatehealthcare.co.uk  |  Tel: 0333 321 8996